Analysis of the modification site in a small molecule-modified peptide by ion trap/time-of-flight hybrid mass spectrometry

被引:7
作者
Mano, Nariyasu
Kamota, Mio
Inohana, Yusuke
Yamaguchi, Shin-ichi
Goto, Junichi
机构
[1] Tohoku Univ Hosp, Dept Pharmaceut Sci, Aoba Ku, Sendai, Miyagi 9808574, Japan
[2] Tohoku Univ, Grad Sch Pharmaceut Sci, Aoba Ku, Sendai, Miyagi 9808578, Japan
[3] Shimadzu Co Ltd, Analyt Instruments Div, Nakagyo Ku, Kyoto 6048511, Japan
关键词
ion trap; mass spectometry; modification site; MSn analysis; small molecule-modified peptide; time-of-flight;
D O I
10.1007/s00216-006-0696-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Ion trap/time-of-flight hybrid mass spectrometers are powerful tools for the detailed structural analysis of modified peptides. We have analyzed Met-Lys-bradykinin modified with deoxycholate at the amino-terminus or the epsilon-amino group as model peptides. These two modified peptides produced fragment ions with the same nominal but different exact masses in tandem mass spectrometry with low-energy collision-induced dissociation. Accurate high-resolution analysis coupled with MS3 allowed us to distinguish the deoxycholate modification sites in the modified peptides.
引用
收藏
页码:682 / 688
页数:7
相关论文
共 11 条
[1]   TANDEM MASS-SPECTROMETRY APPLIED TO PROTEIN-STRUCTURE PROBLEMS [J].
BIEMANN, K .
BIOCHEMICAL SOCIETY TRANSACTIONS, 1989, 17 (01) :237-243
[2]  
DEHOFFMANN E, 2002, MASS SPECTROMETRY PR
[3]  
GOTO J, 1980, CHEM PHARM BULL, V28, P1258
[4]  
GOTO J, 1979, CHEM PHARM BULL, V27, P1402
[5]   A nonenzymatic modification of the amino-terminal domain of histone H3 by bile acid acyl adenylate [J].
Mano, N ;
Kasuga, K ;
Kobayashi, N ;
Goto, J .
JOURNAL OF BIOLOGICAL CHEMISTRY, 2004, 279 (53) :55034-55041
[6]   Analysis of the antigen binding site of anti-deoxycholate monoclonal antibody using a novel affinity labeling reagent, acyl adenylate [J].
Mano, N ;
Nagaya, Y ;
Saito, S ;
Kobayashi, N ;
Goto, J .
BIOCHEMISTRY, 2004, 43 (07) :2041-2048
[7]   DETECTION OF ELECTROSPRAY-IONIZATION USING A QUADRUPOLE ION-TRAP STORAGE REFLECTRON TIME-OF-FLIGHT MASS-SPECTROMETER [J].
MICHAEL, SM ;
CHIEN, BM ;
LUBMAN, DM .
ANALYTICAL CHEMISTRY, 1993, 65 (19) :2614-2620
[8]   Sodium trifluoroacetate as a tune/calibration compound for positive- and negative-ion electrospray ionization mass spectrometry in the mass range of 100-4000 Da [J].
Moini, M ;
Jones, BL ;
Rogers, RM ;
Jiang, LF .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1998, 9 (09) :977-980
[9]   Analysis of neutral oligosaccharides for structural characterization by matrix-assisted laser desorption/ionization quadrupole ion trap time-of-flight mass spectrometry [J].
Ojima, N ;
Masuda, K ;
Tanaka, K ;
Nishimura, O .
JOURNAL OF MASS SPECTROMETRY, 2005, 40 (03) :380-388
[10]   PROPOSAL FOR A COMMON NOMENCLATURE FOR SEQUENCE IONS IN MASS-SPECTRA OF PEPTIDES [J].
ROEPSTORFF, P ;
FOHLMAN, J .
BIOMEDICAL MASS SPECTROMETRY, 1984, 11 (11) :601-601