A Novel Method for Measuring Polyethylene Hip Liner Wear Using a Coordinate Measuring Machine and a Dual Stylus Probe

被引:0
|
作者
Cunkelman, Benjamin P. [1 ]
Jang, Byoungwook [1 ]
Van Citters, Douglas W. [1 ]
Collier, John P. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, 14 Engn Dr, Hanover, NH 03755 USA
关键词
CROSS-LINKED POLYETHYLENE; CONVENTIONAL POLYETHYLENE; CLINICAL-PERFORMANCE; HEAD PENETRATION; DOUBLE-BLIND; ARTHROPLASTY;
D O I
10.1115/1.4032107
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ex vivo high-resolution measurement of highly crosslinked (HXL) polyethylene hip liner wear is necessary to characterize the in vivo performance of these polymers that exhibit increased wear resistance. Current studies focus on using a coordinate measuring machine (CMM) to acquire data representing the bearing surface( s) of HXL hip liners and use this data to determine linear and volumetric wear. However, these current techniques are subject to error in both data acquisition and data analysis. The purpose of this study was to identify these sources of error and present a novel method for HXL wear measurement that minimizes these contributions to error: our novel methods use a CMM to measure both the articular and backside surfaces of HXL hip liners for subsequent data analysis in Geomagic Control and MATLAB. Our method involves a vertical orientation of the hip liner to enable one CMM scan of both sides of the hip liner. This method minimizes identified sources of error and proves to be an effective approach for data acquisition of HXL hip liner wear. We also find that our data analysis technique of calculating changes in wall thicknesses is effective in accounting for errors associated with data analysis. Validation of this technique occurred via measurement of two never-implanted HXL hip liners of different sizes (28 mm and 32 mm). In comparing the 32 mm hip liner to its corresponding computer-aided design (CAD) model, we found that our data acquisition technique led to a 0.0019 mm discrepancy between the scanned liner and its CAD model in measured thickness at the pole. We calculated 0.0588 mm and 0.0800 of linear wear for the 28 mm and 32 mm hip liners, respectively, based on our data analysis algorithm. We hypothesize that these reported linear wear values of the never-implanted hip liners are due to machining tolerances of the hip liners themselves.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] PERFORMANCE OF A COORDINATE MEASURING MACHINE WITH A LASER PROBE
    LYNCH, E
    TAN, R
    TAY, WM
    SMALES, FC
    ALLRED, H
    JOURNAL OF DENTAL RESEARCH, 1990, 69 : 344 - 344
  • [2] Probe orientation for coordinate measuring machine systems using design models
    Jackman, J
    Park, DK
    ROBOTICS AND COMPUTER-INTEGRATED MANUFACTURING, 1998, 14 (03) : 229 - 236
  • [3] COORDINATE MEASURING MACHINE VARIATIONS FOR DIFFERENT PROBE CONFIGURATIONS
    Tulcan, Aurel
    Cioana, Cristian
    Stan, Daniel
    Cosma, Cristian
    MODTECH 2012: NEW FACE OF T M C R, VOLS I AND II, 2012, : 989 - 992
  • [4] Validation of an Optical Coordinate Measuring Machine for the Measurement of Wear at the Taper Interface in Total Hip Replacement
    Cook, R. B.
    Maul, C.
    Strickland, A. M.
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 362 - 378
  • [5] Development of A Novel Virtual Coordinate Measuring Machine
    Hu, Yang
    Yang, Qingping
    Wei, Peng
    I2MTC: 2009 IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3, 2009, : 220 - 223
  • [6] Minimum radius method for measuring curved edges on a coordinate measuring machine
    Grigoriev, Sergey N.
    Zakharov, Oleg, V
    Wang, Jian
    Masterenko, Dmitriy A.
    Li, Chen
    Ivanova, Tatyana N.
    MEASUREMENT, 2025, 250
  • [7] Measurement Control and Radius Compensation of the Probe for Coordinate Measuring Machine
    Wang, Hong-Min
    Shi, Pei-Lin
    ICMS2010: PROCEEDINGS OF THE THIRD INTERNATIONAL CONFERENCE ON MODELLING AND SIMULATION ICMS2010, VOL 3: MODELLING AND SIMULATION IN INDUSTRIAL APPLICATION, 2010, : 115 - 118
  • [8] Probe Error Analysis of Articulated Arm Coordinate Measuring Machine
    Cai Zeliang
    Luo Zai
    Liu Hui
    TENTH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2019, 11053
  • [9] Probe Parameters Calibration for Articulated Arm Coordinate Measuring Machine
    Cheng Wentao
    Fei Yetai
    Yu Liandong
    Yang Ruichang
    6TH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2010, 7544
  • [10] Coordinate measuring machine variations for selected probe head configurations
    Berisso, Kevin
    Ollison, Troy
    Journal of Industrial Technology, 2010, 26 (01): : 1 - 11