Yet more evidence that myelin protons can be directly imaged with UTE sequences on a clinical 3T scanner: Bicomponent <mml:msubsup>T2*</mml:msubsup> analysis of native and deuterated ovine brain specimens

被引:28
作者
Fan, Shu-Juan [1 ]
Ma, Yajun [1 ]
Zhu, Yanchun [1 ]
Searleman, Adam [1 ]
Szeverenyi, Nikolaus M. [1 ]
Bydder, Graeme M. [1 ]
Du, Jiang [1 ]
机构
[1] Univ Calif San Diego, Dept Radiol, 200 West Arbor Dr, San Diego, CA 92103 USA
基金
美国国家卫生研究院;
关键词
T-2*; bicomponent; myelin; white matter; UTE; inversion recovery; WHITE-MATTER; MAGNETIC-RESONANCE; LONG-T-2; SUPPRESSION; MULTIPLE-SCLEROSIS; ALZHEIMERS-DISEASE; PULSE SEQUENCES; TIME; MRI; WATER; T-2;
D O I
10.1002/mrm.27052
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeUTE sequences with a minimal nominal TE of 8 mu s have shown promise for direct imaging of myelin protons (T-2,<1 ms). However, there is still debate about the efficiency of 2D slice-selective UTE sequences in exciting myelin protons because the half excitation pulses used in these sequences have a relatively long duration (e.g., 0.3-0.6 ms). Here, we compared UTE and inversion-recovery (IR) UTE sequences used with either hard or half excitation pulses (durations 32 mu s or 472 mu s, respectively) for imaging myelin in native and deuterated ovine brain at 3T. MethodsFreshly frozen ovine brains were dissected into approximate to 2mm-thick pure white matter and approximate to 3 to 8mm-thick cerebral hemisphere specimens, which were imaged before and/or after different immersion time in deuterium oxide. ResultsBicomponent analysis of UTE signals obtained with hard excitation pulses detected an ultrashort T-2 component (STC) fraction (f(S)) of 0% to 10% in native specimens, and up to approximate to 86% in heavily deuterated specimens. f(S) values were significantly affected by the TIs used in IR-UTE sequences with either hard or half excitation pulses in native specimens but not in heavily deuterated specimens. The STC was in the range of 150 to 400 mu s in all UTE and IR-UTE measurements obtained with either hard or half excitation pulses. ConclusionOur results further support myelin protons as the major source of the ultrashort signals seen on IR-UTE images and demonstrate the potential of IR-UTE sequences with half excitation pulses for directly imaging myelin using clinical scanners. Magn Reson Med 80:538-547, 2017. (c) 2017 International Society for Magnetic Resonance in Medicine.
引用
收藏
页码:538 / 547
页数:10
相关论文
共 45 条
[1]   Characterization of Cerebral White Matter Properties Using Quantitative Magnetic Resonance Imaging Stains [J].
Alexander, Andrew L. ;
Hurley, Samuel A. ;
Samsonov, Alexey A. ;
Adluru, Nagesh ;
Hosseinbor, Ameer Pasha ;
Mossahebi, Pouria ;
Tromp, Do P. M. ;
Zakszewski, Elizabeth ;
Field, Aaron S. .
BRAIN CONNECTIVITY, 2011, 1 (06) :423-446
[2]   Multi-gradient-echo myelin water fraction imaging: Comparison to the multi-echo-spin-echo technique [J].
Alonso-Ortiz, Eva ;
Levesque, Ives R. ;
Pike, G. Bruce .
MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (03) :1439-1446
[3]   Diffusion tensor imaging (DTI)-based white matter mapping in brain research: A review [J].
Assaf, Yaniv ;
Pasternak, Ofer .
JOURNAL OF MOLECULAR NEUROSCIENCE, 2008, 34 (01) :51-61
[4]   Simultaneous spin-echo and gradient-echo BOLD measurements by dynamic MRS [J].
Cao, Peng ;
Hyder, Fahmeed ;
Zhou, Iris Y. ;
Zhang, Jevin W. ;
Xie, Victor B. ;
Tsang, Adrian ;
Wu, Ed X. .
NMR IN BIOMEDICINE, 2017, 30 (09)
[5]   Optimization of RF Excitation to Maximize Signal and T2 Contrast of Tissues With Rapid Transverse Relaxation [J].
Carl, Michael ;
Bydder, Mark ;
Du, Jiang ;
Takahashi, Atsushi ;
Han, Eric .
MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (02) :481-490
[6]   DYNAMIC AND CHEMICAL FACTORS AFFECTING WATER PROTON RELAXATION BY MACROMOLECULES [J].
CECKLER, TL ;
WOLFF, SD ;
YIP, V ;
SIMON, SA ;
BALABAN, RS .
JOURNAL OF MAGNETIC RESONANCE, 1992, 98 (03) :637-645
[7]   Gleaning Multicomponent T1 and T2 Information From Steady-State Imaging Data [J].
Deoni, Sean C. L. ;
Rutt, Brian K. ;
Arun, Tarunya ;
Pierpaoli, Carlo ;
Jones, Derek K. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (06) :1372-1387
[8]   Compartmental study of T1 and T2 in rat brain and trigeminal nerve in vivo [J].
Does, MD ;
Gore, JC .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (02) :274-283
[9]   Ultrashort echo time (UTE) magnetic resonance imaging of the short T2 components in white matter of the brain using a clinical 3T scanner [J].
Du, Jiang ;
Ma, Guolin ;
Li, Shihong ;
Carl, Michael ;
Szeverenyi, Nikolaus M. ;
VandenBerg, Scott ;
Corey-Bloom, Jody ;
Bydder, Graeme M. .
NEUROIMAGE, 2014, 87 :32-41
[10]   Inversion recovery ultrashort echo time imaging of ultrashort T2 tissue components in ovine brain at 3 T: a sequential D2O exchange study [J].
Fan, Shu-Juan ;
Ma, Yajun ;
Chang, Eric Y. ;
Bydder, Graeme M. ;
Du, Jiang .
NMR IN BIOMEDICINE, 2017, 30 (10)