In-Place Calibration With Single Measurement in Time-Domain Microwave Breast Imaging

被引:3
|
作者
Kwon, Sollip [1 ]
Lee, Seungjun [1 ]
机构
[1] Ewha Womans Univ, Dept Elect Engn, Seoul 120750, South Korea
基金
新加坡国家研究基金会;
关键词
Calibration; clock jitter; microwave imaging; time-domain measurements; CANCER DETECTION; RADAR;
D O I
10.1109/LAWP.2016.2569818
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Microwave breast imaging requires a calibration process, which extracts the tumor response from the received signal. Existing calibration methods require more than two measurements. In this letter, we propose a calibration method called "in-place calibration" for time-domain microwave imaging. The method can obtain the received signal and the reference signal in a single measurement with the antenna array in place. It can minimize the measurement errors that occur due to environmental changes during calibration. The method also minimizes the error due to random clock jitter. The proposed in-place calibration is compared to array rotation calibration and differential calibration using homogeneous and heterogeneous breast phantoms. In most cases, in-place calibration provided better results than array rotation calibration.
引用
收藏
页码:206 / 209
页数:4
相关论文
共 50 条
  • [1] Verification of In place Calibration for Time-domain Microwave Imaging Development of Time-domain Microwave Breast Imaging System for Experimental Verification of In place Calibration
    Kwon, Sollip
    Son, Seong-Ho
    Lee, Kwang-Jae
    Lee, Seungjun
    2016 URSI ASIA-PACIFIC RADIO SCIENCE CONFERENCE (URSI AP-RASC), 2016, : 688 - 690
  • [2] Experimental Demonstration of In-Place Calibration for Time Domain Microwave Imaging System
    Kwon, S.
    Son, S.
    Lee, K.
    RADIO SCIENCE, 2018, 53 (04) : 429 - 439
  • [3] Performance Evaluation of In-Place Calibration in Microwave Imaging for Breast Cancer Detection
    Kwon, Sollip
    Lee, Seungjun
    2016 10TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2016,
  • [4] Improved Calibration for an Experimental Time-Domain Microwave Imaging System
    Santorelli, Adam
    Kirshin, Evgeny
    Porter, Emily
    Popovic, Milica
    Schwartz, Joshua
    2013 7TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2013, : 801 - 805
  • [5] A Time-Domain Measurement System for UWB Microwave Imaging
    Shao, Wenyi
    Edalati, Arezou
    McCollough, Todd R.
    McCollough, William J.
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2018, 66 (05) : 2265 - 2275
  • [6] Time-Domain Microwave Radar Applied to Breast Imaging: Measurement Reliability in a Clinical Setting
    Porter, Emily
    Santorelli, Adam
    Popovic, Milica
    PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER, 2014, 149 : 119 - 132
  • [7] Microwave Breast Imaging: Time-Domain Experiments on Tissue Phantoms
    Porter, Emily
    Santorelli, Adam
    Coates, Mark
    Popovic, Milica
    2011 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION (APSURSI), 2011, : 695 - 698
  • [8] TIME-DOMAIN MICROWAVE TARGET IMAGING
    YEUNG, WK
    EVANS, S
    IEE PROCEEDINGS-H MICROWAVES ANTENNAS AND PROPAGATION, 1985, 132 (06) : 345 - 350
  • [9] Instantaneous microwave imaging with time-domain measurements for breast cancer detection
    Kwon, S.
    Lee, S.
    ELECTRONICS LETTERS, 2013, 49 (10) : 639 - 640
  • [10] A Clinical Prototype for Microwave Breast Imaging Using Time-Domain Measurements
    Porter, Emily
    Kirshin, Evgeny
    Santorelli, Adam
    Popovic, Milica
    2013 7TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2013, : 830 - U862