Determining the optimal dosimetric leaf gap setting for rounded leaf-end multileaf collimator systems by simple test fields

被引:35
作者
Yao, Weiguang [1 ]
Farr, Jonathan B. [1 ]
机构
[1] St Jude Childrens Res Hosp, Dept Radiat Oncol, Memphis, TN 38105 USA
关键词
Optimization of dosimetric leaf gap; tongue and groove effect; individual QA; IMRT DELIVERY; MLC; CALIBRATION; UNCERTAINTIES; ACCURACY;
D O I
10.1120/jacmp.v16i4.5321
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Individual QA for IMRT/VMAT plans is required by protocols. Sometimes plans cannot pass the institute's QA criteria. For the Eclipse treatment planning system (TPS) with rounded leaf-end multileaf collimator (MLC), one practical way to improve the agreement of planned and delivered doses is to tune the value of dosimetric leaf gap (DLG) in the TPS from the measured DLG. We propose that this step may be necessary due to the complexity of the MLC system, including dosimetry of small fields and the tongue-and-groove (T&G) effects, and report our use of test fields to obtain linac-specific optimal DLGs in TPSs. More than 20 original patient plans were reoptimized with the linac-specific optimal DLG value. We examined the distribution of gaps and T&G extensions in typical patient plans and the effect of using the optimal DLG on the distribution. The QA pass rate of patient plans using the optimal DLG was investigated. The dose-volume histograms (DVHs) of targets and organs at risk were checked. We tested three MLC systems (Varian millennium 120 MLC, high-definition 120 MLC, and Siemens 160 MLC) installed in four Varian linear accelerators (linacs) (TrueBEAM STx, Trilogy, Clinac 2300 iX, and Clinac 21 EX) and 1 Siemens linac (Artiste). With an optimal DLG, the individual QA for all those patient plans passed the institute's criteria (95% in DTA test or gamma test with 3%/3 mm/10%), even though most of these plans had failed to pass QA when using original DLGs optimized from typical patient plans or from the optimization process (automodeler) of Pinnacle TPS. Using either our optimal DLG or one optimized from typical patient plans or from the Pinnacle optimization process yielded similar DVHs.
引用
收藏
页码:65 / 77
页数:13
相关论文
共 26 条
[1]   A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy [J].
Arnfield, MR ;
Siebers, JV ;
Kim, JO ;
Wu, QW ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (10) :2231-2241
[2]   Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones [J].
Bassinet, C. ;
Huet, C. ;
Derreumaux, S. ;
Brunet, G. ;
Chea, M. ;
Baumann, M. ;
Lacornerie, T. ;
Gaudaire-Josset, S. ;
Trompier, F. ;
Roch, P. ;
Boisserie, G. ;
Clairand, I. .
MEDICAL PHYSICS, 2013, 40 (07)
[3]   Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT? [J].
Betzel, Gregory T. ;
Yi, Byong Yong ;
Niu, Ying ;
Yu, Cedric X. .
MEDICAL PHYSICS, 2012, 39 (10) :5882-5890
[4]   Lateral loss and dose discrepancies of multileaf collimator segments in intensity modulated radiation therapy [J].
Cheng, CW ;
Das, IJ ;
Huq, MS .
MEDICAL PHYSICS, 2003, 30 (11) :2959-2968
[5]   Small fields:: Nonequilibrium radiation dosimetry [J].
Das, Indra J. ;
Ding, George X. ;
Ahnesjo, Anders .
MEDICAL PHYSICS, 2008, 35 (01) :206-215
[6]   The MLC tongue-and-groove effect on IMRT dose distributions [J].
Deng, J ;
Pawlicki, T ;
Chen, Y ;
Li, JS ;
Jiang, SB ;
Ma, CM .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04) :1039-1060
[7]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[8]   Calibration and quality assurance for rounded leaf-end MLC systems [J].
Graves, MN ;
Thompson, AV ;
Martel, MK ;
McShan, DL ;
Fraass, BA .
MEDICAL PHYSICS, 2001, 28 (11) :2227-2233
[9]   Dose uncertainty due to aperture effects in dynamic fields [J].
Higgins, P. D. ;
Alaei, P. .
MEDICAL PHYSICS, 2006, 33 (07) :2418-2425
[10]   Verification of dosimetric accuracy on the TrueBeam STx: Rounded leaf effect of the high definition MLC [J].
Kielar, Kayla N. ;
Mok, Ed ;
Hsu, Annie ;
Wang, Lei ;
Luxton, Gary .
MEDICAL PHYSICS, 2012, 39 (10) :6360-6371