Calibration of cylindrical detectors using a simplified theoretical approach

被引:35
作者
Abbas, Mahmoud I. [1 ]
Nafee, Sherif [1 ]
Selim, Younis S. [1 ]
机构
[1] Univ Alexandria, Fac Sci, Phys Dept, Alexandria 21121, Egypt
关键词
gamma spectroscopy; HPGe detector; NaI(Tl) detector; photon path length;
D O I
10.1016/j.apradiso.2006.05.005
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The calibration of cylindrical detectors using different types of radioactive sources is a matter of routine. The most accurate method, that of experiment, is limited by several factors when the energy interval is, broad, requiring a relatively large number of primary standards, implying considerable investment of money and time. Several other techniques can be used instead, including Monte Carlo simulations and semi-empirical methods. Calculations based on the first technique require good definition of the geometry and materials, including the dead layer and window thickness together with an accurate set of cross-sections. The second technique requires two different types of experimental input, the first being from use of sources emitting cascade 7 rays and the second from use of sources emitting isolated 7 rays in order to cover the wide energy range and provide coincidence-summing corrections, respectively. Here, we introduce a new theoretical approach based on the Direct Statistical method proposed by Selim and Abbas to calculate the total and the full-energy peak (photopeak) efficiencies for both point and thin circular disk sources for scintillation and semiconductor detectors. The present method combines calculation of the average path length covered by the photon inside the detector active volume and the geometrical solid angle Omega, to obtain a simple formula for the different efficiencies. Results from the present model were tested against data sets obtained with previous treatments in order to underline how simple and fast our calculations are. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1057 / 1064
页数:8
相关论文
共 23 条
[1]   HPGe detector photopeak efficiency calculation including self-absorption and coincidence corrections for cylindrical sources using compact analytical expressions [J].
Abbas, MI ;
Selim, YS ;
Bassiouni, M .
RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) :429-431
[2]   HPGe detector photopeak efficiency calculation including self-absorption and coincidence corrections for Marinelli beaker sources using compact analytical expressions [J].
Abbas, MI .
APPLIED RADIATION AND ISOTOPES, 2001, 54 (05) :761-768
[3]  
ABBAS MI, 2005, ARAB J NUCL SCI APPL, V38, P351
[4]   MONTE-CARLO CALCULATION OF EFFICIENCIES OF RIGHT-CIRCULAR CYLINDRICAL NAI DETECTORS FOR ARBITRARILY LOCATED POINT SOURCES [J].
BEAM, GB ;
WIELOPOLSKI, L ;
GARDNER, RP ;
VERGHESE, K .
NUCLEAR INSTRUMENTS & METHODS, 1978, 154 (03) :501-508
[6]   DETERMINATION OF FULL ENERGY PEAK EFFICIENCY OF NAI(TL) DETECTORS [J].
CESANA, A ;
TERRANI, M .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1978, 29 (07) :427-429
[7]   Cascade summing corrections for HPGe spectrometers by the Monte Carlo method [J].
Dias, MS ;
Takeda, MN ;
Koskinas, MF .
APPLIED RADIATION AND ISOTOPES, 2002, 56 (1-2) :105-109
[8]  
ELGHARABAWY HA, 2005, IN PRESS NUCL INSTRU
[9]  
Hubbell J. H., 1995, 5632 NISTIR
[10]   RELATIVE PHOTOPEAK EFFICIENCIES AND PHOTOFRACTIONS OF A 2'' BY 2'' CSI(TL) CRYSTAL [J].
IRFAN, M ;
PRASAD, RDG .
NUCLEAR INSTRUMENTS & METHODS, 1970, 88 (02) :165-&