Gas-Liquid Two-Phase Flow Correction Method for Digital CMF

被引:19
作者
Hou, Qi-Li [1 ]
Xu, Ke-Jun [1 ,2 ]
Fang, Min [1 ]
Shi, Yan [1 ]
Tao, Bo-Bo [1 ]
Jiang, Rong-Wei [1 ]
机构
[1] Hefei Univ Technol, Sch Elect & Automat Engn, Hefei 230009, Peoples R China
[2] Engn Technol Res Ctr Ind Automat, Hefei 230009, Peoples R China
关键词
Artificial neural network; Coriolis mass flowmeter (CMF); digital drive method; digital zero-crossing detection method; gas-liquid two-phase flow; CORIOLIS MASS; DRIVE; METER;
D O I
10.1109/TIM.2014.2310031
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
There are large measurement errors in Coriolis mass flowmeter (CMF) when it is used to measure gas-liquid two-phase flow without compensation. In particular, the flowtube of CMF may stall when using analog transmitters and CMF cannot operate normally because of the high damping caused by gas-liquid two-phase. To solve above problems, a gas-liquid two-phase flow rig is developed to perform the two-phase flow experiments. A digital drive method using multiplying digital-to-analog converter and direct digital synthesizer is applied to CMF to maintain the flowtube oscillation under two-phase flow condition. A digital zero-crossing detection method based on Lagrange interpolation is adopted to calculate the frequency and phase difference of the sensor output signals in order to synthesize the digital drive signal and realize the flow measurement. The artificial neural network using back propagation algorithm is applied to correct measurement errors. A digital Coriolis mass flow transmitter is developed with a digital signal processor to control the digital drive, and realize the two-phase flow measurement and correction. By connecting the developed transmitter with a CMF025 type primary instrument, the gas-liquid two-phase flow experiments are conducted to validate the performance of the transmitter. The testing results show that the transmitter can maintain the flowtube oscillation and reduce the metering errors.
引用
收藏
页码:2396 / 2404
页数:9
相关论文
共 19 条
[1]  
Chen Kun, 2010, 2010 IEEE Symposium on Industrial Electronics and Applications (ISIEA 2010), P56, DOI 10.1109/ISIEA.2010.5679495
[2]   Non-linear Control of the Oscillation Amplitude of a Coriolis Mass-Flow Meter [J].
Clarke, D. W. .
EUROPEAN JOURNAL OF CONTROL, 1998, 4 (03) :196-207
[3]  
Freeman B. S., 1998, U. S. patent, Patent No. 5804741
[4]   An aeroelastic model of Coriolis mass and density meters operating on aerated mixtures [J].
Gysling, Daniel L. .
FLOW MEASUREMENT AND INSTRUMENTATION, 2007, 18 (02) :69-77
[5]   Coriolis meters in two phase conditions [J].
Hemp, J ;
Yeung, H .
COMPUTING & CONTROL ENGINEERING JOURNAL, 2003, 14 (04) :36-36
[6]  
Henry M. P., 2003, U.S. Patent, Patent No. [6 505 519, 6505519]
[7]  
Henry M. P., 2005, U.S. Patent, Patent No. [695 076 0 B2, 6950760]
[8]   Two-phase flow metering of heavy oil using a Coriolis mass flow meter: A case study [J].
Henry, Manus ;
Tombs, Michael ;
Duta, Mihaela ;
Zhou, Feibiao ;
Mercado, Ronaldo ;
Kenyery, Frank ;
Shen, Joseph ;
Morles, Martin ;
Garcia, Carlos ;
Langansan, Robbie .
FLOW MEASUREMENT AND INSTRUMENTATION, 2006, 17 (06) :399-413
[9]   A self-validating digital Coriolis mass-flow meter: an overview [J].
Henry, MP ;
Clarke, DW ;
Archer, N ;
Bowles, J ;
Leahy, MJ ;
Liu, RP ;
Vignos, J ;
Zhou, FB .
CONTROL ENGINEERING PRACTICE, 2000, 8 (05) :487-506
[10]   Development of Coriolis mass flowmeter with digital drive and signal processing technology [J].
Hou, Qi-Li ;
Xu, Ke-Jun ;
Fang, Min ;
Liu, Cui ;
Xiong, Wen-Jun .
ISA TRANSACTIONS, 2013, 52 (05) :692-700