Lead extraction outcomes in patients with congenital heart disease

被引:14
作者
Fender, Erin A. [1 ]
Killu, Ammar M. [1 ]
Cannon, Bryan C. [1 ]
Friedman, Paul A. [1 ]
Mcleod, Christopher J. [1 ]
Hodge, David O. [1 ]
Broberg, Craig S. [2 ]
Henrikson, Charles A. [2 ]
Cha, Yong-Mei [1 ]
机构
[1] Mayo Clin, St Marys Campus,Mail Code RO MB 04 506CAR, Rochester, MN 55902 USA
[2] Oregon Hlth & Sci Univ, Knight Cardiovasc Inst, Mail Code UHN 62 3181 SW Sam Jackson Pk Rd, Portland, OR 97239 USA
来源
EUROPACE | 2017年 / 19卷 / 03期
关键词
Cardiac implantable electronic device; Lead extraction; Laser lead extraction; Congenital heart disease; GO CURRENT CONTROVERSIES; YOUNG-ADULTS; CHILDREN; COMPLICATIONS; MANAGEMENT; STAY;
D O I
10.1093/europace/euw049
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Patients with congenital heart disease (CHD) are at increased risk for intracardiac device malfunction and infection that may necessitate extraction; however, the risk of extraction is poorly understood. This study addresses the safety of extraction in patients with structural heart disease and previous cardiac surgery. Methods and results This retrospective study included 40 CHD and 80 matched control patients, who underwent transvenous lead extractions between 2001 and 2014. Only leads >12 months were included. There were 77 leads in CHD patients and 146 in controls. The mean age was 38 +/- 16 years in CHD patients. Ninety per cent of CHD patients had >= 1 cardiac surgeries when compared with 21% of controls (P < 0.001). The number of abandoned leads was significantly different (17 vs. 3, P < 0.001). Lead age was similar with an average duration of 83 6 87 months in CHD patients and 62 6 65 months in controls (P = 0.24). There was no significant difference in extraction techniques. Manual traction was successful in 40% of CHD patients and 47% of controls, and advanced techniques were used in 60 and 53% of CHD patients and controls, respectively. Complete extraction was achieved in 94% of the patients in both groups. There was no significant difference in complications. Conclusion Lead extraction can be safely performed in patients with CHD. Despite anatomic abnormalities and longer implantation times, the difficulty of lead extraction in patients with CHD is comparable with controls.
引用
收藏
页码:441 / 446
页数:6
相关论文
共 21 条
  • [1] Multi-Institutional Study of Implantable Defibrillator Lead Performance in Children and Young Adults: Results of the Pediatric Lead Extractability and Survival Evaluation (PLEASE) Study
    Atallah, Joseph
    Erickson, Christopher C.
    Cecchin, Frank
    Dubin, Anne M.
    Law, Ian H.
    Cohen, Mitchell I.
    LaPage, Martin J.
    Cannon, Bryan C.
    Chun, Terrence U. H.
    Freedenberg, Vicki
    Gierdalski, Marcin
    Berul, Charles I.
    [J]. CIRCULATION, 2013, 127 (24) : 2393 - 2402
  • [2] How adequate are the current methods of lead extraction? A review of the efficiency and safety of transvenous lead extraction methods
    Buiten, Maurits S.
    van der Heijden, Aafke C.
    Schalij, Martin J.
    van Erven, Lieselot
    [J]. EUROPACE, 2015, 17 (05): : 689 - 700
  • [3] Lead Extraction in Pediatric and Congenital Heart Disease Patients
    Cecchin, Frank
    Atallah, Joseph
    Walsh, Edward P.
    Triedman, John K.
    Alexander, Mark E.
    Berul, Charles I.
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (05) : 437 - 444
  • [4] Implantable cardioverter defibrillator lead complications and laser extraction in children and young adults with congenital heart disease: Implications for implantation and management
    Cooper, JM
    Stephenson, EA
    Berul, CI
    Walsh, EP
    Epstein, LM
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (04) : 344 - 349
  • [5] Lead Extraction in Children and Young Adults Using Different Techniques
    Dilber, Embiya
    Karagoz, Tevfik
    Celiker, Alpay
    [J]. MEDICAL PRINCIPLES AND PRACTICE, 2009, 18 (05) : 356 - 359
  • [6] Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease
    Fortescue, EB
    Berul, CI
    Cecchin, F
    Walsh, EP
    Triedman, JK
    Alexander, ME
    [J]. HEART RHYTHM, 2004, 1 (02) : 150 - 159
  • [7] Device complications in adult congenital heart disease
    Hayward, Robert M.
    Dewland, Thomas A.
    Moyers, Brian
    Vittinghoff, Eric
    Tanel, Ronn E.
    Marcus, Gregory M.
    Tseng, Zian H.
    [J]. HEART RHYTHM, 2015, 12 (02) : 338 - 344
  • [8] How to prevent, recognize, and manage complications of lead extraction. Part III: Procedural factors
    Henrikson, Chares A.
    Brinker, Jeffrey A.
    [J]. HEART RHYTHM, 2008, 5 (09) : 1352 - 1354
  • [9] Should They Stay or Should They Go? Current Controversies in Lead Extraction Think Before You Pull-Not Every Lead Has to Come Out
    Henrikson, Charles A.
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (04) : 409 - 412
  • [10] Laser lead extraction in adult congenital heart disease
    Khairy, Paul
    Roux, Jean-Francois
    Dubuc, Marc
    Thibault, Bernard
    Guerra, Peter G.
    Macle, Laurent
    Mercier, Lise-Andree
    Dore, Annie
    Roy, Denis
    Talajic, Mario
    Page, Pierre