Compatibility of Nonoriginal Abutments With Implants: Evaluation of Microgap at the Implant-Abutment Interface, With Original and Nonoriginal Abutments

被引:358
作者
Duraisamy, Revathi [1 ]
Krishnan, Chitra Shankar [2 ]
Ramasubramanian, Hariharan [2 ]
Sampathkumar, Jayakrishnakumar [2 ]
Mariappan, Saravanakumar [2 ]
Sivaprakasam, Azhagarasan Navarasampatti [2 ]
机构
[1] Saveetha Univ, Saveetha Inst Med & Tech Sci, Saveetha Dent Coll & Hosp, Dept Prosthodont, 162 Poonamalle High Rd, Chennai 600077, Tamil Nadu, India
[2] Ragas Dent Coll & Hosp, Dept Prosthodont, Chennai, Tamil Nadu, India
关键词
compatible abutment; implant-abutment junction; internal connection; misfit; scanning electron microscopy; FIT; ACCURACY; MISFIT; COMPONENTS;
D O I
10.1097/ID.0000000000000885
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to evaluate the fit of nonoriginal abutments to implants at the implant abutment junction. Materials and Methods: Twenty titanium implants from a single manufacturer were randomly divided into 2 groups of ten each. Ten titanium premachined original abutments (group I) and ten titanium premachined nonoriginal abutments (from different manufacturerd-group II) were connected to the implants with the recommended manufacturer torque level and then embedded into autopolymerizing clear acrylic resin blocks. After overnight curing, these blocks were vertically sectioned using water jet sectioning machine and evaluated under scanning electron microscope following the sequential cleaning procedures. The microgap at the implant-abutment interface for all the samples was measured using pixel counting software and subjected to statistical analysis using nonparametric Mann-Whitney U test. Results: The mean microgap at the implant-abutment interface at the external, middle, and internal points was 1.597, 1.399, and 1.831 mm, respectively, for group I and 2.395, 2.488, and 3.339 mm, respectively, for group II samples. Nonparametric Mann-Whitney U test showed statistically significant difference between 2 groups at the midpoint for the nonoriginal abutments compared with the original ones. Conclusion: Within the limitations of the study, the mean microgap at the implant-abutment interface at the platform level at the external, middle, and internal points for both original abutments and nonoriginal abutments was found to be within clinically acceptable limits.
引用
收藏
页码:289 / 295
页数:7
相关论文
共 37 条
[1]   Comparison of Fit Accuracy between Procera® Custom Abutments and Three Implant Systems [J].
Alves da Cunha, Tiago de Morais ;
Correia de Araujo, Roberto Paulo ;
Barbosa da Rocha, Paulo Vicente ;
Pazos Amoedo, Rosa Maria .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2012, 14 (06) :890-895
[2]  
[Anonymous], BRAZ DENT SCI
[3]  
[Anonymous], 2014, J DENT BIOMECH
[4]  
Apicella D, 2010, ANN STOMATOL, V1, P9
[5]  
Baixe S, 2010, INT J ORAL MAX IMPL, V25, P455
[6]  
Baldassarri M, 2012, INT J ORAL MAX IMPL, V27, P537
[7]  
Barbosa Gustavo Augusto Seabra, 2008, Braz. Dent. J., V19, P358, DOI 10.1590/S0103-64402008000400013
[8]  
Baumgarten H, 2014, J IMPLANT RECONS DEN, V5
[9]  
Binon PP, 2000, INT J ORAL MAX IMPL, V15, P76
[10]   The fit of cast and premachined implant abutments [J].
Byrne, D ;
Houston, F ;
Cleary, R ;
Claffey, N .
JOURNAL OF PROSTHETIC DENTISTRY, 1998, 80 (02) :184-192