Improving the measurement accuracy of CRD fibre amplified loop gas sensing system by using a digital LMS adaptive filter

被引:24
作者
Ni, N. [1 ]
Chan, C. C. [1 ]
机构
[1] Nanyang Technol Univ, Sch Chem & Biomed Engn, Singapore 639798, Singapore
关键词
cavity ring-down fibre amplified loop gas sensing system; digital least mean square ( LMS) adaptive filter; amplified spontaneous emission (ASE) noise; measurement accuracy;
D O I
10.1088/0957-0233/17/8/042
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A mathematical model of a cavity ring-down (CRD) fibre amplified loop gas sensing system is initially proposed. A digital least mean square (LMS) adaptive filter is designed and demonstrated the reduction of the detection error due to an amplified spontaneous emission (ASE) noise which is induced by an erbium doped fibre amplifier (EDFA) in the CRD loop. The simulation results show that the detection error of 16 ppm is obtained.
引用
收藏
页码:2349 / 2354
页数:6
相关论文
共 10 条
  • [1] ARYA V, THESIS VIRGINIA POLY
  • [2] ATHERTON K, 2002, OPT FIBR SENS C, V1, P581
  • [3] BUSCH KW, 1999, CAVITY RINGDOWN SPEC
  • [4] Study on noise characteristic of gain-clamped erbium-doped fiber-ring lasing amplifier
    Cai, M
    Liu, XM
    Cui, JC
    Tang, PS
    Peng, JD
    [J]. IEEE PHOTONICS TECHNOLOGY LETTERS, 1997, 9 (08) : 1093 - 1095
  • [5] IMPROVED AVERAGE POWER ANALYSIS TECHNIQUE FOR ERBIUM-DOPED FIBER AMPLIFIERS
    HODGKINSON, TG
    [J]. IEEE PHOTONICS TECHNOLOGY LETTERS, 1992, 4 (11) : 1273 - 1275
  • [6] IFEACHOR EC, 2002, DIGIT SIGNAL PROCESS, P645
  • [7] DEPENDENCE OF OPTICAL AMPLIFIER NOISE-FIGURE ON RELATIVE-INTENSITY-NOISE
    JACOBS, I
    [J]. JOURNAL OF LIGHTWAVE TECHNOLOGY, 1995, 13 (07) : 1461 - 1465
  • [8] The HITRAN 2004 molecular spectroscopic database
    Rothman, LS
    Jacquemart, D
    Barbe, A
    Benner, DC
    Birk, M
    Brown, LR
    Carleer, MR
    Chackerian, C
    Chance, K
    Coudert, LH
    Dana, V
    Devi, VM
    Flaud, JM
    Gamache, RR
    Goldman, A
    Hartmann, JM
    Jucks, KW
    Maki, AG
    Mandin, JY
    Massie, ST
    Orphal, J
    Perrin, A
    Rinsland, CP
    Smith, MAH
    Tennyson, J
    Tolchenov, RN
    Toth, RA
    Vander Auwera, J
    Varanasi, P
    Wagner, G
    [J]. JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2005, 96 (02) : 139 - 204
  • [9] An investigation of an optical fibre amplifier loop for intra-cavity and ring-down cavity loss measurements
    Stewart, G
    Atherton, K
    Yu, HB
    Culshaw, B
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (07) : 843 - 849
  • [10] Trace gas analysis by diode laser cavity ring-down spectroscopy
    Yan, WB
    Krusen, C
    Optics, T
    Dudek, J
    Lehmann, K
    Rabinowitz, P
    [J]. 2002 IEEE/SEMI ADVANCED SEMICONDUCTOR MANUFACTURING CONFERENCE AND WORKSHOP: ADVANCING THE SCIENCE OF SEMICONDUCTOR MANUFACTURING EXCELLENCE, 2002, : 319 - 323