Determining the molar mass of a plasma substitute succinylated gelatin by size exclusion chromatography - multi-angle laser light scattering, sedimentation equilibrium and conventional size exclusion chromatography

被引:11
作者
Kaur, M
Jumel, K
Hardie, KR
Hardman, A
Meadows, J
Melia, CD
机构
[1] Univ Nottingham, Sch Biosci, NCMH, Nottingham LE12 5RD, England
[2] Univ Nottingham, Sch Pharm, Pharmaceut Sci Res Inst, Nottingham NG7 2RD, England
[3] Maelor Pharmaceut Ltd, Wrexham LL14 3JG, Wales
关键词
molar mass determination; sedimentation equilibrium; universal calibration; plasma substitute; succinylated gelatin;
D O I
10.1016/S0021-9673(02)00350-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The clinical effectiveness of succinylated gelatin as a plasma substitute depends strongly on its molar mass, determined conventionally by size exclusion chromatography (SEC). This study evaluates different SEC calibration standards in comparison with two independent "absolute" methods for determining the weight average molar mass (M-w) of a succinylated gelatin sample. SEC calibrated using succinylated gelatin fractions correlated well with size exclusion chromatography-multi-angle laser light scattering (SEC-MALLS) and sedimentation equilibrium whereas SEC calibrated with unmodified gelatin, sodium polystyrene sulfonates or pullulans overestimated M-w by over 20%. Universal calibration was equivocal. The problems associated with the preparation of succinylated gelatin fractions suggest that an absolute method such as SEC-MALLS may be a more suitable choice for determining the M-w in succinylated gelatins. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:139 / 148
页数:10
相关论文
共 38 条
[1]  
Barth HG, 1998, ANAL CHEM, V70, p251R
[2]   ETUDE PAR CHROMATOGRAPHIE EN PHASE LIQUIDE DE POLYSTYRENES LINEAIRES ET RAMIFIES DE STRUCTURES CONNUES [J].
BENOIT, H ;
GRUBISIC, Z ;
REMPP, P ;
DECKER, D ;
ZILLIOX, JG .
JOURNAL DE CHIMIE PHYSIQUE, 1966, 63 (11-1) :1507-&
[3]  
BEUTEL J, 1983, 4 IAG C PHOT GEL FIR, P65
[4]  
BILLINGHAM NC, 1977, MOLAR MASS MEASUREME
[5]   A STUDY OF GELATIN MOLECULES, AGGREGATES AND GELS [J].
BOEDTKER, H ;
DOTY, P .
JOURNAL OF PHYSICAL CHEMISTRY, 1954, 58 (11) :968-983
[6]  
BRUZZESE E, 1983, MED MIL, V133, P283
[7]   MSTARA and MSTARI: Interactive PC algorithms for simple, model independent evaluation of sedimentation equilibrium data [J].
Colfen, H ;
Harding, SE .
EUROPEAN BIOPHYSICS JOURNAL WITH BIOPHYSICS LETTERS, 1997, 25 (5-6) :333-346
[8]  
*COMM METH TEST PH, 1997, PHOT GEL IND JAP, P32
[9]   SIZE-EXCLUSION CHROMATOGRAPHY OF STRONG POLY-ELECTROLYTES ON POROUS-GLASS COLUMNS [J].
DUBIN, PL ;
TECKLENBURG, MM .
ANALYTICAL CHEMISTRY, 1985, 57 (01) :275-279
[10]   MOLECULAR CONFIGURATION OF GELATIN [J].
GOUINLOCK, EV ;
FLORY, PJ ;
SCHERAGA, HA .
JOURNAL OF POLYMER SCIENCE, 1955, 16 (82) :383-395