A feasibility study of projection-based energy weighting based on a photon-counting detector in contrast-enhanced digital subtraction mammography: A simulation study

被引:1
作者
Choi, Sunghoon [1 ]
Lee, Seungwan
Choi, Yu-Na
Kim, Hee-Joung
机构
[1] Yonsei Univ, Dept Radiol Sci, Wonju 220710, South Korea
基金
新加坡国家研究基金会;
关键词
Digital subtraction mammography; Projection-based energy-weighting; Energy-resolved photon-counting detector; CNR; COMPUTED-TOMOGRAPHY; MONTE-CARLO; ANODE; CDTE;
D O I
10.3938/jkps.64.1900
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Contrast media, such as iodine and gadolinium, are generally used in digital subtraction mammography to enhance the contrast between target and background materials. In digital subtraction mammography, where one image (with contrast medium) is subtracted from another (anatomical background) to facilitate visualization of the tumor structure, tumors can be more easily distinguished after the injection of a contrast medium. In order to have more an effective method to increase the contrast-to-noise ratio (CNR), we applied a projection-based energy-weighting method. The purpose of this study is to demonstrate the feasibility of using the projection-based energy-weighting method in digital subtraction mammography. Unlike some other previous studies, we applied the projection-based energy-weighting method to more practical mammography conditions by using the Monte Carlo method to simulate four different iodine solutions embedded in a breast phantom comprised of 50% adipose and 50% glandular tissues. We also considered an optimal tube voltage and anode/filter combination in digital iodine contrast media mammography in order to maximize the figure-of-merit (FOM). The simulated source energy was from 20 to 45 keV to prevent electronic noise and include the k-edge energy of iodine (33.2 keV). The results showed that the projection-based energy-weighting improved the CNR by factors of 1.05-1.86 compared to the conventionally integrated images. Consequently, the CNR of digital subtraction mammography images can be improved by using projection-based energy-weighting with photon-counting detectors.
引用
收藏
页码:1900 / 1906
页数:7
相关论文
共 25 条
[1]  
[Anonymous], 2010, CAMBRIDGE DICT STAT
[2]   Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment [J].
Boone, JM .
RADIOLOGY, 1999, 213 (01) :23-37
[3]   Numerical modeling of charge sharing in CdZnTe pixel detectors [J].
Chen, CMH ;
Boggs, SE ;
Bolotnikov, AE ;
Cook, WR ;
Harrison, FA ;
Schindler, SM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) :270-276
[4]   The Effect of Photon Energy Weighting on X-ray Imaging Based on a Photon Counting Detector [J].
Choi, Yu-Na ;
Lee, Seung-Wan ;
Cho, Hyo-Min ;
Ryu, Hyun-Ju ;
Lee, Young-Jin ;
Kim, Hee-Joung .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2011, 59 (05) :3114-3119
[5]  
Cranley K., 1997, 78 I PHYS ENG MED
[6]   Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study [J].
Dance, DR ;
Klang, AT ;
Sandborg, M ;
Skinner, CL ;
Smith, IAC ;
Carlsson, GA .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (874) :1056-1067
[7]  
David Freedman Robert Pisani Roger Purves., 2007, Statistics, VFourth
[8]   INFLUENCE OF ANODE AND FILTER MATERIAL ON IMAGE QUALITY AND GLANDULAR DOSE FOR SCREEN FILM MAMMOGRAPHY [J].
DESPONDS, L ;
DEPEURSINGE, C ;
GRECESCU, M ;
HESSLER, C ;
SAMIRI, A ;
VALLEY, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (09) :1165-1182
[9]   OPTIMIZATION OF SPECTRAL SHAPE IN DIGITAL MAMMOGRAPHY - DEPENDENCE ON ANODE MATERIAL, BREAST THICKNESS, AND LESION TYPE [J].
FAHRIG, R ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (09) :1473-1481
[10]   Contrast-enhanced spectral mammography with a photon-counting detector [J].
Fredenberg, Erik ;
Hemmendorff, Magnus ;
Cederstrom, Bjorn ;
Aslund, Magnus ;
Danielsson, Mats .
MEDICAL PHYSICS, 2010, 37 (05) :2017-2029