A concept for a 3D-printed patient-specific stereotaxy platform for brain biopsy -a canine cadaver study-

被引:13
作者
Mueller, Marcel [1 ]
Winkler, Dirk [2 ]
Moebius, Robert [2 ]
Sauerstein, Tobias [1 ]
Scholz, Sebastian [1 ]
Gutrnann, Sarah [3 ]
Flegel, Thomas [3 ]
Meixensberger, Juergen [2 ]
Drossel, Welf-Guntram [1 ]
Grunert, Ronny [1 ]
机构
[1] Fraunhofer Inst Machine Tools & Forming Technol, Nothnitzer Str 44, D-01187 Dresden, Germany
[2] Univ Clin Leipzig, Fac Med, Dept Neurosurg, Liebigstr 20, D-04103 Leipzig, Germany
[3] Univ Leipzig, Fac Vet Med, Dept Small Anim Med, Tierkliniken 23, D-04103 Leipzig, Germany
关键词
Brain biopsy; 3D print; Canine cadaver; Stereotactic frame; ACCURACY; SYSTEM; DOGS;
D O I
10.1016/j.rvsc.2019.02.007
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The treatment of intracranial lesions requires a precise diagnosis with subsequent identification of an adequate therapeutic approach. Stereotactic tumor biopsy may be considered the safest neurosurgical procedure in terms of anticipated results and potential surgical complications. The aim of the present paper was to demonstrate a new method of stereotactic biopsy, based on a patient-specific 3D printed platform in dogs. The system was tested on two canine cadavers, a small (Shih Tzu) and a large (Labrador) breed. Imaginary biopsy targets were defined in a superficial (caudate nucleus) and a deep (piriform lobe) position. Based on 3 Tesla MRI, individualized stereotactic platforms were designed using a semi-automatic approach, and manufactured additively using ABS M30. A pre- and intra-operative CT was performed to compare the planned vs. the realized needle position for precision analyses of the procedure. The target points varied with a precision between 0.09 mm and 0.48 mm. Manufacturing time required 480 to 700 min per platform. The presented patient-specific stereotactic system seems a suitable instrument for application in small animal neurosurgery. In particular, the implementation of relevant stereotactic data may help performing the procedure in rapid sequence and with higher precision than currently-used systems. Required adjustments and adaptions to the respective anatomical conditions are omitted and make the procedure reliable and safe.
引用
收藏
页码:79 / 84
页数:6
相关论文
共 20 条
[1]   Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation [J].
Bjartmarz, Hjalmar ;
Rehncrona, Stig .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2007, 85 (05) :235-242
[2]   DESCRIPTION AND VALIDATION OF A MAGNETIC RESONANCE IMAGING-GUIDED STEREOTACTIC BRAIN BIOPSY DEVICE IN THE DOG [J].
Chen, Annie V. ;
Wininger, Fred A. ;
Frey, Stephen ;
Comeau, Roch M. ;
Bagley, Rodney S. ;
Tucker, Russell L. ;
Schneider, Adam R. ;
Gay, John M. .
VETERINARY RADIOLOGY & ULTRASOUND, 2012, 53 (02) :150-156
[3]   Frameless stereotactic neurosurgery: Two steps towards the holy grail of surgical navigation [J].
Eljamel, MS .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1999, 72 (2-4) :125-128
[4]   SonoWand, an ultrasound-based neuronavigation system [J].
Gronningsaeter, A ;
Kleven, A ;
Ommedal, S ;
Aarseth, TE ;
Lie, T ;
Lindseth, F ;
Lango, T ;
Unsgard, G .
NEUROSURGERY, 2000, 47 (06) :1373-1379
[5]   An analysis of stereotactic biopsy of brain tumors and nonneoplastic lesions: a prospective clinicopathologic study [J].
Heper, AO ;
Erden, E ;
Savas, A ;
Ceyhan, K ;
Erden, I ;
Akyar, S ;
Kanpolat, Y .
SURGICAL NEUROLOGY, 2005, 64 :82-88
[6]   Novel MRI and CT compatible stereotactic brain biopsy system in dogs using patient-specific facemasks [J].
James, M. D. ;
Bova, F. J. ;
Rajon, D. A. ;
Carrera-Justiz, S. ;
Clemmons, R. M. .
JOURNAL OF SMALL ANIMAL PRACTICE, 2017, 58 (11) :615-621
[7]   Single fraction stereotactic radiation therapy (stereotactic radiosurgery) is a feasible method for treating intracranial meningiomas in dogs [J].
Kelsey, Krista L. ;
Gieger, Tracy L. ;
Nolan, Michael W. .
VETERINARY RADIOLOGY & ULTRASOUND, 2018, 59 (05) :632-638
[8]   Computer-assisted thoracic pedicle screw placement -: An in vitro feasibility study [J].
Kim, KD ;
Johnson, JP ;
Bloch, O ;
Masciopinto, JE .
SPINE, 2001, 26 (04) :360-364
[9]   Modification and application of a Pelorus Mark III Stereotactic System for CT-guided brain biopsy in 50 dogs [J].
Koblik, PD ;
LeCouteur, RA ;
Higgins, RJ ;
Fick, J ;
Kortz, GD ;
Sturges, BK ;
Pascoe, PJ .
VETERINARY RADIOLOGY & ULTRASOUND, 1999, 40 (05) :424-433
[10]  
Koch-Institut R, 2019, KREBS DEUTSCHLAND 20, V12, DOI [10.17886/rkipubl-2017-007, DOI 10.17886/RKIPUBL-2017-007]