Bacteria-modified amperometric immunosensor for a Brucella melitensis antibody assay

被引:15
作者
Li, ZZ
Gong, FC
Shen, GL
Yu, RQ [1 ]
机构
[1] Hunan Univ, Coll Chem & Chem Engn, State Key Lab ChemoBiosensing & Chemometr, Changsha 410082, Peoples R China
[2] Ling Ling Teachers Coll, Dept Chem & Biol, Yongzhou 425006, Peoples R China
关键词
D O I
10.2116/analsci.18.625
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel amperometric immunosensor setup is described which uses horseradish peroxidase (HRP) as a label in conjunction with a current-based Brucella sensor. The Bacteria modified immunosensor was constructed by using a biocomposite formed by dispersing graphite powder into a mixture of Brucella melitensis and silicate polymer gel. The enzyme-labeled antibody can readily diffuse toward the encapsulated antigen (Brucella melitensis), which retains its binding properties, and the association reaction is easily detected at the surface exposed to the solution. The use of an o-aminophenol (o-AP) substrate and amperometric detection at -150 mV (vs. SCE) results in a relatively low detection limit of 3.5 ng/ml and a linear detection range of 3.5 ng/ml to 200 ng/ml. Based on an optimized parameter, the prepared sensor was used to detect the Brucella melitensis antibody in serum samples by using a competitive binding assay. The results demonstrate the feasibility of employing the proposed immunosensor for the detection for Brucella melitensis antibody in a clinical analysis.
引用
收藏
页码:625 / 630
页数:6
相关论文
共 22 条
  • [11] Amperometric hydrogen peroxide biosensor based on immobilization of peroxidase in chitosan matrix crosslinked with glutaraldehyde
    Miao, YQ
    Tan, SN
    [J]. ANALYST, 2000, 125 (09) : 1591 - 1594
  • [12] Inert metal-modified, composite ceramic-carbon, amperometric biosensors: Renewable, controlled reactive layer
    Sampath, S
    Lev, O
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (13) : 2015 - 2021
  • [13] Amperometric immunosensors based on rigid conducting immunocomposites
    Santandreu, M
    Cespedes, F
    Alegret, S
    MartinezFabregas, E
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (11) : 2080 - 2085
  • [14] Advances in electrochemical immunosensors
    Skladal, P
    [J]. ELECTROANALYSIS, 1997, 9 (10) : 737 - 745
  • [15] Flow injection immunoanalysis based on a magnetoimmunosensor system
    Solé, S
    Alegret, S
    Céspedes, F
    Fàbregas, E
    Díez-Caballero, T
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (08) : 1462 - 1467
  • [16] SOL-GEL-DERIVED CERAMIC-CARBON COMPOSITE ELECTRODES - INTRODUCTION AND SCOPE OF APPLICATIONS
    TSIONSKY, M
    GUN, G
    GLEZER, V
    LEV, O
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (10) : 1747 - 1753
  • [17] Screen-printable sol-gel enzyme-containing carbon inks
    Wang, J
    Pamidi, PVA
    Park, DS
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (15) : 2705 - 2708
  • [18] Sol-gel-derived metal-dispersed carbon composite amperometric biosensors
    Wang, J
    Pamidi, PVA
    Park, DS
    [J]. ELECTROANALYSIS, 1997, 9 (01) : 52 - 55
  • [19] AFFINITY OF ANTIFLUORESCEIN ANTIBODIES ENCAPSULATED WITHIN A TRANSPARENT SOL-GEL GLASS
    WANG, R
    NARANG, U
    PRASAD, PN
    BRIGHT, FV
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (19) : 2671 - 2675
  • [20] WU JG, 1989, CLIN IMMUNOLOGY ANAL