Improving PET imaging for breast cancer using virtual pinhole PET half-ring insert

被引:22
作者
Mathews, Aswin John [1 ]
Komarov, Sergey [2 ]
Wu, Heyu [3 ]
O'Sullivan, Joseph A. [1 ]
Tai, Yuan-Chuan [2 ]
机构
[1] Washington Univ, Dept Elect & Syst Engn, St Louis, MO 63130 USA
[2] Washington Univ, Dept Radiol, St Louis, MO 63110 USA
[3] Sinogram Technol Beijing Co Ltd, Beijing, Peoples R China
基金
美国国家卫生研究院;
关键词
POSITRON EMISSION MAMMOGRAPHY; F-18; FLUORODEOXYGLUCOSE; RECONSTRUCTION; RESOLUTION; SCANNER; SYSTEM; DESIGN; DEVICE; FEASIBILITY; TOMOGRAPHY;
D O I
10.1088/0031-9155/58/18/6407
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A PET insert with detector having smaller crystals and placed near a region of interest in a conventional PET scanner can improve image resolution locally due to the virtual-pinhole PET (VP-PET) effect. This improvement is from the higher spatial sampling of the imaging area near the detector. We have built a prototype half-ring PET insert for head-and-neck cancer imaging applications. In this paper, we extend the use of the insert to breast imaging and show that such a system provides high resolution images of breast and axillary lymph nodes while maintaining the full imaging field of view capability of a clinical PET scanner. We characterize the resolution and contrast recovery for tumors across the imaging field of view. First, we model the system using Monte Carlo methods to determine its theoretical limit of improvement. Simulations were conducted with hot spherical tumors embedded in background activity at tumor-to-background contrast ranging from 3: 1 to 12:1. Tumors are arranged in a Derenzo-like pattern with their diameters ranging from 2 to 12 mm. Experimental studies were performed using a chest phantom with cylindrical breast attachment. Tumors of different sizes arranged in a Derenzo-like pattern with tumor-to-background ratio of 6: 1 are inserted into the breast phantom. Imaging capability of mediastinum and axillary lymph nodes is explored. Both Monte Carlo simulations and experiment show clear improvement in image resolution and contrast recovery with VP-PET half-ring insert. The degree of improvement in resolution and contrast recovery depends on location of the tumor. The full field of view imaging capability is shown to be maintained. Minor artifacts are introduced in certain regions.
引用
收藏
页码:6407 / 6427
页数:21
相关论文
共 30 条
  • [1] Design and evaluation of the Clear-PEM scanner for positron emission mammography
    Abreu, MC
    Aguiar, JD
    Almeida, FG
    Almeida, P
    Bento, P
    Carrico, B
    Ferreira, M
    Ferreira, NC
    Gonçalves, F
    Leong, C
    Lopes, F
    Lousa, P
    Martins, MV
    Matela, N
    Mendes, PR
    Moura, R
    Nobre, J
    Oliveira, N
    Ortigao, C
    Peralta, L
    Pereira, R
    Rego, J
    Ribeiro, R
    Rodrigues, P
    Sampaio, J
    Santos, AI
    Silva, L
    Silva, JC
    Sousa, P
    Teixeira, IC
    Teixeira, JP
    Trindade, A
    Varela, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 71 - 77
  • [2] American Cancer Society, 2011, BREAST CANC FACTS FI
  • [3] Breast imaging with positron emission tomography and fluorine-18 fluorodeoxyglucose:: Use and limitations
    Avril, N
    Rosé, CA
    Schelling, M
    Dose, J
    Kuhn, W
    Bense, S
    Weber, W
    Ziegler, S
    Graeff, H
    Schwaiger, M
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2000, 18 (20) : 3495 - 3502
  • [4] Metabolic characterization of breast tumors with positron emission tomography using F-18 fluorodeoxyglucose
    Avril, N
    Dose, J
    Janicke, F
    Bense, S
    Ziegler, S
    Laubenbacher, C
    Romer, W
    Pache, H
    Herz, M
    Allgayer, B
    Nathrath, W
    Graeff, H
    Schwaiger, M
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 1996, 14 (06) : 1848 - 1857
  • [5] Model-based normalization for iterative 3D PET image reconstruction
    Bai, B
    Li, Q
    Holdsworth, CH
    Asma, E
    Tai, YC
    Chatziioannou, A
    Leahy, RM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (15) : 2773 - 2784
  • [6] Development of the YAP-PEM scanner for breast cancer imaging
    Camarda, Manuela
    Belcari, Nicola
    Del Guerra, Alberto
    Galeotti, Stefano
    Morsani, Fabio
    Herbert, Deborah J.
    Vaiano, Angela
    [J]. PHYSICA MEDICA, 2006, 21 : 114 - 116
  • [7] CLINTHORNE NH, 2003, IEEE NUCL SCI S, V3, P1997
  • [8] Delfino EP, 2010, IEEE NUCL SCI CONF R, P3442, DOI 10.1109/NSSMIC.2010.5874446
  • [9] maxPET: A dedicated mammary and axillary region PET imaging system for breast cancer
    Doshi, NK
    Silverman, RW
    Shao, Y
    Cherry, SR
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 811 - 815
  • [10] Design and evaluation of an LSO PET detector for breast cancer imaging
    Doshi, NK
    Shao, YP
    Silverman, RW
    Cherry, SR
    [J]. MEDICAL PHYSICS, 2000, 27 (07) : 1535 - 1543