Improved Accuracy with 3D Planning and Patient-Specific Instruments During Simulated Pelvic Bone Tumor Surgery

被引:135
作者
Cartiaux, Olivier [1 ]
Paul, Laurent [1 ]
Francq, Bernard G. [2 ]
Banse, Xavier [1 ,3 ]
Docquier, Pierre-Louis [1 ,3 ]
机构
[1] UCL, IREC, B-1200 Brussels, Belgium
[2] Catholic Univ Louvain, Inst Stat Biostat & Sci Actuarielles, B-1348 Louvain, Belgium
[3] Clin Univ St Luc, Serv Orthope & Traumatol Appareil Locomoteur, B-1200 Brussels, Belgium
关键词
Tumor resection; Safe margin; Patient-specific cutting guides; Quantitative surgery; ISO-based evaluation; ASSISTED ORTHOPEDIC-SURGERY; TOTAL KNEE ARTHROPLASTY; RESECTION; RECONSTRUCTION; OSTEOSARCOMA; PERSPECTIVES; TEMPLATES; OSTEOTOMY; ALIGNMENT;
D O I
10.1007/s10439-013-0890-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In orthopaedic surgery, resection of pelvic bone tumors can be inaccurate due to complex geometry, limited visibility and restricted working space of the pelvis. The present study investigated accuracy of patient-specific instrumentation (PSI) for bone-cutting during simulated tumor surgery within the pelvis. A synthetic pelvic bone model was imaged using a CT-scanner. The set of images was reconstructed in 3D and resection of a simulated periacetabular tumor was defined with four target planes (ischium, pubis, anterior ilium, and posterior ilium) with a 10-mm desired safe margin. Patient-specific instruments for bone-cutting were designed and manufactured using rapid-prototyping technology. Twenty-four surgeons (10 senior and 14 junior) were asked to perform tumor resection. After cutting, ISO1101 location and flatness parameters, achieved surgical margins and the time were measured. With PSI, the location accuracy of the cut planes with respect to the target planes averaged 1 and 1.2 mm in the anterior and posterior ilium, 2 mm in the pubis and 3.7 mm in the ischium (p < 0.0001). Results in terms of the location of the cut planes and the achieved surgical margins did not reveal any significant difference between senior and junior surgeons (p = 0.2214 and 0.8449, respectively). The maximum differences between the achieved margins and the 10-mm desired safe margin were found in the pubis (3.1 and 5.1 mm for senior and junior surgeons respectively). Of the 24 simulated resection, there was no intralesional tumor cutting. This study demonstrates that using PSI technology during simulated bone cuts of the pelvis can provide good cutting accuracy. Compared to a previous report on computer assistance for pelvic bone cutting, PSI technology clearly demonstrates an equivalent value-added for bone cutting accuracy than navigation technology. When in vivo validated, PSI technology may improve pelvic bone tumor surgery by providing clinically acceptable margins.
引用
收藏
页码:205 / 213
页数:9
相关论文
共 38 条
[1]   Recent advances in navigation-assisted musculoskeletal tumor resection [J].
Abraham, John A. .
CURRENT ORTHOPAEDIC PRACTICE, 2011, 22 (04) :297-302
[2]  
[Anonymous], 2008, F183908E1 INT ASTM
[3]  
Anselmetti B., 2003, TOLERANCEMENT COTATI, V3, P334
[4]  
Brown H., 1999, Applied mixed models in medicine
[5]  
Cartiaux O, 2011, EUR MUSC REV, V6, P125
[6]  
Cartiaux O., 2012, Computer Aided Surgery
[7]   Surgical inaccuracy of tumor resection and reconstruction within the pelvis - An experimental study [J].
Cartiaux, Olivier ;
Docquier, Pierre-Louis ;
Paul, Laurent ;
Francq, Bernard G. ;
Cornu, Olivier H. ;
Delloye, Christian ;
Raucent, Benoit ;
Dehez, Bruno ;
Banse, Xavier .
ACTA ORTHOPAEDICA, 2008, 79 (05) :695-702
[8]   Computer-Assisted and Robot-Assisted Technologies to Improve Bone-Cutting Accuracy When Integrated with a Freehand Process Using an Oscillating Saw [J].
Cartiaux, Olivier ;
Paul, Laurent ;
Docquier, Pierre-Louis ;
Raucent, Benoit ;
Dombre, Etienne ;
Banse, Xavier .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (11) :2076-2082
[9]   Accuracy in planar cutting of bones: an ISO-based evaluation [J].
Cartiaux, Olivier ;
Paul, Laurent ;
Docquier, Pierre-Louis ;
Francq, Bernard G. ;
Raucent, Benoit ;
Dombre, Etienne ;
Banse, Xavier .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (01) :77-84
[10]   Computer-assisted sacral tumor resection - A case report [J].
Cho, Hwan Seong ;
Kang, Hyun Guy ;
Kim, Han-Soo ;
Han, Ilkyu .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (07) :1561-1566