Backscatter dose effects for high atomic number materials being irradiated in the presence of a magnetic field: A Monte Carlo study for the MRI linac

被引:8
作者
Ahmad, Syed Bilal [1 ,2 ]
Sarfehnia, Arman [2 ,3 ]
Kim, Anthony [2 ,3 ]
Wronski, Matt [2 ,3 ]
Sahgal, Arjun [2 ,3 ]
Keller, Brian M. [2 ,3 ]
机构
[1] Sunnybrook Res Inst, 2075 Bayview Ave, Toronto, ON M4N 3M5, Canada
[2] Sunnybrook Hlth Sci Ctr, Odette Canc Ctr, 2075 Bayview Ave, Toronto, ON M4N 3M5, Canada
[3] Univ Toronto, Dept Radiat Oncol, 27 Kings Coll Circle, Toronto, ON M5S 1A1, Canada
关键词
backscatter; MRI linac; inhomogeneities; interface effects; Monte Carlo; MV RADIOTHERAPY ACCELERATOR; PHOTON BEAMS; SURFACE ORIENTATION; GEANT4; INTERFACES; PERTURBATION; PHYSICS; ARTHROPLASTY; TRANSPORT; ENERGIES;
D O I
10.1118/1.4955175
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To quantify and explain the backscatter dose effects for clinically relevant high atomic number materials being irradiated in the presence of a 1.5 T transverse magnetic field. Methods: Interface effects were investigated using Monte Carlo simulation techniques. We used GPUMCD (v5.1) and GEANT4 (V10.1) for this purpose. GPUMCD is a commercial software written for the Elekta AB, MRI linac. Dose was scored using GPUMCD in cubic voxels of side 1 and 0 5 mm, in two different virtual phantoms of dimensions 20 x 20 x 20 cm and 5 x 5 x 13.3 cm, respectively. A photon beam was generated from a point 143.5 cm away from the isocenter with energy distribution sampled from a histogram representing the true Elekta, MRI linac photon spectrum. A slab of variable thickness and position containing either bone, aluminum, titanium, stainless steel, or one of the two different dental filling materials was inserted as an inhomogeneity in the 20 x 20 x 20 cm phantom. The 5 x 5 x 13.3 cm phantom was used as a clinical test case in order to explain the dose perturbation effects for a head and neck cancer patient. The back scatter dose factor (BSDF) was defined as the ratio of the doses at a given depth with and without the presence of the inhomogeneity. Backscattered electron fluence was calculated at the inhomogeneity interface using GEANT4. A 1.5 T magnetic field was applied perpendicular to the direction of the beam in both phantoms, identical to the geometry in the Elekta MRI linac. Results: With the application of a 1.5 T magnetic field, all the BSDF's were reduced by 12%-47%, compared to the no magnetic field case. The corresponding backscattered electron fluence at the interface was also reduced by 45%-64%. The reduction in the BSDF at the interface, due to the application of the magnetic field, is manifested in a different manner for each material. In the case of bone, the dose drops at the interface contrary to the expected increase when no magnetic field is applied. In the case of aluminum, the dose at the interface is the same with and without the presence of the aluminum. For all of the other materials the dose increases at the interface. Conclusions: The reduction in dose at the interface, in the presence of the magnetic field, is directly related to the reduction in backscattered electron fluence. This reduction occurs due to two different reasons. First, the electron spectrum hitting the interface is changed when the magnetic field is turned on, which results in changes in the electron scattering probability. Second, some electrons that have curved trajectories due to the presence of the magnetic field are absorbed by the higher density side of the interface and no longer contribute to the backscattered electron fluence. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:4665 / 4673
页数:9
相关论文
共 33 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   Evaluation of a commercial MRI Linac based Monte Carlo dose calculation algorithm with GEANT4 [J].
Ahmad, Syed Bilal ;
Sarfehnia, Arman ;
Paudel, Moti Raj ;
Kim, Anthony ;
Hissoiny, Sami ;
Sahgal, Arjun ;
Keller, Brian .
MEDICAL PHYSICS, 2016, 43 (02) :894-907
[3]   Comparison of Geant4 electromagnetic physics models against the NIST reference data [J].
Amako, K ;
Guatelli, S ;
Ivanchenko, VN ;
Maire, M ;
Mascialino, B ;
Murakami, K ;
Nieminen, P ;
Pandola, L ;
Parlati, S ;
Pia, MG ;
Piergentili, M ;
Sasaki, T ;
Urban, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (04) :910-918
[4]   Geometry and physics of the Geant4 toolkit for high and medium energy applications [J].
Apostolakis, J. ;
Asai, M. ;
Bogdanov, A. G. ;
Burkhardt, H. ;
Cosmo, G. ;
Elles, S. ;
Folger, G. ;
Grichine, V. M. ;
Gumplinger, P. ;
Heikkinen, A. ;
Hrivnacova, I. ;
Ivanchenko, V. N. ;
Jacquemier, J. ;
Koi, T. ;
Kokoulin, R. P. ;
Kossov, M. ;
Kurashige, H. ;
McLaren, I. ;
Link, O. ;
Maire, M. ;
Pokorski, W. ;
Sasaki, T. ;
Starkov, N. ;
Urban, L. ;
Wright, D. H. .
RADIATION PHYSICS AND CHEMISTRY, 2009, 78 (10) :859-873
[5]   A deterministic solution of the first order linear Boltzmann transport equation in the presence of external magnetic fields [J].
Aubin, J. St. ;
Keyvanloo, A. ;
Vassiliev, O. ;
Fallone, B. G. .
MEDICAL PHYSICS, 2015, 42 (02) :780-793
[6]   Effect of dental restorations and prostheses on radiotherapy dose distribution: a Monte Carlo study [J].
Chin, David W. H. ;
Treister, Nathaniel ;
Friedland, Bernard ;
Cormack, Robert A. ;
Tishler, Roy B. ;
Makrigiorgos, G. Mike ;
Court, Laurence E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (01) :80-89
[7]   Metal Artifact Reduction With MAVRIC SL at 3-T MRI in Patients With Hip Arthroplasty [J].
Choi, Soo-Jung ;
Koch, Kevin M. ;
Hargreaves, Brian A. ;
Stevens, Kathryn J. ;
Gold, Garry E. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (01) :140-147
[8]   BACKSCATTER DOSE PERTURBATION IN KILOVOLTAGE PHOTON BEAMS AT HIGH ATOMIC-NUMBER INTERFACES [J].
DAS, IJ ;
CHOPRA, KL .
MEDICAL PHYSICS, 1995, 22 (06) :767-773
[9]   BACKSCATTER DOSE PERTURBATION AT HIGH ATOMIC-NUMBER INTERFACES IN MEGAVOLTAGE PHOTON BEAMS [J].
DAS, IJ ;
KAHN, FM .
MEDICAL PHYSICS, 1989, 16 (03) :367-375
[10]   INTERFACE DOSE PERTURBATION AS A MEASURE OF MEGAVOLTAGE PHOTON-BEAM ENERGY [J].
DAS, IJ ;
KHAN, FM ;
GERBI, BJ .
MEDICAL PHYSICS, 1988, 15 (01) :78-81