Numerical Evaluation on Irregular Field Generated by Screening Current in High-Field REBCO Coil for Whole-Body MRI

被引:9
|
作者
Nakazono, Kohei [1 ]
Ueda, Hiroshi [2 ]
Ishiyama, Atsushi [1 ]
Noguchi, So [3 ]
Miyazaki, Hiroshi [4 ]
Tosaka, Taizo [4 ]
Kurusu, Tsutomu [4 ]
Nomura, Shunji [4 ]
Urayama, Shinichi [5 ]
Fukuyama, Hidenao [5 ]
机构
[1] Waseda Univ, Tokyo 1698555, Japan
[2] Okayama Univ, Okayama 7008530, Japan
[3] Hokkaido Univ, Sapporo, Hokkaido 0600814, Japan
[4] Toshiba Co Ltd, Yokohama, Kanagawa 2300045, Japan
[5] Kyoto Univ, Kyoto 6068507, Japan
关键词
High-temperature superconducting coil; irregular magnetic field; MRI; screening current; three-dimensional electromagnetic field analysis;
D O I
10.1109/TASC.2016.2645839
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In recent years, high-field magnets wound with RE-BCO tape and their applications to magnetic resonance imaging (MRI) have undergone considerable research and development. MRI coil systems require a highly accurate field in the imaging space. However, in REBCO coils, the irregular magnetic field generated by the screening current is induced by the radial component of the magnetic field in the REBCO tape, and may deteriorate the field quality both spatially and temporally. Therefore, a design process with consideration for the effect of the screening-current field is required for the reduction and correction of this field for a high-accuracy-field REBCO coil. In this study, the screening-current distribution in the REBCO tape and the irregular magnetic field were analyzed by using the simulation we developed for a REBCO model coil system for whole-body MRI. We additionally discuss the effect of the screening-current field in the REBCO coil based on the coil shape and arrangement to use it as feedback to the design stage.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Reduction of Irregular Magnetic Field Generated by Screening Current in REBCO Coil
    Ueda, Hiroshi
    Saito, Jun
    Ariya, Yuta
    Mochida, Ayumu
    Wang, Tao
    Wang, Xudong
    Agatsuma, Koh
    Ishiyama, Atsushi
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2015, 25 (03)
  • [2] Evaluation of Irregular Magnetic Field Generated by Screening Current in REBCO Coils for High Accuracy Field
    Matsumi, Ayako
    Ueda, Hiroshi
    Noguchi, So
    Wang, Tao
    Ishiyama, Atsushi
    Miyazaki, Hiroshi
    Tosaka, Taizo
    Nomura, Shunji
    Kurusu, Tsutomu
    Urayama, Shinichi
    Fukuyama, Hidenao
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (04)
  • [3] Numerical Simulation on Magnetic Field Generated by Screening Current in 10-T-Class REBCO Coil
    Ueda, Hiroshi
    Imaichi, Yohei
    Wang, Tao
    Ishiyama, Atsushi
    Noguchi, So
    Iwai, Sadanori
    Miyazaki, Hiroshi
    Tosaka, Taizo
    Nomura, Shunji
    Kurusu, Tsutomu
    Urayama, Shinichi
    Fukuyama, Hidenao
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (04)
  • [4] Design and Performance Analysis of Gradient Coil for High-Field HTS REBCO MRI Magnet
    Park, Hyunsoo
    Kim, Geonyoung
    Choi, Kibum
    Park, Jeonghwan
    Kim, Jaemin
    Choi, Seyong
    Sim, Kideok
    Hahn, Seungyong
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (05) : 1 - 5
  • [5] Numerical Evaluation of Screening Current-Induced Magnetic Field in No-Insulation REBCO Coil Systems
    Ueda S.
    Ishiyama A.
    Ueda H.
    Noguchi S.
    Ueda, Satomi (atsushi@waseda.jp), 1600, Institute of Electrical and Electronics Engineers Inc. (31):
  • [6] Efficient high-frequency body coil for high-field MRI
    Vaughan, JT
    Adriany, G
    Snyder, CJ
    Tian, J
    Thiel, T
    Bolinger, L
    Liu, H
    DelaBarre, L
    Ugurbil, K
    MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (04) : 851 - 859
  • [7] Optimization in design of actively shielded whole-body open high-field superconducting MRI magnet
    Feng Zhong-Kui
    Hu Ge-Li
    Xu Ying
    Zhu Guang
    Zhou Feng
    Dai Yin-Ming
    Wang Qiu-Liang
    ACTA PHYSICA SINICA, 2013, 62 (23)
  • [8] Screening current effect on the stress and strain distribution in REBCO high-field magnets: experimental verification and numerical analysis
    Yan, Yufan
    Xin, Canjie
    Guan, Mingzhi
    Liu, Huajun
    Tan, Yunfei
    Qu, Timing
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2020, 33 (05):
  • [9] Ultra-high field magnets for whole-body MRI
    Warner, Rory
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2016, 29 (09):
  • [10] Numerical evaluation of the fields induced by body motion in or near high-field MRI scanners
    Crozier, S
    Liu, F
    PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 2005, 87 (2-3): : 267 - 278