On-line intracardiac echocardiography alone for Amplatzer Septal Occluder selection and device deployment in adult patients with atrial septal defect

被引:24
|
作者
Zanchetta, N [1 ]
机构
[1] Cittadella Gen Hosp, Cardiovasc Dept, Padua, Italy
关键词
intracardiac echocardiography; atrial septal defect; Amplatzer Septal Occluder;
D O I
10.1016/j.ijcard.2003.04.022
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: During the last few years, several different devices have been proposed for atrial septal defect (ASD) percutaneous closure. For the Amplatzer Septal Occluder (ASO) device, accurate balloon sizing is considered of paramount importance because the prosthesis waist has to be exactly adjusted to the defect diameter (+/- 1 mm). In this study, we aimed to demonstrate the possibility of marked misinterpreting of the actual defect size using the balloon technique in patients with secundum ASD and to evaluate the accuracy of intracardiac echocardiography (ICE) measurements as a new method for selecting the size of ASO device. Methods: Between February 1999 and December 2000, 166 consecutive adult patients underwent percutaneous transvenous secundum ASD occlusion using the ASO device. In 124 patients (control group), ASD were closed by conventional methods. In 13 patients (pilot group), balloon pulling technique was used in size selection, whereas ICE was used on-line to monitor device placement and off-line to assess its possibilities for accurate quantitative measurements and qualitative evaluation. In 31 patients (study group), ICE was used as the sole imaging tool both for guiding device selection and monitoring the procedure. All patients underwent complete transthoracic echocardiographic study before discharge and during follow-up visits at 3 and 12 months. Results: Successful device implantation was accomplished in 163 of the 166 patients (98.2%). Short-term follow-up results were available in all eligible patients at least 3 months. Complete occlusion was demonstrated in 91.4% and 92.2% of patients in the control and pilot groups, respectively, increasing to 97.3% in the study group (p < 0.01 vs. both control and pilot groups). There were no significant differences in mean ASO diameters in the control and pilot groups (20 +/- 7.7 and 22 +/- 5.4 mm, respectively), whereas the mean size of the devices used in the study group was significantly larger (27.4 +/- 6.2 mm, p < 0.01 vs. both control and pilot groups). In the pilot group, the underestimation effect of the balloon strategy was evident, with a mean 12.3% larger diameter required on ICE measurements. Moreover, a misalignment between the ASO and the atrial septum was seen on ICE in 9 of 13 patients of the pilot group, whereas good apposition of the ASO on the septum secundum was seen in all patients of the study group. Conclusion: ICE is a safe and effective method for selecting ASO size and continuous monitoring of the procedure. In contrast to the previously reported implantation procedure (device-to-defect ratio 1:1), a device 10-20% larger than invasively measured stretched defect diameter should be chosen and implanted on the basis of the ICE data. (C) 2003 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:61 / 68
页数:8
相关论文
共 50 条
  • [1] Real time three-dimensional transthoracic echocardiography for guiding Amplatzer septal occluder device deployment in patients with atrial septal defect
    Chen, Fong L.
    Hsiung, Ming C.
    Hsieh, Kai S.
    Li, Yi C.
    Chou, Ming C.
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2006, 23 (09): : 763 - 770
  • [2] Preprocedural Transthoracic Echocardiography Can Predict Amplatzer Septal Occluder Device Size for Transcatheter Atrial Septal Defect Closure
    Sah, Serena P.
    Bartakian, Sergio
    El-Said, Howaida
    Molkara, Delaram P.
    Printz, Beth
    Moore, John W.
    CONGENITAL HEART DISEASE, 2016, 11 (06) : 656 - 662
  • [3] The Amplatzer® Septal Occluder, a transcatheter device for atrial septal defect closure
    Fu, Yun-Ching
    Hijazi, Ziyad M.
    EXPERT REVIEW OF MEDICAL DEVICES, 2008, 5 (01) : 25 - 31
  • [4] Real-time 3D transoesophageal echocardiography for guiding Amplatzer septal occluder device deployment in an adult patient with atrial septal defect
    Acar, Philippe
    Massabuau, Pierre
    Elbaz, Meyer
    EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2008, 9 (06): : 822 - 823
  • [5] Comparison of the Occlutech® Figulla® Septal Occluder and Amplatzer® Septal Occluder for Atrial Septal Defect Device Closure
    Supaporn Roymanee
    Worakan Promphan
    Nakharin Tonklang
    Kanjarut Wongwaitaweewong
    Pediatric Cardiology, 2015, 36 : 935 - 941
  • [6] Amplatzer Septal Occluder Closure of Atrial Septal Defect: Evaluation of Transthoracic Echocardiography, Cardiac CT, and Transesophageal Echocardiography
    Ko, Sheung-Fat
    Liang, Chi-Di
    Yip, Hon-Kan
    Huang, Chung-Cheng
    Ng, Shu-Hang
    Huang, Chien-Fu
    Chen, Min-Chi
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (06) : 1522 - 1529
  • [7] Early clinical experience with use of the 'Amplatzer Septal Occluder' device for atrial septal defect
    Wilkinson, JL
    Goh, TH
    CARDIOLOGY IN THE YOUNG, 1998, 8 (03) : 295 - 302
  • [8] Dislocation of Amplatzer Septal Occluder Device after Closure of Secundum Atrial Septal Defect
    Piatkowski, Radoslaw
    Kochanowski, Janusz
    Scislo, Piotr
    Kochman, Janusz
    Opolski, Grzegorz
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2010, 23 (09) : 1007.e1 - 1007.e2
  • [9] Experience in one centre using the buttoned device for occlusion of atrial septal defect:: comparison with the Amplatzer septal occluder
    Godart, F
    Rey, C
    Francart, C
    Cajot, MA
    Brevière, GM
    Vaksmann, G
    Devos, P
    Coullet, JM
    Vincentelli, A
    CARDIOLOGY IN THE YOUNG, 2000, 10 (05) : 527 - 533
  • [10] An 84-year-old patient with an atrial septal defect rescued using the Amplatzer septal occluder device
    Sato, Tetsuya
    Hioka, Toru
    Akagi, Teiji
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (05) : 758 - 762