26.5-40 GHz Ka-band pulsed EPR spectrometer

被引:51
作者
Astashkin, A. V. [1 ]
Enemark, J. H. [1 ]
Raitsimring, A. [1 ]
机构
[1] Univ Arizona, Dept Chem, Tucson, AZ 85721 USA
关键词
K-a-band; pulsed EPR spectrometer; ESEEM; ENDOR; ELDOR;
D O I
10.1002/cmr.b.20064
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The design and performance of a new homebuilt K-a-band pulsed electron paramagnetic resonance (EPR) spectrometer is described. The spectrometer is unique in that it is broadband and allows experiments to be performed at any operational frequency within the range from 26.5 to 40 GHz. The high-power microwave amplifier employed in this spectrometer allowed us to use resonators with low quality factor and to achieve a minimum inversion microwave pulse duration of 11 ns. As a result, this spectrometer can be used for all types of pulsed EPR experiments, including electron spin echo envelope modulation (ESEEM), electron - nuclear double resonance (ENDOR), and electron-electron double resonance (ELDOR) with a difference between the pumping and observation frequencies of up to several hundred megahertz. The timing is controlled by an arbitrary waveform generator with 12 digital outputs and 1 ns time resolution. The spectrometer control/data acquisition software is home-written and allows an easy implementation of any one- or two-dimensional experiment using sweep definition tables. (C) 2006 Wiley Periodicals, Inc.
引用
收藏
页码:125 / 136
页数:12
相关论文
共 35 条
[1]   Two- and four-pulse ESEEM studies of the heme binding center of a low-spin ferriheme protein: the importance of a multi-frequency approach [J].
Astashkin, AV ;
Raitsimring, AM ;
Walker, FA .
CHEMICAL PHYSICS LETTERS, 1999, 306 (1-2) :9-17
[2]   Pulsed EPR investigations of systems modeling molybdenum enzymes:: Hyperfine and quadrupole parameters of OXO-17O in [MO17O(SPh)4]- [J].
Astashkin, AV ;
Neese, F ;
Raitsimring, AM ;
Cooney, JJA ;
Bultman, E ;
Enemark, JH .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2005, 127 (47) :16713-16722
[3]   Refocused primary echo: A zero dead time detection of the electron spin echo envelope modulation [J].
Astashkin, AV ;
Raitsimring, AM .
JOURNAL OF MAGNETIC RESONANCE, 2000, 143 (02) :280-291
[4]  
ASTASHKIN AV, 1998, 40 ROCK MOUNT C AN C
[5]  
BORBAT PP, 1994, 36 ROCK MOUNT C AN C
[6]   NEW PULSE ENDOR TECHNIQUE [J].
DAVIES, ER .
PHYSICS LETTERS A, 1974, A 47 (01) :1-2
[7]   Q-band pulsed electron spin-echo spectrometer and its application to ENDOR and ESEEM [J].
Davoust, CE ;
Doan, PE ;
Hoffman, BM .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1996, 119 (01) :38-44
[8]  
de Beer R., 1979, Delft Progress Report, V4, P63
[9]  
Dikanov S. A., 1992, ELECT SPIN ECHO MODU
[10]  
ENEMARK J, 2002, CONSTRUCTION VARIABL