Geometric Aspects of Optimizing the Acceptance of a Static Mass Analyzer

被引:3
|
作者
Sachenko, V. D. [1 ]
Antonov, A. S. [2 ]
Gall, L. N. [1 ]
Berdnikov, A. S. [1 ]
机构
[1] Russian Acad Sci, Inst Analyt Instrumentat, St Petersburg 190103, Russia
[2] Russian Acad Sci, Ioffe Inst, St Petersburg 194021, Russia
关键词
static mass spectrometers; resolution; sensitivity; transmission; acceptance; optimization; TOROIDAL ELECTRON SPECTROMETER; RESOLUTION; DESIGN;
D O I
10.1134/S1061934820140129
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The axial aberration of the ion-optical system of a static mass analyzer is one of the main factors limiting the transmission and sensitivity of the instrument. We described an algorithm for selecting the optimal collimating system for a mass analyzer, which makes it possible to decrease this aberration effectively, with minimal loss in the intensity of the mass spectral lines. The algorithm is based on the representation of the considered aberration in the phase space as an ellipse and the optimal approximation of this aberration ellipse by a polygon with a given number of vertices. In the absence of constraints, the solution of the corresponding optimization problems is in phase polygons generated by regular polygons inscribed in a unit circle or circumscribed around a unit circle with the same number of vertices. After that, the unit circle along with the polygons is "stretched" to the size of the specified aberration ellipse along its main axes. Additional freedom of the rotation of the original polygon around the center of the unit circle can be used to position auxiliary slits of the collimating system more conveniently.
引用
收藏
页码:1781 / 1789
页数:9
相关论文
共 50 条
  • [1] Geometric Aspects of Optimizing the Acceptance of a Static Mass Analyzer
    V. D. Sachenko
    A. S. Antonov
    L. N. Gall
    A. S. Berdnikov
    Journal of Analytical Chemistry, 2020, 75 : 1781 - 1789
  • [2] Optimization of Vertical Acceptance of the Static Mass Analyzer
    Sachenko, V. D.
    Antonov, A. S.
    Gall, L. N.
    Berdnikov, A. S.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2020, 75 (13) : 1660 - 1664
  • [3] Optimization of Vertical Acceptance of the Static Mass Analyzer
    V. D. Sachenko
    A. S. Antonov
    L. N. Gall
    A. S. Berdnikov
    Journal of Analytical Chemistry, 2020, 75 : 1660 - 1664
  • [4] A Mass Analyzer Based on a Magnetic Mirror
    Sachenko, V. D.
    Antonov, A. S.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2024, 79 (13) : 1985 - 1991
  • [5] Time-of-flight Mass Analyzer
    Volny, Michael
    CHEMICKE LISTY, 2020, 114 (02): : 113 - 118
  • [6] THE THERAPEUTIC ASPECTS OF ACCEPTANCE
    WEISSMAN, J
    APPLETON, C
    PERSPECTIVES IN PSYCHIATRIC CARE, 1995, 31 (01) : 19 - 23
  • [7] A simplified toroidal ion trap mass analyzer
    Taylor, Nicholas
    Austin, Daniel E.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2012, 321 : 25 - 32
  • [8] On Some Improvements of Ion Microprobe Mass Analyzer
    Kuzema, O. S.
    Kuzema, P. O.
    JOURNAL OF NANO- AND ELECTRONIC PHYSICS, 2013, 5 (03)
  • [9] Optimizing the Ion-Optical Design of a Portable Static Mass Spectrometer for Medical Diagnostics
    Antonov, A. S.
    Aruev, N. N.
    Kuleshov, D. O.
    Sachenko, V. D.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2023, 78 (13) : 1868 - 1873
  • [10] Identification of Geometric Parameters Influencing the Impedance Sensitivity of Piezoelectric Mass Sensors
    Zhao, Jian
    Huang, Yu
    Wen, Xin
    Zhang, Shumin
    Liu, Pengbo
    IEEE SENSORS JOURNAL, 2020, 20 (24) : 14740 - 14746