摘要
An analytical validation of the precision and accuracy of the perturbation peak (PP) method for determination of single and competitive thermodynamic isotherm parameters was performed using frontal analysis as a reference. The isotherm parameters of 11alpha-hydroxyprogesterone were determined in an achiral system and the isotherm parameters of (+)-methyl L-mandelate and (-)-methyl D-mandelate were determined in a chiral system, both for the single components and for the competitive binary mixture. The experimental errors in the PP method using different injection techniques were investigated, and we devised a new injection technique for the determination of competitive isotherm parameters that considerably reduced the experimental errors and also made both perturbation peaks detectable. We showed that the PP method with the new injection technique can be used to determine isotherm parameters directly from a racemic mixture. These parameters agreed well with those determined using several enantiomer ratios. Elution-band profiles simulated using the isotherm parameters showed excellent agreement with experimental profiles.
引用
收藏
页码:4856 / 4865
页数:10
相关论文
共 20 条
[1]   Quantification of single solute and competitive adsorption isotherms using a closed-loop perturbation method [J].
Blümel, C ;
Hugo, P ;
Seidel-Morgenstern, A .
JOURNAL OF CHROMATOGRAPHY A, 1999, 865 (1-2) :51-71
[2]   Comparison between adsorption isotherm determination techniques and overloaded band profiles on four batches of monolithic columns [J].
Cavazzini, A ;
Felinger, A ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1012 (02) :139-149
[3]   COMPLETE DESIGN OF A SIMULATED MOVING-BED [J].
CHARTON, F ;
NICOUD, RM .
JOURNAL OF CHROMATOGRAPHY A, 1995, 702 (1-2) :97-112
[4]   Adsorption isotherms of nonionic surfactants in SBA-15 measured by micro-column chromatography [J].
Findenegg, G. H. ;
Eltekov, A. Y. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1150 (1-2) :236-240
[5]   THEORETICAL-STUDY OF HIGH-CONCENTRATION ELUTION PROFILES AND LARGE SYSTEM PEAKS IN NONLINEAR CHROMATOGRAPHY [J].
FORNSTEDT, T ;
GUICHON, G .
ANALYTICAL CHEMISTRY, 1994, 66 (13) :2116-2128
[6]   Nonlinear effects in LC and chiral LC [J].
Fornstedt, T ;
Guiochon, G .
ANALYTICAL CHEMISTRY, 2001, 73 (21) :608A-617A
[7]   Theoretical and experimental study of binary perturbation peaks with focus on peculiar retention behaviour and vanishing peaks in chiral liquid chromatography [J].
Forssén, P ;
Lindholm, J ;
Fornstedt, T .
JOURNAL OF CHROMATOGRAPHY A, 2003, 991 (01) :31-45
[8]   Study of the mass transfer kinetics in a monolithic column [J].
Gritti, F ;
Piatkowski, W ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2003, 983 (1-2) :51-71
[9]   Design of the simulated moving bed process based on adsorption isotherm measurements using a perturbation method [J].
Heuer, C ;
Küsters, E ;
Plattner, T ;
Seidel-Morgenstern, A .
JOURNAL OF CHROMATOGRAPHY A, 1998, 827 (02) :175-191
[10]  
*ICH, 2002, ITN C HARM TECHN REQ