On-line intracardiac echocardiography alone for Amplatzer Septal Occluder selection and device deployment in adult patients with atrial septal defect

被引:24
|
作者
Zanchetta, N [1 ]
机构
[1] Cittadella Gen Hosp, Cardiovasc Dept, Padua, Italy
关键词
intracardiac echocardiography; atrial septal defect; Amplatzer Septal Occluder;
D O I
10.1016/j.ijcard.2003.04.022
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: During the last few years, several different devices have been proposed for atrial septal defect (ASD) percutaneous closure. For the Amplatzer Septal Occluder (ASO) device, accurate balloon sizing is considered of paramount importance because the prosthesis waist has to be exactly adjusted to the defect diameter (+/- 1 mm). In this study, we aimed to demonstrate the possibility of marked misinterpreting of the actual defect size using the balloon technique in patients with secundum ASD and to evaluate the accuracy of intracardiac echocardiography (ICE) measurements as a new method for selecting the size of ASO device. Methods: Between February 1999 and December 2000, 166 consecutive adult patients underwent percutaneous transvenous secundum ASD occlusion using the ASO device. In 124 patients (control group), ASD were closed by conventional methods. In 13 patients (pilot group), balloon pulling technique was used in size selection, whereas ICE was used on-line to monitor device placement and off-line to assess its possibilities for accurate quantitative measurements and qualitative evaluation. In 31 patients (study group), ICE was used as the sole imaging tool both for guiding device selection and monitoring the procedure. All patients underwent complete transthoracic echocardiographic study before discharge and during follow-up visits at 3 and 12 months. Results: Successful device implantation was accomplished in 163 of the 166 patients (98.2%). Short-term follow-up results were available in all eligible patients at least 3 months. Complete occlusion was demonstrated in 91.4% and 92.2% of patients in the control and pilot groups, respectively, increasing to 97.3% in the study group (p < 0.01 vs. both control and pilot groups). There were no significant differences in mean ASO diameters in the control and pilot groups (20 +/- 7.7 and 22 +/- 5.4 mm, respectively), whereas the mean size of the devices used in the study group was significantly larger (27.4 +/- 6.2 mm, p < 0.01 vs. both control and pilot groups). In the pilot group, the underestimation effect of the balloon strategy was evident, with a mean 12.3% larger diameter required on ICE measurements. Moreover, a misalignment between the ASO and the atrial septum was seen on ICE in 9 of 13 patients of the pilot group, whereas good apposition of the ASO on the septum secundum was seen in all patients of the study group. Conclusion: ICE is a safe and effective method for selecting ASO size and continuous monitoring of the procedure. In contrast to the previously reported implantation procedure (device-to-defect ratio 1:1), a device 10-20% larger than invasively measured stretched defect diameter should be chosen and implanted on the basis of the ICE data. (C) 2003 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:61 / 68
页数:8
相关论文
共 50 条
  • [41] Left atrial appendage occlusion guided by intracardiac echocardiography in a patient with a 34 mm atrial septal defect occluder: a case report
    Korsholm, Kasper
    Jensen, Jesper Moller
    Nielsen-Kudsk, Jens Erik
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2023, 7 (12)
  • [42] Simultaneous transcatheter closure of an atrial septal defect with an Amplatzer septal occluder and radiofrequency ablation of an accessory connection
    Kannankeril, PJ
    Wax, DF
    Deal, BJ
    Johnsrude, CL
    Duffy, CE
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2000, 51 (01) : 55 - 57
  • [43] Efficacy and safety of atrial septal defect closure using Occlutech Figulla Flex II compared with Amplatzer Septal Occluder
    Rie Nakayama
    Yoichi Takaya
    Teiji Akagi
    Nobuhisa Watanabe
    Takashi Miki
    Koji Nakagawa
    Norihisa Toh
    Hiroshi Ito
    Heart and Vessels, 2021, 36 : 704 - 709
  • [44] Efficacy and safety of atrial septal defect closure using Occlutech Figulla Flex II compared with Amplatzer Septal Occluder
    Nakayama, Rie
    Takaya, Yoichi
    Akagi, Teiji
    Watanabe, Nobuhisa
    Miki, Takashi
    Nakagawa, Koji
    Toh, Norihisa
    Ito, Hiroshi
    HEART AND VESSELS, 2021, 36 (05) : 704 - 709
  • [45] Histopathological Workup of an Amplatzer Atrial Septal Defect Occluder After Surgical Removal
    Matthias Sigler
    Christian Jux
    Peter Ewert
    Pediatric Cardiology, 2006, 27 : 775 - 776
  • [46] Percutaneous closure of atrial septal defect via transjugular approach with the amplatzer septal occluder after unsuccessful attempt using the CardioSEAL device
    Sullebarger, JT
    Sayad, D
    Gerber, L
    Ettedgui, J
    Jimmo-Waumans, S
    Alcebo, PC
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (02) : 262 - 265
  • [47] Histopathological workup of an Amplatzer Atrial Septal Defect occluder after surgical removal
    Sigler, Matthias
    Jux, Christian
    Ewert, Peter
    PEDIATRIC CARDIOLOGY, 2006, 27 (06) : 775 - 776
  • [48] Unconventional deployment of atrial septal occluder in a patient with atrial septal defect, dextrocardia, and interrupted inferior vena cava
    Yerram, Sreekanth
    Aramalla, Sunitha
    Bhyravavajhala, Srinivas
    CARDIOLOGY IN THE YOUNG, 2020, 30 (08) : 1206 - 1208
  • [49] Transcatheter atrial septal defect closure with the amplatzer septal occluder: Five-year follow-up
    Yew, G
    Wilson, NJ
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (02) : 193 - 196
  • [50] Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding nitinol double disc device (Amplatzer septal occluder): multicentre UK experience
    Chan, KC
    Godman, MJ
    Walsh, K
    Wilson, N
    Redington, A
    Gibbs, JL
    HEART, 1999, 82 (03) : 300 - 306