Radial fast spin-echo method for T2-weighted imaging and T2 mapping of the liver

被引:62
作者
Altbach, MI
Outwater, EK
Trouard, TP
Krupinski, EA
Theilmann, RJ
Stopeck, AT
Kono, M
Gmitro, AF
机构
[1] Univ Arizona, Dept Radiol, Tucson, AZ 85724 USA
[2] Univ Arizona, Opt Sci Ctr, Tucson, AZ USA
[3] Univ Arizona, Biomed Engn Program, Tucson, AZ USA
[4] Univ Arizona, Arizona Canc Ctr, Tucson, AZ USA
关键词
radial MRI; projection reconstruction MRI; T2; measurements; fast spin-echo; diffusion weighting; flow suppression;
D O I
10.1002/jmri.10142
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate a multishot radial fast-spin echo (RAD-FSE) methods developed to improve the quality of abdominal T2-weighted imaging as well as the characterization of focal liver lesions. Materials and Methods: The RAD-FSE sequence used in this work consisted of a preparatory period followed by a short echo train (ETL = 16). A novel radial k-space trajectory was used to minimize streaking artifacts due to T2 variations and motion. Small diffusion gradients (b = 1.2 mm/s(2)) were used to improve flow suppression. The quality of images obtained with RAD-FSE was compared to multishot 2DFT fast spin-echo (2DFT-FSE) and half-Fourier acquistion single-shot turbo-spin-echo (HASTE) images using data from 16 patients. A postprocessing algorithm was used to generate multiple high-resolution images (at different effective TE values) as well as a T2 map from a single RAD-FSE data set. the T2 maps were used to differentiate malignant from benign lesions for a set of 33 lesions ranging from 0.8-194 cm(3). Results: RAD-FSE produces high-resolution images of the liver in a breath-hold without the motion artifacts of 2DFT-FSE methods, and without the blurriness and loss of small lesion detectability of HASTE. The inclusion of diffusion weighting in RAD-FSE decreases the signal from blood in hepatic vessels, which improves lesion visualization. the T2 values obtained by postprocessing a single RAD-FSE data set can differentiate malignant from benign lesions. The mean T2 values obtained for malignancies, hemangiomas, and cysts are 108 +/- 30 msec, 240 +/- 14 msec, and 572 +/- 334 msec, respectively. Conclusion: These results indicate that RAD-FSE produces abdominal images of higher quality than 2DFT-FSE and HASTE. In addition, lesions can be characterized using T2 maps generated from a single RAD-FSE data set.
引用
收藏
页码:179 / 189
页数:11
相关论文
共 48 条
  • [1] Abe Y, 2000, Radiat Med, V18, P7
  • [2] ALTBACH MI, 2000, P 6 ANN M ISMRM DENV, P1621
  • [3] Preliminary experience with a new double-echo half-Fourier single-shot turbo spin echo acquisition in the characterization of liver lesions
    Bosmans, H
    Gryspeerdt, S
    VanHoe, L
    VanOostende, S
    DeJaegere, T
    Kiefer, B
    Baert, AL
    Marchal, G
    [J]. MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 1997, 5 (01) : 79 - 84
  • [4] Hepatic T2-weighted MRI: A prospective comparison of sequences, including breath-hold, half-Fourier turbo spin echo (HASTE)
    Coates, GG
    Borrello, JA
    McFarland, EG
    Mirowitz, SA
    Brown, JJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (03): : 642 - 649
  • [5] FACTORS INFLUENCING CONTRAST IN FAST SPIN-ECHO MR IMAGING
    CONSTABLE, RT
    ANDERSON, AW
    ZHONG, J
    GORE, JC
    [J]. MAGNETIC RESONANCE IMAGING, 1992, 10 (04) : 497 - 511
  • [6] THE LOSS OF SMALL OBJECTS IN VARIABLE TE IMAGING - IMPLICATIONS FOR FSE, RARE, AND EPI
    CONSTABLE, RT
    GORE, JC
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) : 9 - 24
  • [7] Signal characteristics of focal liver lesions on double echo T2-weighted conventional spin echo MRI: Observer performance versus quantitative measurements of T2 relaxation times
    Fenlon, HM
    Tello, R
    deCarvalho, VLS
    Yucel, EK
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2000, 24 (02) : 204 - 211
  • [8] Liver masses: Replacement of conventional T2-weighted spin-echo MR imaging with breath-hold MR imaging
    Gaa, J
    Hatabu, H
    Jenkins, RL
    Finn, JP
    Edelman, RR
    [J]. RADIOLOGY, 1996, 200 (02) : 459 - 464
  • [9] PROJECTION RECONSTRUCTION TECHNIQUES FOR REDUCTION OF MOTION EFFECTS IN MRI
    GLOVER, GH
    PAULY, JM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (02) : 275 - 289
  • [10] VALUE OF T1-RELAXATION AND T2-RELAXATION TIMES FROM ECHOPLANAR MR IMAGING IN THE CHARACTERIZATION OF FOCAL HEPATIC-LESIONS
    GOLDBERG, MA
    HAHN, PF
    SAINI, S
    COHEN, MS
    REIMER, P
    BRADY, TJ
    MUELLER, PR
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (05) : 1011 - 1017