Physiologic and hematologic concerns of rotary blood pumps: what needs to be improved?

被引:12
作者
Pirbodaghi, Tohid [1 ,2 ,3 ]
Asgari, Siavash [4 ]
Cotter, Chris [3 ]
Bourque, Kevin [3 ]
机构
[1] Univ Bern, Grad Sch Cellular & Biomed Sci, CH-3010 Bern, Switzerland
[2] Univ Bern, ARTORG Cardiovasc Engn, CH-3010 Bern, Switzerland
[3] Thoratec Corp, Burlington, MA USA
[4] Yale Univ, Sch Med, Dept Surg Cardiac Sect, New Haven, CT USA
关键词
Ventricular assist devices; Rotary blood pumps; Heart failure; Continuous flow; Pulsatile flow; VENTRICULAR ASSIST DEVICE; VON-WILLEBRAND-SYNDROME; END-ORGAN FUNCTION; MECHANICAL CIRCULATORY SUPPORT; CONTINUOUS-FLOW; SHEAR-STRESS; HEART; PULSATILE; NONPULSATILE; PLATELET;
D O I
10.1007/s10741-013-9389-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Over the past few decades, advances in ventricular assist device (VAD) technology have provided a promising therapeutic strategy to treat heart failure patients. Despite the improved performance and encouraging clinical outcomes of the new generation of VADs based on rotary blood pumps (RBPs), their physiologic and hematologic effects are controversial. Currently, clinically available RBPs run at constant speed, which results in limited control over cardiac workload and introduces blood flow with reduced pulsatility into the circulation. In this review, we first provide an update on the new challenges of mechanical circulatory support using rotary pumps including blood trauma, increased non-surgical bleeding rate, limited cardiac unloading, vascular malformations, end-organ function, and aortic valve insufficiency. Since the non-physiologic flow characteristic of these devices is one of the main subjects of scientific debate in the literature, we next emphasize the latest research regarding the development of a pulsatile RBP. Finally, we offer an outlook for future research in the field.
引用
收藏
页码:259 / 266
页数:8
相关论文
共 69 条
  • [1] Bridge to Transplantation: Current Outcomes
    Aggarwal, Sanjeev
    Pagani, Francis D.
    [J]. JOURNAL OF CARDIAC SURGERY, 2010, 25 (04) : 455 - 461
  • [2] Altemose GT, 1997, J HEART LUNG TRANSPL, V16, P765
  • [3] Peripheral vascular reactivity in patients with pulsatile vs axial flow left ventricular assist device support
    Amir, O
    Radovancevic, B
    Delgado, RM
    Kar, B
    Radovancevic, R
    Henderson, M
    Cohn, WE
    Smart, FW
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (04) : 391 - 394
  • [4] A novel counterpulsation mode of rotary left ventricular assist devices can enhance myocardial perfusion
    Ando, Masahiko
    Takewa, Yoshiaki
    Nishimura, Takashi
    Yamazaki, Kenji
    Kyo, Shunei
    Ono, Minoru
    Tsukiya, Tomonori
    Mizuno, Toshihide
    Taenaka, Yoshiyuki
    Tatsumi, Eisuke
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2011, 14 (03) : 185 - 191
  • [5] [Anonymous], J HEART LUNG TRANSPL, V27, P1096
  • [6] [Anonymous], J HEART LUNG TRANSPL, V15, P840
  • [7] Pulsatile operation of a centrifugal ventricular assist device with magnetic bearings
    Bearnson, GB
    Olsen, DB
    Khanwilkar, PS
    Long, JW
    Allaire, PE
    Maslen, EH
    [J]. ASAIO JOURNAL, 1996, 42 (05) : M620 - M624
  • [8] Left ventricular assist device and drug therapy for the reversal of heart failure
    Birks, Emma J.
    Tansley, Patrick D.
    Hardy, James
    George, Robert S.
    Bowles, Christopher T.
    Burke, Margaret
    Banner, Nicholas R.
    Khaghani, Asghar
    Yacoub, Magdi H.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (18) : 1873 - 1884
  • [9] In vivo assessment of a rotary left ventricular assist device-induced artificial pulse in the proximal and distal aorta
    Bourque, Kevin
    Dague, Charles
    Farrar, David
    Harms, Kelly
    Tamez, Dan
    Cohn, William
    Tuzun, Egemeun
    Poirier, Victor
    Frazier, O. Howard
    [J]. ARTIFICIAL ORGANS, 2006, 30 (08) : 638 - 642
  • [10] Current state of ventricular assist devices
    Caccamo M.
    Eckman P.
    John R.
    [J]. Current Heart Failure Reports, 2011, 8 (2) : 91 - 98