Influence of daily imaging on plan quality and normal tissue toxicity for prostate cancer radiotherapy

被引:20
作者
Bell, Katharina [1 ]
Heitfeld, Marina [1 ]
Licht, Norbert [1 ]
Ruebe, Christian [1 ]
Dzierma, Yvonne [1 ]
机构
[1] Univ Saarland, Med Ctr, Dept Radiotherapy & Radiat Oncol, Kirrberger Str,Geb 6-5, D-66421 Homburg, Germany
来源
RADIATION ONCOLOGY | 2017年 / 12卷
关键词
Imaging dose distribution; Linac-based imaging; IGRT; NTCP modelling; Kilovoltage imaging; Megavoltage imaging; Prostate; BEAM COMPUTED-TOMOGRAPHY; SET-UP ERRORS; GUIDED RADIOTHERAPY; RADIATION-THERAPY; CARBON TARGET; PTV MARGIN; SYSTEM; CT; PROBABILITY; IMPLEMENTATION;
D O I
10.1186/s13014-016-0757-9
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Modern radiotherapy offers various possibilities for image guided verification of patient positioning. Different clinically relevant IGRT (image guided radiotherapy) scenarios were considered with regard to their influence on dosimetric plan quality and normal tissue complication probability (NTCP). Methods: This study is based on treatment plans of 50 prostate patients. We evaluate the clinically performed IGRT and simulate the influence of different daily IGRT scenarios on plan quality. Imaging doses of planar and cone-beam-CT (CBCT) images for three different energies (6 MV, 1 MV and 121 kV) were added to the treatment plans. The plan quality of the different scenarios was assessed by a visual inspection of the dose distribution and dose-volume-histogram (DVH) and a statistical analysis of DVH criteria. In addition, an assessment of the normal tissue complication probability was performed. Results: Daily 1MV-CBCTs result in undesirable high dose regions in the target volume. The DVH shows that the scenarios with actual imaging performed, daily kV-CBCT and daily 6MV imaging (1x CBCT, 4x planar images per week) do not differ exceedingly from the original plan; especially imaging with daily kV-CBCT has little influence to the sparing of organs at risk. In contrast, daily 1MV-CBCT entails an additional dose of up to two fraction doses. Due to the additional dose amount some DVH constraints for plan acceptability could no longer be satisfied, especially for the daily 1MV-CBCT scenario. This scenario also shows increased NTCP for the rectum. Conclusion: Daily kV-CBCT has negligible influence on plan quality and is commendable for the clinical routine. If no kV-modality is available, a daily IGRT scenario with one CBCT per week and planar axial images on the other days should be preferred over daily MV-CBCT.
引用
收藏
页数:11
相关论文
共 32 条
  • [1] MEGAVOLTAGE CONE BEAM COMPUTED TOMOGRAPHY DOSE AND THE NECESSITY OF REOPTIMIZATION FOR IMAGING DOSE-INTEGRATED INTENSITY-MODULATED RADIOTHERAPY FOR PROSTATE CANCER
    Akino, Yuichi
    Koizumi, Masahiko
    Sumida, Iori
    Takahashi, Yutaka
    Ogata, Toshiyuki
    Ota, Seiichi
    Isohashi, Fumiaki
    Konishi, Koji
    Yoshioka, Yasuo
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05): : 1715 - 1722
  • [2] Imaging dose from cone beam computed tomography in radiation therapy
    Alaei, Parham
    Spezi, Emiliano
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (07): : 647 - 658
  • [3] Dose calculation and treatment plan optimization including imaging dose from kilovoltage cone beam computed tomography
    Alaei, Parham
    Spezi, Emiliano
    Reynolds, Margaret
    [J]. ACTA ONCOLOGICA, 2014, 53 (06) : 839 - 844
  • [4] Commissioning kilovoltage cone-beam CT beams in a radiation therapy treatment planning system
    Alaei, Parham
    Spezi, Emiliano
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (06): : 19 - 33
  • [5] Inclusion of the dose from kilovoltage cone beam CT in the radiation therapy treatment plans
    Alaei, Parham
    Ding, George
    Guan, Huaiqun
    [J]. MEDICAL PHYSICS, 2010, 37 (01) : 244 - 248
  • [6] Imaging doses from the Elekta Synergy X-ray cone beam CT system
    Amer, A.
    Marchant, T.
    Sykes, J.
    Czajka, J.
    Moore, C.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (954) : 476 - 482
  • [7] [Anonymous], THESIS
  • [8] Image-guided radiotherapy for prostate cancer with cone beam CT: dosimetric effects of imaging frequency and PTV margin
    Ariyaratne, Hemal
    Chesham, Hayley
    Pettingell, John
    Alonzi, Roberto
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 121 (01) : 103 - 108
  • [9] Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT
    Beltran, Chris
    Lukose, Renin
    Gangadharan, Bijumon
    Bani-Hashemi, Ali
    Faddegon, Bruce A.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (03): : 37 - 48
  • [10] THE BENEFITS OF INCLUDING CLINICAL FACTORS IN RECTAL NORMAL TISSUE COMPLICATION PROBABILITY MODELING AFTER RADIOTHERAPY FOR PROSTATE CANCER
    Defraene, Gilles
    Van den Bergh, Laura
    Al-Mamgani, Abrahim
    Haustermans, Karin
    Heemsbergen, Wilma
    Van den Heuvel, Frank
    Lebesque, Joos V.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (03): : 1233 - 1242