Relationship between the ablation index, lesion formation, and incidence of steam pops

被引:38
作者
Mori, Hitoshi [1 ]
Kato, Ritsushi [1 ]
Sumitomo, Naokata [2 ]
Ikeda, Yoshifumi [1 ]
Goto, Koji [1 ]
Tanaka, Sayaka [1 ]
Asano, So [1 ]
Tahara, Mai [1 ]
Nagase, Takahiko [1 ]
Iwanaga, Shiro [1 ]
Muramatsu, Toshihiro [1 ]
Matsumoto, Kazuo [1 ]
机构
[1] Saitama Med Univ, Dept Cardiol, Int Med Ctr, 1397-1 Yamane Hidaka, Saitama 3501298, Japan
[2] Saitama Med Univ, Int Med Ctr, Dept Pediat Cardiol, Hidaka, Japan
关键词
ablation index; Catheter ablation; lesion size; radiofrequency; steam pop; PULMONARY VEIN ISOLATION; CATHETER CONTACT FORCE; RADIOFREQUENCY; RECONNECTION; CONDUCTION; IMPEDANCE; ELECTRODE; SIZE;
D O I
10.1002/joa3.12195
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The ablation index (AI) is reported to be useful for a durable pulmonary vein isolation (PVI). However, there have been no studies investigating the relationship between the power, contact force (CF), AI, and steam pops. Methods Using an in vitro model, ablation energy was delivered until a steam pop occurred and the time to the steam pop and AI when the steam pop occurred were measured. The experiment was performed with a combination of various powers (20, 30, 40, and 50 W) and contact forces (CFs) (10, 30, and 50 g) 20 times for each setting. The analysis consisted of two protocols. The first protocol was a comparison between the ablation power and several parameters under the same CF (10, 30, and 50 g). The second protocol was a comparison between the CF and several parameters under the same power (20, 30, 40, and 50 W). The correlation between the lesion formation and ablation parameters was evaluated. Results The AI value when steam pops occurred varied depending on the ablation settings. All AI median values were <500 under an ablation power of 50 W. On other hand, the median ablation time up to the steam pop was more than 46 seconds, but all median values of the AI were more than 550 under an ablation with 20 W. Conclusions The AI cannot predict steam pops. A low power and long duration ablation could obtain a high AI value. However, high-power ablation could not obtain a high AI value because of an early occurrence of steam pops.
引用
收藏
页码:636 / 644
页数:9
相关论文
共 19 条
[1]   High-power short-duration versus standard radiofrequency ablation: Insights on lesion metrics [J].
Bourier, Felix ;
Duchateau, Josselin ;
Vlachos, Konstantinos ;
Lam, Anna ;
Martin, Claire A. ;
Takigawa, Masateru ;
Kitamura, Takeshi ;
Frontera, Antonio ;
Cheniti, Ghassen ;
Pambrun, Thomas ;
Klotz, Nicolas ;
Denis, Arnaud ;
Derval, Nicolas ;
Cochet, Hubert ;
Sacher, Frederic ;
Hocini, Meleze ;
Haissaguerre, Michel ;
Jais, Pierre .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (11) :1570-1575
[2]  
Calkins H, 2012, HEART RHYTHM, V9, P632, DOI 10.1016/j.hrthm.2011.12.016
[3]   Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation [J].
Cappato, R ;
Negroni, S ;
Pecora, D ;
Bentivegna, S ;
Lupo, PP ;
Carolei, A ;
Esposito, C ;
Furlanello, F ;
De Ambroggi, L .
CIRCULATION, 2003, 108 (13) :1599-1604
[4]   Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values [J].
Das, Moloy ;
Loveday, Jonathan J. ;
Wynn, Gareth J. ;
Gomes, Sean ;
Saeed, Yawer ;
Bonnett, Laura J. ;
Waktare, Johan E. P. ;
Todd, Derick M. ;
Hall, Mark C. S. ;
Snowdon, Richard L. ;
Modi, Simon ;
Gupta, Dhiraj .
EUROPACE, 2017, 19 (05) :775-783
[5]   Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine Heart Electrogram Amplitude, Impedance, and Electrode Temperature Are Poor Predictors of Electrode-Tissue Contact Force and Lesion Size [J].
Ikeda, Atsushi ;
Nakagawa, Hiroshi ;
Lambert, Hendrik ;
Shah, Dipen C. ;
Fonck, Edouard ;
Yulzari, Aude ;
Sharma, Tushar ;
Pitha, Jan V. ;
Lazzara, Ralph ;
Jackman, Warren M. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (06) :1174-1180
[6]   EFFICAS II: optimization of catheter contact force improves outcome of pulmonary vein isolation for paroxysmal atrial fibrillation [J].
Kautzner, Josef ;
Neuzil, Petr ;
Lambert, Hendrik ;
Peichl, Petr ;
Petru, Jan ;
Cihak, Robert ;
Skoda, Jan ;
Wichterle, Dan ;
Wissner, Erik ;
Yulzari, Aude ;
Kuck, Karl-Heinz .
EUROPACE, 2015, 17 (08) :1229-1235
[7]   High-Power and Short-Duration Ablation for Pulmonary Vein Isolation Biophysical Characterization [J].
Leshem, Eran ;
Zilberman, Israel ;
Tschabrunn, Cory M. ;
Barkagan, Michael ;
Contreras-Valdes, Fernando M. ;
Govari, Assaf ;
Anter, Elad .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (04) :467-479
[8]  
Nakagawa H., 2018, CIRCULATION, V128
[9]  
Nakagawa H., 2013, HEART RHYTHM, V10, pS481
[10]   Resumption of electrical conduction in previously isolated pulmonary veins - Rationale for a different strategy? [J].
Nanthakumar, K ;
Plumb, VJ ;
Epstein, AE ;
Veenhuyzen, GD ;
Link, D ;
Kay, GN .
CIRCULATION, 2004, 109 (10) :1226-1229