Development of a centrifugal pump with thick blades

被引:1
|
作者
Kim, WG
Chung, CH
Yang, WS
Park, YN
Kim, HI
Kim, HC
Kang, SH
机构
[1] Seoul Natl Univ, Coll Med, Dept Thorac & Cardiovasc Surg, Seoul 110744, South Korea
[2] Seoul Natl Univ, Heart Res Inst, Seoul 110744, South Korea
[3] Seoul Natl Univ, Coll Med, Dept Mech Engn, Seoul 110744, South Korea
[4] Seoul Natl Univ, Coll Engn, Dept Biomed Engn, Seoul 110744, South Korea
关键词
centrifugal pump; thick blades; tip clearance;
D O I
10.1046/j.1525-1594.2000.06482-3.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have developed a centrifugal blood pump with thick impeller blades (60% of pitch) to obtain a small tip clearance. An unshrouded impeller with 6 backward curved thick blades was used to reduce the dead zone between the shroud and upper casing, A streamline angle in volute was uniform in circumferential direction by continuity and angular momentum conservation. To prove the effectiveness of small tip clearance, performance and hemolysis tests were conducted on pumps with a tip clearance of 05, 1.5, and 2.0 mm at exit with the blade thickness of 60% of pitch, and with that of 1.0, 2.0, and 2.5 mm at exit with the thickness of 40% of pitch. The results showed that the smaller the tip clearance, the better the hydrodynamic and hemolytic performance. The best result was seen in the pump with tip clearance of 0.5 mm with a blade thickness of 60% of pitch. These results suggest that a centrifugal pump with thick blades and a small tip clearance can be a promising alternative as a cardiopulmonary bypass pump.
引用
收藏
页码:160 / 164
页数:5
相关论文
共 50 条
  • [31] Performance and internal flow condition of mini centrifugal pump with splitter blades
    Shigemitsu, T.
    Fukutomi, J.
    Kaji, K.
    Wada, T.
    26TH IAHR SYMPOSIUM ON HYDRAULIC MACHINERY AND SYSTEMS, PTS 1-7, 2013, 15
  • [32] The flow field in a centrifugal pump with a large tongue gap and back blades
    Jiancheng Cai
    Jie Pan
    Andrew Guzzomi
    Journal of Mechanical Science and Technology, 2014, 28 : 4455 - 4464
  • [33] Influence of Splitter Blades on the Cavitation Performance of a Double Suction Centrifugal Pump
    Yang, Wei
    Xiao, Ruofu
    Wang, Fujun
    Wu, Yulin
    ADVANCES IN MECHANICAL ENGINEERING, 2014,
  • [34] Investigation of Self-Priming Process of a Centrifugal Pump with Double Blades
    QIANHeng
    MOUJiegang
    RENYun
    ZHUZhibing
    LIUNuojia
    ZHENGShuihua
    WUDenghao
    JournalofThermalScience, 2021, 30 (03) : 849 - 858
  • [35] Investigation of Self-Priming Process of a Centrifugal Pump with Double Blades
    Heng Qian
    Jiegang Mou
    Yun Ren
    Zhibing Zhu
    Nuojia Liu
    Shuihua Zheng
    Denghao Wu
    Journal of Thermal Science, 2021, 30 : 849 - 858
  • [36] Performance and internal flow condition of mini centrifugal pump with splitter blades
    Shigemitsu, Toru
    Fukutomi, Junichiro
    Kaji, Kensuke
    Wada, Takashi
    International Journal of Fluid Machinery and Systems, 2013, 6 (01) : 11 - 17
  • [37] Development of a novel centrifugal pump: Magnetic rotary pump
    Naganuma, S
    Yambe, T
    Sonobe, T
    Kobayashi, S
    Nitta, S
    ARTIFICIAL ORGANS, 1997, 21 (07) : 746 - 750
  • [38] DEVELOPMENT OF THE CHANNEL TYPE CENTRIFUGAL PUMP
    Kuleshov, A. P.
    Itkin, G. P.
    Baybikov, A. S.
    VESTNIK TRANSPLANTOLOGII I ISKUSSTVENNYH ORGANOV, 2018, 20 (03): : 32 - 39
  • [39] TRENDS IN DEVELOPMENT OF CENTRIFUGAL PUMP CONSTRUCTION
    PEKRUN, M
    PUMPS AND THEIR APPLICATIONS, 1971, (57): : 255 - &
  • [40] Development of Liquid Nitrogen Centrifugal Pump
    Abe, M.
    Takayama, T.
    Sagiyama, R.
    Tsuchiya, H.
    Torii, Y.
    Nakamura, M.
    Hoshino, Y.
    Odashima, Y.
    25TH INTERNATIONAL CONFERENCE ON LOW TEMPERATURE PHYSICS (LT25), PART 1: CRYOGENIC TECHNOLOGIES AND APPLICATIONS, 2009, 150