The new design and validation of an epithermal neutron flux detector using 71Ga(n, γ)72Ga reaction for BNCT

被引:8
作者
Guan, X. C. [1 ]
Gong, Y. [1 ]
Murata, I [2 ]
Wang, T. S. [1 ]
机构
[1] Lanzhou Univ, Sch Nucl Sci & Technol, Tianshui South Rd 222, Lanzhou 730000, Gansu, Peoples R China
[2] Osaka Univ, Grad Sch Engn, Div Sustainable Energy & Environm Engn, Yamada Oka 2-1, Suita, Osaka 5650871, Japan
关键词
Beam-line instrumentation (beam position and profile monitors; beam-intensity monitors; bunch length monitors); Instrumentation for neutron sources; Neutron detectors (cold; thermal; fast neutrons); Solid state detectors; MONTE-CARLO; OPTIMIZATION; PERFORMANCE; MONITOR;
D O I
10.1088/1748-0221/14/06/P06016
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The epithermal neutron (0.5 eV-10 keV) flux is an essential characteristic for the boron neutron capture therapy (BNCT) neutron beam. The epithermal neutron flux measurement of a BNCT neutron beam is always a critical issue to its quality assessment and radiation dose evaluation. In this work, a cylindrical activation detector using Ga-71(n, gamma)Ga-72 reaction is designed by Monte Carlo simulations to determine the epithermal neutron flux of the BNCT neutron beam. In the detector, the activation material, i.e., gallium nitride (GaN) wafer, is positioned in the geometrical center of a high-density polyethylene (HDPE) cylinder covered with cadmium (Cd) foil. On completion of the design, the performance of the detector is validated by Monte Carlo simulations using reactor- and accelerator-based BNCT neutron beams. It is concluded from the performance validation results that the detector can accurately determine the epithermal neutron flux of the BNCT neutron beam. The designed epithermal neutron flux detector will be efficiently applicable to characterize the field of the BNCT neutron beam.
引用
收藏
页数:10
相关论文
共 20 条
[1]  
[Anonymous], 2003, LAUR031987 LOS AL NA
[2]  
[Anonymous], 2010, CAMBRIDGE DICT STAT, DOI DOI 10.1017/CBO9780511779633
[3]   Beam shaping assembly study for BNCT facility based on a 2.5MeV proton accelerator on Li target [J].
Fantidis, J. G. .
JOURNAL OF THEORETICAL AND APPLIED PHYSICS, 2018, 12 (04) :249-256
[4]  
Fantidis JG, 2013, INT J RADIAT RES, V11, P207
[5]   Conceptual design of BNCT facility based on the TRR medical room [J].
Golshanian, M. ;
Rajabi, A. A. ;
Kasesaz, Y. .
JOURNAL OF INSTRUMENTATION, 2017, 12
[6]   Performance verification of an epithermal neutron flux monitor using accelerator-based BNCT neutron sources [J].
Guan, X. ;
Murata, I. ;
Wang, T. .
JOURNAL OF INSTRUMENTATION, 2017, 12
[7]   The neutron flux monitors from 20 keV to 1 MeV for BNCT: performance study using accelerator-based neutron sources [J].
Guan, X. C. ;
Gong, Y. ;
Murata, I. ;
Wang, T. S. .
JOURNAL OF INSTRUMENTATION, 2018, 13
[8]  
Guan X.J., 2016, THESIS
[9]   Monte Carlo optimization of an epithermal neutron flux monitor for BNCT [J].
Guan, Xingcai ;
Murata, Isao ;
Wang, Tieshan .
JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2017, 54 (10) :1118-1122
[10]   Performance testing of the neutron flux monitors from 10 keV to 1 MeV developed for BNCT: A preliminary study [J].
Guan, Xingcai ;
Manabe, Masanobu ;
Tamaki, Shingo ;
Sato, Fuminobu ;
Murata, Isao ;
Wang, Tieshan .
APPLIED RADIATION AND ISOTOPES, 2017, 125 :119-123