A new self-assembled layer-by-layer glucose biosensor based on chitosan biopolymer entrapped enzyme with nitrogen doped graphene

被引:62
作者
Barsan, Madalina M. [1 ]
David, Melinda [1 ,2 ]
Florescu, Monica [3 ]
Tugulea, Laura [2 ]
Brett, Christopher M. A. [1 ]
机构
[1] Univ Coimbra, Dept Quim, Fac Ciencias & Tecnol, P-3004535 Coimbra, Portugal
[2] Univ Bucuresti, Fac Fizica, Magurele 077125, Romania
[3] Univ Transilvania Brasov, Fac Med, Brasov 500019, Romania
关键词
Layer-by layer; Quartz crystal microbalance; Nitrogen doped graphene; Enzyme; Chitosan; CARBON NANOTUBE; OXIDASE; FILM; IMMOBILIZATION; ELECTRODES; NANOMATERIALS;
D O I
10.1016/j.bioelechem.2014.06.004
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The layer-by-layer (LbL) technique has been used for the construction of a new enzyme biosensor. Multilayer films containing glucose oxidase, GOx, and nitrogen-doped graphene (NG) dispersed in the biocompatible positively-charged polymer chitosan (chit(+)(NG + GOx)), together with the negatively charged polymer poly(styrene sulfonate), PSS-, were assembled by alternately immersing a gold electrode substrate in chit(+)(NG + GOx) and PSS- solutions. Gravimetric monitoring during LbL assembly by an electrochemical quartz microbalance enabled investigation of the adsorption mechanism and deposited mass for each monolayer. Cyclic voltammetry and electrochemical impedance spectroscopy were used to characterize the LbL modified electrodes, in order to establish the contribution of each monolayer to the overall electrochemical properties of the biosensor. The importance of NG in the biosensor architecture was evaluated by undertaking a comparative study without NG in the chit layer. The GOx biosensor's analytical properties were evaluated by fixed potential chronoamperometry and compared with similar reported biosensors. The biosensor operates at a low potential of -0.2 V vs., Ag/AgCl, exhibiting a high sensitivity of 10.5 mu A cm(-2) mM(-1), and a detection limit of 64 mu M. This study shows a simple approach in developing new biosensor architectures, combining the advantages of nitrogen-doped graphene with the LbL technique for enzyme immobilization. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:46 / 52
页数:7
相关论文
共 34 条
[1]  
[Anonymous], 1987, ANALYST, V112, P199
[2]   Metal decorated graphene nanosheets as immobilization matrix for amperometric glucose biosensor [J].
Baby, Tessy Theres ;
Aravind, S. S. Jyothirmayee ;
Arockiadoss, T. ;
Rakhi, R. B. ;
Ramaprabhu, S. .
SENSORS AND ACTUATORS B-CHEMICAL, 2010, 145 (01) :71-77
[3]   Biosensors based on carbon nanotubes [J].
Balasubramanian, Kannan ;
Burghard, Marko .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 385 (03) :452-468
[4]   Carbon nanotube modified carbon cloth electrodes: Characterisation and application as biosensors [J].
Barsan, Madalina M. ;
Carvalho, Ricardo C. ;
Zhong, Yu ;
Sun, Xueliang ;
Brett, Christopher M. A. .
ELECTROCHIMICA ACTA, 2012, 85 :203-209
[5]  
Barsoukov E., 2005, IMPEDANCE SPECTROSCO, DOI [10.1002/0471716243, DOI 10.1002/0471716243]
[6]   Layer-by-layer biomolecular assemblies for enzyme sensors, immunosensing, and nano architectures [J].
Campàs, M ;
O'Sullivan, C .
ANALYTICAL LETTERS, 2003, 36 (12) :2551-2569
[7]   The effect of the layer structure on the activity of immobilized enzymes in ultrathin films [J].
Caseli, Luciano ;
dos Santos, David S., Jr. ;
Foschini, Mauricio ;
Goncalves, Debora ;
Oliveira, Osvaldo N., Jr. .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2006, 303 (01) :326-331
[8]   Development and application of oxysilane sol-gel electrochemical glucose biosensors based on cobalt hexacyanoferrate modified carbon film electrodes [J].
Florescu, Monica ;
Barsan, Madalina ;
Pauliukaite, Rasa ;
Brett, Christopher M. A. .
ELECTROANALYSIS, 2007, 19 (2-3) :220-226
[9]   Nitrogen doping effects on the structure of graphene [J].
Geng, Dongsheng ;
Yang, Songlan ;
Zhang, Yong ;
Yang, Jinli ;
Liu, Jian ;
Li, Ruying ;
Sham, Tsun-Kong ;
Sun, Xueliang ;
Ye, Siyu ;
Knights, Shanna .
APPLIED SURFACE SCIENCE, 2011, 257 (21) :9193-9198
[10]   Application of functionalised carbon nanotubes immobilised into chitosan films in amperometric enzyme biosensors [J].
Ghica, Mariana Emilia ;
Pauliukaite, Rasa ;
Fatibello-Filho, Orlando ;
Brett, Christopher M. A. .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 142 (01) :308-315