A New Computational Fluid Dynamics Method for In-Depth Investigation of Flow Dynamics in Roller Pump Systems

被引:19
作者
Zhou, Xiaoming [1 ]
Liang, Xin M. [2 ]
Zhao, Gang [2 ]
Su, Youchao [1 ]
Wang, Yang [1 ]
机构
[1] Univ Elect Sci & Technol China, Sch Mech Elect & Ind Engn, Chengdu 611731, Sichuan, Peoples R China
[2] Univ Sci & Technol China, Dept Elect Sci & Technol, Ctr Biomed Engn, Hefei 230026, Anhui, Peoples R China
基金
中国国家自然科学基金;
关键词
Roller pump; Computational fluid dynamics; Wall shear stress; Cell damage; CARDIOPULMONARY BYPASS; BLOOD DAMAGE; CENTRIFUGAL; HEMOLYSIS; PREDICTION; SURGERY; SUPPORT; TRAUMA;
D O I
10.1111/aor.12319
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Roller pumps are commonly used in circulatory assist devices to deliver blood, but the inherent high mechanical stresses (especially wall shear stress) may cause considerable damage to cells. Conventional experimental approaches to evaluate and reduce device-induced cell damage require considerable effort and resources. In this work, we describe the use of a new computational fluid dynamics method to more effectively study roller pump systems. A generalized parametric model for the fluid field in a typical roller pump system is presented first, and analytical formulations of the moving boundary are then derived. Based on the model and formulations, the dynamic geometry and mesh of the fluid field can be updated automatically according to the time-dependent roller positions. The described method successfully simulated the pulsing flow generated by the pump, offering a convenient way to visualize the inherent flow pattern and to assess shear-induced cell damage. Moreover, the highly reconfigurable model and the semiautomated simulation process extend the usefulness of the presented method to a wider range of applications. Comparison studies were conducted, and valuable indications about the detailed effects of structural parameters and operational conditions on the produced wall shear stress were obtained. Given the good consistency between the simulated results and the existing experimental data, the presented method displays promising potential to more effectively guide the development of improved roller pump systems which produce less mechanical damage to cells.
引用
收藏
页码:E106 / E117
页数:12
相关论文
共 26 条
[1]   Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics [J].
Apel, J ;
Paul, R ;
Klaus, S ;
Siess, T ;
Reul, H .
ARTIFICIAL ORGANS, 2001, 25 (05) :341-347
[2]   A review of computational fluid dynamics analysis of blood pumps [J].
Behbahani, M. ;
Behr, M. ;
Hormes, M. ;
Steinseifer, U. ;
Arora, D. ;
Coronado, O. ;
Pasquali, M. .
EUROPEAN JOURNAL OF APPLIED MATHEMATICS, 2009, 20 :363-397
[3]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[4]   MODEL FOR A GENERAL MECHANICAL BLOOD DAMAGE PREDICTION [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :583-589
[5]   Hemolysis During Cardiac Extracorporeal Membrane Oxygenation: A Case-Control Comparison of Roller Pumps and Centrifugal Pumps in a Pediatric Population [J].
Byrnes, Jonathan ;
McKamie, Wes ;
Swearingen, Christopher ;
Prodhan, Parthak ;
Bhutta, Adnan ;
Jaquiss, Robert ;
Imamura, Michiaki ;
Fiser, Richard .
ASAIO JOURNAL, 2011, 57 (05) :456-461
[6]   FURTHER DECREASE IN SUBCLINICAL HEMOLYSIS UTILIZING 12.7-MM TUBING IN THE ARTERIAL ROLLER HEAD [J].
CRANE, KA ;
BROWN, D ;
ANDERSON, R ;
BEGEIMAN, KM .
ANNALS OF THORACIC SURGERY, 1983, 35 (04) :463-465
[7]  
De Soma F. M. J. J., 2008, CARDIOPULMONARY BYPA, p[35, 43]
[8]  
Fung Y, 2013, Biomechanics: mechanical properties of living tissues
[9]   Axial Magnetic Bearing Development for the BiVACOR Rotary BiVAD/TAH [J].
Greatrex, Nicholas A. ;
Timms, Daniel L. ;
Kurita, Nobuyuki ;
Palmer, Edward W. ;
Masuzawa, Toru .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (03) :714-721
[10]   Prospective Randomized Clinical Study of Arterial Pumps Used for Routine on Pump Coronary Bypass Grafting [J].
Keyser, Andreas ;
Hilker, Michael K. ;
Diez, Claudius ;
Philipp, Alois ;
Foltan, Maik ;
Schmid, Christof .
ARTIFICIAL ORGANS, 2011, 35 (05) :534-542