Use of Cardiac Output to Improve Measurement of Input Function in Quantitative Dynamic Contrast-Enhanced MRI

被引:32
作者
Zhang, Jeff L. [1 ]
Rusinek, Henry [1 ]
Bokacheva, Louisa [1 ]
Chen, Qun [1 ]
Storey, Pippa [1 ]
Lee, Vivian S. [1 ]
机构
[1] NYU, Sch Med, Dept Radiol, New York, NY 10003 USA
关键词
cardiac output; dynamic contrast-enhanced MRI; arterial input function; FAST GRADIENT-ECHO; MAGNETIC-RESONANCE; IN-VIVO; CARDIOVASCULAR APPLICATIONS; TUMOR ANGIOGENESIS; T-1; MEASUREMENT; RENAL-FUNCTION; PORCINE MODEL; BLOOD-FLOW; DCE-MRI;
D O I
10.1002/jmri.21891
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To validate a new method for converting MR arterial signal intensity versus time curves to arterial input functions (AIFs). Materials and Methods: The method constrains AIF with patient's cardiac output (Q). Monte Carlo simulations of MR renography and tumor perfusion protocols were carried out for comparison with two alternative methods: direct measurement and population-averaged input function. MR renography was performed to assess the method's inter- and intraday reproducibility for renal parameters. Results: In simulations of tumor perfusion, the precision of the parameters (K(trans) and v(e)) computed using the proposed method was improved by at least a factor of three compared to direct measurement. Similar improvements were obtained in simulations of MR renography. Volunteer study for testing interday reproducibility confirmed the improvement of precision in renal parameters when using the proposed method compared to conventional methods. In another patient study (two injections within one session), the proposed method significantly increased the correlation coefficient (R) between GFR of the two exams (0.92 vs. 0.83) compared to direct measurement. Conclusion: A new method significantly improves the precision of dynamic contrast-enhanced (DCE) parameters. The method may be especially useful for analyzing repeated DCE examinations, such as monitoring tumor therapy or angiotensin converting enzyme-inhibitor renography. Key Words: cardiac output; dynamic contrast-enhanced MRI; arterial input function
引用
收藏
页码:656 / 665
页数:10
相关论文
共 54 条
[41]   Phase I trial of the antivascular agent combretastatin A4 phosphate on a 5-day schedule to patients with cancer: Magnetic resonance imaging evidence for altered tumor blood flow [J].
Stevenson, JP ;
Rosen, M ;
Sun, WJ ;
Gallagher, M ;
Haller, DG ;
Vaughn, D ;
Giantonio, B ;
Zimmer, R ;
Petros, WP ;
Stratford, M ;
Chaplin, D ;
Young, SL ;
Schnall, M ;
O'Dwyer, PJ .
JOURNAL OF CLINICAL ONCOLOGY, 2003, 21 (23) :4428-4438
[42]   Cardiovascular applications of magnetic resonance flow and velocity measurements [J].
Szolar, DH ;
Sakuma, H ;
Higgins, CB .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 1996, 6 (01) :78-89
[43]  
Taylor AT, 1998, J NUCL MED, V39, P1297
[44]   Modeling tracer kinetics in dynamic Gd-DTPA MR imaging [J].
Tofts, PS .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (01) :91-101
[45]   Dynamic contrast-enhanced MRI in the diagnosis and management of breast cancer [J].
Turnbull, Lindsay W. .
NMR IN BIOMEDICINE, 2009, 22 (01) :28-39
[46]  
van der Schaaf I, 2006, AM J NEURORADIOL, V27, P46
[47]   T1 measurements incorporating flip angle calibration and correction in vivo [J].
Wang, Jinghua ;
Qiu, Maolin ;
Kim, Hyeonjin ;
Constable, R. Todd .
JOURNAL OF MAGNETIC RESONANCE, 2006, 182 (02) :283-292
[48]   Factors influencing flip angle mapping in MRI:: RF pulse shape, slice-select gradients, off-resonance excitation, and Bo inhomogeneities [J].
Wang, Jinghua ;
Mao, Weihua ;
Qiu, Maolin ;
Smith, Michael B. ;
Constable, R. Todd .
MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (02) :463-468
[49]   Feasibility of using limited-population-based arterial input function for pharmacokinetic modeling of osteosarcoma dynamic contrast-enhanced MRI data [J].
Wang, Ya ;
Huang, Wei ;
Panicek, David A. ;
Schwartz, Lawrence H. ;
Koutcher, Jason A. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) :1183-1189
[50]   Novel method for rapid, simultaneous T1, T2*, and proton density quantification [J].
Warntjes, J. B. M. ;
Dahlqvist, O. ;
Lundberg, P. .