Second-Generation Electronic Ligament Balancing for Knee Arthroplasty: A Cadaver Study

被引:11
|
作者
Nielsen, Evan S. [1 ]
Hsu, Albert [1 ]
Patil, Shantanu [1 ]
Colwell, Clifford W., Jr. [1 ]
D'Lima, Darryl D. [1 ]
机构
[1] Scripps Hlth, Scripps Clin, Shiley Ctr Orthopaed Res & Educ, La Jolla, CA USA
关键词
total knee arthroplasty; soft-tissue balance; ligament balance; knee instability; balanced gap technique; gap balancing; SOFT-TISSUE RELEASE; IMPROVE TOTAL KNEE; PATIENT SATISFACTION; JOINT STABILITY; FAILING TODAY; REPLACEMENT; TKA; INSTABILITY; EXTENSION; ALIGNMENT;
D O I
10.1016/j.arth.2018.02.057
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Knee instability is emerging as a major complication after total knee arthroplasty (TKA), with ligament laxity and component alignment listed as important contributory factors. Knee balancing remains an art and is largely dependent on the surgeon's subjective "feel." The objectives were to measure the accuracy of an electronic balancing device to document the magnitude of correction in knee balance after soft-tissue releases and measure change in knee laxity after medial release. Methods: The accuracy of a second-generation electronic ligament-balancing device was compared with that of 2 mechanical balancing instruments. TKA was performed in 12 cadaver knees. Soft-tissue balance was measured sequentially before TKA, after mounting a trial femoral component, after medial release, and after resecting the posterior cruciate ligament. Coronal laxity of the knee under a 10 Nm valgus moment was measured before and after medial release. Results: The electronic balancing instrument was more accurate than mechanical instruments in measuring distracted gap and distraction force. On average, before TKA, the flexion gap was wider than the extension gap, and the medial gap was tighter than the lateral gap. Medial release increased the medial gap in flexion and increased passive knee valgus laxity. Posterior cruciate ligament release increased the tibiofemoral gap in both flexion and extension with a greater increase in the lateral gap. Conclusion: The second-generation electronic balancing device was significantly more accurate than mechanical instruments and could record knee balance over the entire range of flexion. More accurate soft-tissue balance may enhance outcomes after TKA. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:2293 / 2300
页数:8
相关论文
共 50 条
  • [31] The Use of Electronic Sensor Device to Augment Ligament Balancing Leads to a Lower Rate of Arthrofibrosis After Total Knee Arthroplasty
    Geller, Jeffrey A.
    Lakra, Akshay
    Murtaugh, Taylor
    JOURNAL OF ARTHROPLASTY, 2017, 32 (05) : 1502 - 1504
  • [32] Second-generation bi-cruciate stabilized total knee system has a lower reoperation and revision rate than its predecessor
    Christen, Bernhard
    Kopjar, Branko
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2018, 138 (11) : 1591 - 1599
  • [33] How Effective Is Multiple Needle Puncturing for Medial Soft Tissue Balancing During Total Knee Arthroplasty? A Cadaveric Study
    Koh, In Jun
    Kwak, Dai-Soon
    Kim, Tae Kyun
    Park, In Joo
    In, Yong
    JOURNAL OF ARTHROPLASTY, 2014, 29 (12) : 2478 - 2483
  • [34] Numerical analysis of variations in posterior cruciate ligament properties and balancing techniques on total knee arthroplasty loading
    Zelle, J.
    Heesterbeek, P. J. C.
    Malefijt, M. De Waal
    Verdonschot, N.
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (07) : 700 - 707
  • [35] Eversion or subluxation of patella in soft tissue balancing of total knee arthroplasty?: Results of a cadaver experiment
    Luring, C
    Hüfner, T
    Kendoff, D
    Perlick, L
    Bäthis, H
    Grifka, J
    Krettek, C
    KNEE, 2006, 13 (01) : 15 - 18
  • [36] Pulsatile Lavage During Cementation of Total Knee Arthroplasty - Is Fixation Impaired? A Cadaver Study
    Glowalla, Claudio
    Ertl, Max
    Lenze, Ulrich
    Lazic, Igor
    Burgkart, Rainer
    Lang, Jan J.
    Von Eisenhart-Rothe, Rudiger
    Pohlig, Florian
    IN VIVO, 2022, 36 (02): : 672 - 677
  • [37] Gap-Balancing versus Measured Resection Technique in Total Knee Arthroplasty: A Comparison Study
    Churchill, Jessica L.
    Khlopas, Anton
    Sultan, Assem A.
    Harwin, Steven F.
    Mont, Michael A.
    JOURNAL OF KNEE SURGERY, 2018, 31 (01) : 13 - 16
  • [38] Influence of patellar subluxation on ligament balancing in total knee arthroplasty through a subvastus approach. An in vivo study
    De Keyser, Wouter
    Beckers, Leo
    ACTA ORTHOPAEDICA BELGICA, 2010, 76 (06): : 799 - 805
  • [39] Second-Generation Locking Mechanisms and Ethylene Oxide Sterilization Reduce Tibial Insert Backside Damage in Total Knee Arthroplasty
    Azzam, Michael G.
    Roy, Marcel E.
    Whiteside, Leo A.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (04) : 523 - 530
  • [40] Objective quantification of ligament balancing using VERASENSE in measured resection and modified gap balance Iota knee arthroplasty
    Cho, Kyu-Jin
    Seon, Jong-Keun
    Jang, Won-Young
    Park, Chun-Gon
    Song, Eun-Kyoo
    BMC MUSCULOSKELETAL DISORDERS, 2018, 19