共 50 条
High-Throughput Pesticide Residue Quantitative Analysis Achieved by Tandem Mass Spectrometry with Automated Flow Injection
被引:48
作者:
Nanita, Sergio C.
[1
]
Pentz, Anne M.
[1
]
Bramble, Frederick Q.
[1
]
机构:
[1] DuPont Crop Protect, Stine Haskell Res Ctr, Newark, DE 19714 USA
关键词:
DESORPTION ELECTROSPRAY-IONIZATION;
TIME-OF-FLIGHT;
LIQUID-CHROMATOGRAPHY;
ATMOSPHERIC-PRESSURE;
VEGETABLES;
SOLIDS;
FRUITS;
WATER;
MS;
D O I:
10.1021/ac900226w
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
The use of automated flow injection with MS/MS detection for fast quantitation of agrochemicals in food and water samples was demonstrated in this study. Active ingredients from the sulfonylurea herbicide and carbamate insecticide classes were selected as model systems. Samples were prepared using typical procedures from residue methods, placed in an autosampler, and injected directly into a triple quadrupole instrument without chromatographic separation. The technique allows data acquisition in 15 s per injection, with samples being injected every 65 s, representing a significant improvement from the 15-30 min needed in typical HPLC/MS/MS methods. The availability of HPLC systems is an advantage since they can be used in flow-injection mode (bypassing the column compartment). Adequate accuracy, linearity, and precision (R-2 > 0.99 and RSD < 20%) were obtained using external standards prepared in each control matrix. The limit of quantitation (LOQ) achieved for all analytes was 0.01 mg/kg in food samples and 0.1 ng/mL in water; while limits of detection (LOD) were estimated to be about 0.003 mg/kg and 0.03 ng/mL in food and water, respectively. The advantages and limitations of flow injection MS/MS for ultratrace-level quantitative analysis in complex matrixes are discussed.
引用
收藏
页码:3134 / 3142
页数:9
相关论文
共 50 条