Interplay effect modeling in stereotactic body radiotherapy treatment of liver cancer using volumetric modulated arc therapy

被引:7
作者
Thaper, Deepak [1 ,3 ]
Oinam, Arun S. [2 ]
Kamal, Rose [1 ,3 ]
Singh, Gaganpreet [1 ]
Handa, Bhumika [1 ]
Kumar, Vivek [1 ]
Yadav, Hanuman P. [3 ]
机构
[1] Panjab Univ, Ctr Med Phys, Chandigarh, Punjab, India
[2] PGIMER, Radiotherapy Dept, Reg Canc Ctr, Chandigarh, Punjab, India
[3] Inst Liver & Biliary Sci, Radiat Oncol Dept, New Delhi, India
关键词
SBRT; Interplay effect; Dose Variation; Parameters; Predictive Model; Optimization; DOSIMETRIC IMPACT; ORGAN MOTION; RESPIRATORY MOTION; RADIATION-THERAPY; DOSE DELIVERY; RAPIDARC;
D O I
10.1007/s13246-020-00961-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To model the interplay effect and minimize it by a selection of optimum parameters value using a predictive model for SBRT of liver cancers. Ten cases of liver tumors treated with the VMAT technique were selected retrospectively. The dosimetric error due to the interplay effect was measured with a micro ionization chamber (0.015cm(3)) in a Quasar phantom simulating the moving tumor. The interplay effect dependent parameter's viz. patient breaths per minute, the amplitude of respiration, fractional dose (FD), plan complexity due to different energies (Relative degree of modulation), degree of modulation due to a different level of dose optimization constraints, and dose rate (DR) were measured. For the predictive model, mathematical equations were modeled in python from 300 combinations of proposed parameters using multivariate regression analysis. It was observed that the dose variation reduced from -8.44% to -5.16% for change in the BPM values from 7 to 31 and similarly for amplitude, the dose variation reduced from -9.44% to -4.93% for change in amplitude value from 16 mm to 2 mm. The DR and FD have a prominent effect with R-2 values of 0.990 and 0.880 respectively. The calculated mean square errors of equations excluding amplitude for the predictive model were 0.90 and 0.82 whereas those for equations excluding BPM were 1.31 and 1.41 for 6 MV and 10 MV beams respectively. The values of the parameters can be prospectively optimized by the use of the predictive model according to clinical situations, so dose variation can be minimized.
引用
收藏
页码:123 / 134
页数:12
相关论文
共 24 条
[1]   A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy [J].
Arnfield, MR ;
Siebers, JV ;
Kim, JO ;
Wu, QW ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (10) :2231-2241
[2]   Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation [J].
Bortfeld, T ;
Jokivarsi, K ;
Goitein, M ;
Kung, J ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (13) :2203-2220
[3]   Evaluation of the interplay effect when using RapidArc to treat targets moving in the craniocaudal or right-left direction [J].
Court, Laurence ;
Wagar, Matthew ;
Berbeco, Ross ;
Reisner, Adam ;
Winey, Brian ;
Schofield, Debbie ;
Ionascu, Dan ;
Allen, Aaron M. ;
Popple, Richard ;
Lingos, Tania .
MEDICAL PHYSICS, 2010, 37 (01) :4-11
[4]   Use of a realistic breathing lung phantom to evaluate dose delivery errors [J].
Court, Laurence E. ;
Seco, Joao ;
Lu, Xing-Qi ;
Ebe, Kazuyu ;
Mayo, Charles ;
Ionascu, Dan ;
Winey, Brian ;
Giakoumakis, Nikos ;
Aristophanous, Michalis ;
Berbeco, Ross ;
Rottman, Joerg ;
Bogdanov, Madeleine ;
Schofield, Deborah ;
Lingos, Tania .
MEDICAL PHYSICS, 2010, 37 (11) :5850-5857
[5]   Stereotactic Body Radiation Therapy for Liver Tumors: Current Status and Perspectives [J].
Doi, Hiroshi ;
Beppu, Naohito ;
Kitajima, Kazuhiro ;
Kuribayashi, Kozo .
ANTICANCER RESEARCH, 2018, 38 (02) :591-599
[6]   Three-dimensional versus four-dimensional dose calculation for volumetric modulated arc therapy of hypofractionated treatments [J].
Ehrbar, Stefanie ;
Lang, Stephanie ;
Stieb, Sonja ;
Riesterer, Oliver ;
Stark, Luisa Sabrina ;
Guckenberger, Matthias ;
Kloeck, Stephan .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2016, 26 (01) :45-53
[7]   The influence of plan modulation on the interplay effect in VMAT liver SBRT treatments [J].
Hubley, Emily ;
Pierce, Greg .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 40 :115-121
[8]   An experimental investigation on intra-fractional organ motion effects in lung IMRT treatments [J].
Jiang, SB ;
Pope, C ;
Al Jarrah, KM ;
Kung, JH ;
Bortfeld, T ;
Chen, GTY .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (12) :1773-1784
[9]   Minimizing dose variation from the interplay effect in stereotactic radiation therapy using volumetric modulated arc therapy for lung cancer [J].
Kubo, Kazuki ;
Monzen, Hajime ;
Tamura, Mikoto ;
Hirata, Makoto ;
Ishii, Kentaro ;
Okada, Wataru ;
Nakahara, Ryuta ;
Kishimoto, Shun ;
Kawamorita, Ryu ;
Nishimura, Yasumasa .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (02) :121-127
[10]   A Method Incorporating 4DCT Data for Evaluating the Dosimetric Effects of Respiratory Motion in Single Arc IMAT [J].
Kuo, H. ;
Mah, D. ;
Wu, A. ;
Chuang, K. ;
Hong, L. ;
Yaparpalvi, R. ;
Spierer, M. ;
Kalnicki, S. .
MEDICAL PHYSICS, 2010, 37 (06)