Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA

被引:49
作者
Chaswal, Vibha [1 ]
Weldon, Michael [1 ]
Gupta, Nilendu [1 ]
Chakravarti, Arnab [1 ]
Rong, Yi [1 ]
机构
[1] Ohio State Univ, Dept Radiat Oncol, Columbus, OH 43221 USA
关键词
volumetric-modulated arc therapy; ArcCHECK diode array; quality assurance; angular dependence; ArcCHECK commissioning; ArcCHECK evaluation; ANISOTROPIC ANALYTICAL ALGORITHM; PHOTON DOSE CALCULATION; DOSIMETRIC VERIFICATION; QUALITY-ASSURANCE; QUANTITATIVE-EVALUATION; MODULATED ARC; IMRT; RADIOTHERAPY; WATER; DISTRIBUTIONS;
D O I
10.1120/jacmp.v15i4.4832
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present commissioning and comprehensive evaluation for ArcCHECK as a QA equipment for volumetric-modulated arc therapy (VMAT), using the 6 MV photon beam with and without the flattening filter, and the SNC patient software (version 6.2). In addition to commissioning involving absolute dose calibration, array calibration, and PMMA density verification, ArcCHECK was evaluated for its response dependency on linac dose rate, instantaneous dose rate, radiation field size, beam angle, and couch insertion. Scatter dose characterization, consistency and symmetry of response, and dosimetry accuracy evaluation for fixed aperture arcs and clinical VMAT patient plans were also investigated. All the evaluation tests were performed with the central plug inserted and the homogeneous PMMA density value. Results of gamma analysis demonstrated an overall agreement between ArcCHECK-measured and TPS-calculated reference doses. The diode based field size dependency was found to be within 0.5% of the reference. The dose rate-based dependency was well within 1% of the TPS reference, and the angular dependency was found to be +/- 3% of the reference, as tested for BEV angles, for both beams. Dosimetry of fixed arcs, using both narrow and wide field widths, resulted in clinically acceptable global gamma passing rates on the 3%/3 mm level and 10% threshold. Dosimetry of narrow arcs showed an improvement over published literature. The clinical VMAT cases demonstrated high level of dosimetry accuracy in gamma passing rates.
引用
收藏
页码:212 / 225
页数:14
相关论文
共 25 条
[1]  
[Anonymous], 2009, J APPL CLIN MED PHYS
[2]   Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning [J].
Bragg, Christopher M. ;
Conway, John .
RADIOTHERAPY AND ONCOLOGY, 2006, 81 (03) :315-323
[3]   Basic dosimetric verification in water of the anisotropic analytical algorithm for Varian, Elekta and Siemens linacs [J].
Cozzi, Luca ;
Nicolini, Giorgia ;
Vanetti, Eugenio ;
Clivio, Alessandro ;
Glashoerster, Marco ;
Schiefer, Hans ;
Fogliata, Antonella .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2008, 18 (02) :128-135
[4]   A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation [J].
Depuydt, T ;
Van Esch, A ;
Huyskens, DP .
RADIOTHERAPY AND ONCOLOGY, 2002, 62 (03) :309-319
[5]   Evaluation of a new VMAT QA device, or the "X" and "O" array geometries [J].
Feygelman, Vladimir ;
Zhang, Geoffrey ;
Stevens, Craig ;
Nelms, Benjamin E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :146-168
[6]   Dosimetric validation of the anisotropic analytical algorithm for photon dose calculation: fundamental characterization in water [J].
Fogliata, A ;
Nicolini, G ;
Vanetti, E ;
Clivio, A ;
Cozzi, L .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06) :1421-1438
[7]   Evaluation of the analytical anisotropic algorithm in an extreme water-lung interface phantom using Monte Carlo dose calculations [J].
Gagne, Isabelle Marie ;
Zavgorodni, Sergei .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (01) :33-46
[8]   Collapsed cone convolution and analytical anisotropic algorithm dose calculations compared to VMC plus plus Monte Carlo simulations in clinical cases [J].
Hasenbalg, F. ;
Neuenschwander, H. ;
Mini, R. ;
Born, E. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13) :3679-3691
[9]   Accuracy of out-of-field dose calculations by a commercial treatment planning system [J].
Howell, Rebecca M. ;
Scarboro, Sarah B. ;
Kry, S. F. ;
Yaldo, Derek Z. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) :6999-7008
[10]   COMMISSIONING OF PHOTON BEAMS OF A FLATTENING FILTER-FREE LINEAR ACCELERATOR AND THE ACCURACY OF BEAM MODELING USING AN ANISOTROPIC ANALYTICAL ALGORITHM [J].
Hrbacek, Jan ;
Lang, Stephanie ;
Kloeck, Stephan .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (04) :1228-1237