Management of traumatic implantable cardioverter defibrillator lead perforation of the right ventricle after car accident: a case report

被引:5
作者
Sassone, Biagio [1 ]
Gabrieli, Luca [1 ]
Boggian, Giulio [1 ]
Pilato, Emanuele [2 ]
机构
[1] Azienda USL Bologna, Bentivoglio Hosp, Div Cardiol, I-40010 Bologna, Italy
[2] Univ Bologna, Inst Cardiosurg, I-40138 Bologna, Italy
来源
EUROPACE | 2009年 / 11卷 / 07期
关键词
PACEMAKER;
D O I
10.1093/europace/eup111
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The authors reported a case of traumatic implantable cardioverter defibrillator (ICD) lead perforation of the right ventricular (RV) apex caused by a motor vehicle accident. Clinical and echocardiographic features combined with changes in electrical parameters of the offending lead were decisive for the final diagnosis. Optimal management of ICD lead RV wall perforation is currently unclear. In our report, RV perforation was responsible for cardiac tamponade. This complication was uneventfully managed by open surgical procedure.
引用
收藏
页码:961 / 962
页数:2