A novel approach to EPID-based 3D volumetric dosimetry for IMRT and VMAT QA

被引:22
作者
Alhazmi, Abdulaziz [1 ]
Gianoli, Chiara [1 ]
Neppl, Sebastian [2 ]
Martins, Juliana [1 ]
Veloza, Stella [1 ]
Podesta, Mark [3 ]
Verhaegen, Frank [3 ]
Reiner, Michael [2 ]
Belka, Claus [2 ]
Parodi, Katia [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, Fac Phys, Dept Med Phys, Munich, Germany
[2] Univ Hosp, Dept Radiat Oncol, LMU Munich, Munich, Germany
[3] Maastricht Univ, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, Med Ctr, Maastricht, Netherlands
关键词
EPID dosimetry; pre-treatment QA; dose reconstruction; VMAT; IMRT; 2D DIODE-ARRAY; QUALITY-ASSURANCE; VERIFICATION; POSITION; MODE; FILM;
D O I
10.1088/1361-6560/aac1a6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) are relatively complex treatment delivery techniques and require quality assurance (QA) procedures. Pre-treatment dosimetric verification represents a fundamental QA procedure in daily clinical routine in radiation therapy. The purpose of this study is to develop an EPID-based approach to reconstruct a 3D dose distribution as imparted to a virtual cylindrical water phantom to be used for plan-specific pre-treatment dosimetric verification for IMRT and VMAT plans. For each depth, the planar 2D dose distributions acquired in air were back-projected and convolved by depth-specific scatter and attenuation kernels. The kernels were obtained by making use of scatter and attenuation models to iteratively estimate the parameters from a set of reference measurements. The derived parameters served as a look-up table for reconstruction of arbitrary measurements. The summation of the reconstructed 3D dose distributions resulted in the integrated 3D dose distribution of the treatment delivery. The accuracy of the proposed approach was validated in clinical IMRT and VMAT plans by means of gamma evaluation, comparing the reconstructed 3D dose distributions with Octavius measurement. The comparison was carried out using (3% , 3 mm) criteria scoring 99% and 96% passing rates for IMRT and VMAT, respectively. An accuracy comparable to the one of the commercial device for 3D volumetric dosimetry was demonstrated. In addition, five IMRT and five VMAT were validated against the 3D dose calculation performed by the TPS in a water phantom using the same passing rate criteria. The median passing rates within the ten treatment plans was 97.3%, whereas the lowest was 95%. Besides, the reconstructed 3D distribution is obtained without predictions relying on forward dose calculation and without external phantom or dosimetric devices. Thus, the approach provides a fully automated, fast and easy QA procedure for plan-specific pretreatment dosimetric verification.
引用
收藏
页数:13
相关论文
共 33 条
  • [1] Three dimensional dose verification of VMAT plans using the Octavius 4D dosimetric system
    Arumugam, Sankar
    Xing, Aitang
    Young, Tony
    Thwaites, David
    Holloway, Lois
    [J]. 8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [2] EPID dosimetry for pretreatment quality assurance with two commercial systems
    Bailey, Daniel W.
    Kumaraswamy, Lalith
    Bakhtiari, Mohammad
    Malhotra, Harish K.
    Podgorsak, Matthew B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (04): : 82 - 99
  • [3] Dosimetric characteristics and day-to-day performance of an amorphous-silicon type electronic portal imaging device
    Chang, Chiou-Shiung
    Tseng, Yi-Han
    Hwang, Jing-Min
    Shih, Rompin
    Chuang, Keh-Shih
    [J]. RADIATION MEASUREMENTS, 2016, 91 : 9 - 14
  • [4] Patient specific IMRT quality assurance with film, ionization chamber and detector arrays: Our institutional experience
    Cruz, Wilbert
    Narayanasamy, Ganesh
    Regan, Morgan
    Mavroidis, Panayiotis
    Papanikolaou, Niko
    Ha, Chul S.
    Stathakis, Sotirios
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2015, 115 : 12 - 16
  • [5] ML Segmentation Strategies for Object Interference Compensation in FDG-PET Lesion Quantification
    De Bernardi, E.
    Gallotta, F. Fiorani
    Gianoli, C.
    Zito, F.
    Gerundini, P.
    Baselli, G.
    [J]. METHODS OF INFORMATION IN MEDICINE, 2010, 49 (05) : 537 - 541
  • [6] Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithm
    Fippel, M
    [J]. MEDICAL PHYSICS, 1999, 26 (08) : 1466 - 1475
  • [7] Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors
    Fredh, Anna
    Scherman, Jonas Bengtsson
    Fog, Lotte S.
    Af Rosenschold, Per Munck
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [8] A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment
    Fuangrod, Todsaporn
    Woodruff, Henry C.
    van Uytven, Eric
    McCurdy, Boyd M. C.
    Kuncic, Zdenka
    O'Connor, Daryl J.
    Greer, Peter B.
    [J]. MEDICAL PHYSICS, 2013, 40 (09)
  • [9] Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry
    Greer, PB
    [J]. MEDICAL PHYSICS, 2005, 32 (12) : 3558 - 3568
  • [10] American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for Intensity-modulated Radiation Therapy (IMRT)
    Hartford, Alan C.
    Galvin, James M.
    Beyer, David C.
    Eichler, Thomas J.
    Ibbott, Geoffrey S.
    Kavanagh, Brian
    Schultz, Christopher J.
    Rosenthal, Seth A.
    [J]. AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2012, 35 (06): : 612 - 617