We evaluated the performance characteristics of a new, real-time nucleic acid sequence-based amplification (NASBA) assay that incorporates molecular beacon technology for detection of human immunodeficiency virus type 1 (HIV-1). The quantitative results were comparable to those obtained with three leading commercially available assays. The analytical sensitivity was 37 IU/ml. The NASBA assay detected clinically relevant recombinant viruses and all group M HIV-1 subtypes.
引用
收藏
页码:2280 / 2282
页数:3
相关论文
共 7 条
[1]   RAPID AND SIMPLE METHOD FOR PURIFICATION OF NUCLEIC-ACIDS [J].
BOOM, R ;
SOL, CJA ;
SALIMANS, MMM ;
JANSEN, CL ;
WERTHEIMVANDILLEN, PME ;
VANDERNOORDAA, J .
JOURNAL OF CLINICAL MICROBIOLOGY, 1990, 28 (03) :495-503
[2]   Multicenter evaluation of the NucliSens EasyQ HIV-1 v1.1 assay for the quantitative detection of HIV-1 RNA in plasma [J].
de Mendoza, C ;
Koppelman, M ;
Montès, B ;
Ferre, V ;
Soriano, V ;
Cuypers, H ;
Segondy, M ;
Oosterlaken, T .
JOURNAL OF VIROLOGICAL METHODS, 2005, 127 (01) :54-59
[3]   Characteristics and applications of nucleic acid sequence-based amplification (NASBA) [J].
Deiman, B ;
van Aarle, P ;
Sillekens, P .
MOLECULAR BIOTECHNOLOGY, 2002, 20 (02) :163-179
[4]   Multicenter comparison of Roche COBAS AMPLICOR MONITOR version 1.5, Organon Teknika NucliSens QT with extractor, and Bayer Quantiplex version 3.0 for quantification of human immunodeficiency virus type 1 RNA in plasma [J].
Murphy, DG ;
Côté, L ;
Fauvel, M ;
René, P ;
Vincelette, J .
JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (11) :4034-4041
[5]   Impact of clade diversity on HIV-1 virulence, antiretroviral drug sensitivity and drug resistance [J].
Spira, S ;
Wainberg, MA ;
Loemba, H ;
Turner, D ;
Brenner, BG .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2003, 51 (02) :229-240
[6]   Evaluation of the NucliSens EasyQ assay in HIV-1-infected individuals in South Africa [J].
Stevens, W ;
Wiggill, T ;
Horsfield, P ;
Coetzee, L ;
Scott, LE .
JOURNAL OF VIROLOGICAL METHODS, 2005, 124 (1-2) :105-110
[7]   Quantitative detection of HIV-1 RNA using NucliSens EasyQ HIV-1 assay [J].
Yao, J ;
Liu, Z ;
Ko, LS ;
Pan, G ;
Jiang, Y .
JOURNAL OF VIROLOGICAL METHODS, 2005, 129 (01) :40-46