Predictive wear modeling of the articulating metal-on-metal hip replacements

被引:17
|
作者
Gao, Leiming [1 ]
Dowson, Duncan [2 ]
Hewson, Robert W. [1 ]
机构
[1] Imperial Coll London, Dept Aeronaut, London SW7 2AZ, England
[2] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
关键词
wear model; elastohydrodynamic lubrication (EHL); metal-on-metal total hip joint replacement; Archard law; numerical simulation; ELASTOHYDRODYNAMIC LUBRICATION; JOINT SIMULATOR; CONTACT MECHANICS; IMPLANTS; PERFORMANCE; PARTICLES; INCLINATION; PRESSURE; BEARINGS; DIAMETER;
D O I
10.1002/jbm.b.33568
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The lubrication regime in which artificial hip joints operate adds complexity to the prediction of wear, as the joint operates in both the full fluid film regimespecifically the elastohydrodynamic lubrication (EHL) regimeand the mixed or boundary lubrication regimes, where contact between the bearing surfaces results in wear. In this work, a wear model is developed which considers lubrication for the first time via a transient EHL model of metal-on-metal hip replacements. This is a framework to investigate how the change in film thickness influences the wear, which is important to further investigation of the complex wear procedure, including tribocorrosion, in the lubricated hip implants. The wear model applied here is based on the work of Sharif et al. who adapted the Archard wear law by making the wear rate a function of a relative film thickness nominalized by surface roughness for examining wear of industrial gears. In this work, the gait cycle employed in hip simulator tests is computationally investigated and wear is predicted for two sizes of metal-on-metal total hip replacements. The wear results qualitatively predict the typical wear curve obtained from experimental hip simulator tests, with an initial running-in period before a lower wear rate is reached. The shape of the wear scar has been simulated on both the acetabular cup and the femoral head bearing surfaces. (c) 2015 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 105B: 497-506, 2017.
引用
收藏
页码:497 / 506
页数:10
相关论文
共 50 条
  • [21] The Outcome of Revision Surgery for Failed Metal-on-Metal Total Hip Arthroplasty
    Borton, Zakk M.
    Mumith, Aadil S.
    Nicholls, Alex J.
    Pearce, Andrea R.
    Briant-Evans, Toby W.
    Griffiths, Jamie T.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (08) : 1749 - 1754
  • [22] Effect of walking patterns on the elastohydrodynamic lubrication of metal-on-metal total hip replacements
    Gao, L.
    Fisher, J.
    Tin, Z.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2011, 225 (J6) : 515 - 525
  • [23] Outcome of revision arthroplasty for failed metal-on-metal total hip replacements; is there a relation with metal ions?
    Iqbal, Hafiz J.
    Al-Azzani, Waheeb A. K.
    Jackson-Taylor, Esther
    Clatworthy, Elizabeth
    John, Alun
    HIP INTERNATIONAL, 2017, 27 (03) : 235 - 240
  • [24] Analysis of Run-in-Stage Wear Behavior and Contact Mechanics of Metal-on-Metal Hip Joint Bearings with Different Radial Clearances
    Chen, Yan
    Li, Yunping
    Kurosu, Shingo
    Meng, Qingen
    Tang, Ning
    Koizumi, Yuichiro
    Chiba, Akihiko
    MATERIALS TRANSACTIONS, 2015, 56 (06) : 826 - 834
  • [25] Predicting Wear and Blood Metal Ion Levels in Metal-on-Metal Hip Resurfacing
    Matthies, Ashley K.
    Henckel, Johann
    Cro, Suzie
    Suarez, Alexander
    Noble, Philip C.
    Skinner, John
    Hart, Alister J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2014, 32 (01) : 167 - 174
  • [26] Assessing the material loss of the modular taper interface in retrieved metal-on-metal hip replacements
    Bills, Paul J.
    Racasan, R.
    Tessier, P.
    Blunt, L. A.
    SURFACE TOPOGRAPHY-METROLOGY AND PROPERTIES, 2015, 3 (02):
  • [27] Wear of ceramic coated metal-on-metal bearings used for hip replacement
    Gispert, M. P.
    Serro, A. P.
    Colaco, R.
    Pires, E.
    Saramago, B.
    WEAR, 2007, 263 : 1060 - 1065
  • [28] Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses
    Lord, J. K.
    Langton, D. J.
    Nargol, A. V. F.
    Joyce, T. J.
    WEAR, 2011, 272 (01) : 79 - 87
  • [29] Heart failure after conventional metal-on-metal hip replacements A retrospective cohort study
    Gillam, Marianne H.
    Pratt, Nicole L.
    Inacio, Maria C. S.
    Roughead, Elizabeth E.
    Shakib, Sepehr
    Nicholls, Stephen J.
    Graves, Stephen E.
    Acta Orthopaedica, 2017, 88 (01) : 2 - 9
  • [30] Characterization of metal-wear nanoparticles in pseudotumor following metal-on-metal hip resurfacing
    Xia, Zhidao
    Kwon, Young-Min
    Mehmood, Shahid
    Downing, Clive
    Jurkschat, Kerstin
    Murray, David W.
    NANOMEDICINE-NANOTECHNOLOGY BIOLOGY AND MEDICINE, 2011, 7 (06) : 674 - 681