Using Gas Modifiers to Significantly Improve Sensitivity and Selectivity in a Cylindrical FAIMS Device

被引:40
作者
Purves, Randy W. [1 ]
Ozog, Allison R. [1 ]
Ambrose, Stephen J. [1 ]
Prasad, Satendra [2 ]
Belford, Michael [2 ]
Dunyach, Jean-Jacques [2 ]
机构
[1] Natl Res Council Canada, Saskatoon, SK, Canada
[2] Thermo Fisher Sci, San Jose, CA USA
关键词
FAIMS; DMS; Gas additives; Gas-phase separation; Electrospray; DIFFERENTIAL MOBILITY SPECTROMETRY; CHROMATOGRAPHY-MASS SPECTROMETRY; PER-TRILLION LEVELS; SEPARATION; ACIDS; IONS;
D O I
10.1007/s13361-014-0878-z
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Recent reports describing enhanced performance when using gas additives in a DMS device (planar electrodes) have indicated that comparable benefits are not attainable using FAIMS (cylindrical electrodes), owing to the non-homogeneous electric fields within the analyzer region. In this study, a FAIMS system (having cylindrical electrodes) was modified to allow for controlled delivery of gas additives. An experiment was carried out that illustrates the important distinction between gas modifiers present as unregulated contaminants and modifiers added in a controlled manner. The effect of contamination was simulated by adjusting the ESI needle position to promote incomplete desolvation, thereby permitting ESI solvent vapor into the FAIMS analyzer region, causing signal instability and irreproducible CV values. However, by actively controlling the delivery of the gas modifier, reproducible CV spectra were obtained. The effects of adding different gas modifiers were examined using 15 positive ions having mass-to-charge (m/z) values between 90 and 734. Significant improvements in peak capacity and increases in ion transmission were readily attained by adding acetonitrile vapor, even at trace levels (a parts per thousand currency sign0.1%). Increases in signal intensity were greatest for the low m/z ions; for the six lowest molecular weight species, signal intensities increased by similar to 10- to over 100-fold compared with using nitrogen without gas additives, resulting in equivalent or better signal intensities compared with ESI without FAIMS. These results confirm that analytical benefits derived from the addition of gas modifiers reported with a uniform electric field (DMS) also are observed using a non-homogenous electric field (FAIMS) in the analyser region.
引用
收藏
页码:1274 / 1284
页数:11
相关论文
共 34 条
[1]   Evaluation of carrier gases for use in high-field asymmetric waveform ion mobility spectrometry [J].
Barnett, DA ;
Ells, B ;
Guevremont, R .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2000, 11 (12) :1125-1133
[2]  
Barnett DA, 2000, J MASS SPECTROM, V35, P976, DOI 10.1002/1096-9888(200008)35:8<976::AID-JMS25>3.0.CO
[3]  
2-Q
[4]   Determination of parts-per-trillion levels of chlorate, bromate, and iodate by electrospray ionization/high-field asymmetric waveform ion mobility spectrometry/mass spectrometry [J].
Barnett, DA ;
Guevremont, R ;
Purves, RW .
APPLIED SPECTROSCOPY, 1999, 53 (11) :1367-1374
[5]   Application of ESI-FAIMS-MS to the analysis of tryptic peptides [J].
Barnett, DA ;
Ells, B ;
Guevremont, R ;
Purves, RW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2002, 13 (11) :1282-1291
[6]   Characterization of a temperature-controlled FAIMS system [J].
Barnett, David A. ;
Belford, Michael ;
Dunyach, Jean-Jacques ;
Purves, Randy W. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (09) :1653-1663
[7]  
Belford M., 2013, P 61 ASMS C MASS SPE
[8]   A NEW METHOD OF SEPARATION OF MULTI-ATOMIC IONS BY MOBILITY AT ATMOSPHERIC-PRESSURE USING A HIGH-FREQUENCY AMPLITUDE-ASYMMETRIC STRONG ELECTRIC-FIELD [J].
BURYAKOV, IA ;
KRYLOV, EV ;
NAZAROV, EG ;
RASULEV, UK .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1993, 128 (03) :143-148
[9]  
Eiceman GA, 2014, ION MOBILITY SPECTROMETRY, 3RD EDITION, P119
[10]   Separation of ions from explosives in differential mobility spectrometry by vapor-modified drift gas [J].
Eiceman, GA ;
Krylov, EV ;
Nazarov, EG ;
Miller, RA .
ANALYTICAL CHEMISTRY, 2004, 76 (17) :4937-4944