Development and validation of two phantoms for quality control in cone-beam CT

被引:1
|
作者
Gomes Batista, Wilson Otto [1 ]
机构
[1] Inst Fed Bahia, Rua Emidio dos Santos S-N, BR-40301015 Salvador, BA, Brazil
关键词
Quality control; Development of phantoms; Cone Beam CT; COMPUTED-TOMOGRAPHY; IMAGE QUALITY;
D O I
10.1016/j.apradiso.2017.07.052
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
This study aimed at the development and validation of the models of phantom: CQ-02-IFBA and CQ-01-IFBA, the first filled with water and the second fully with a solid structure in polymethylmethacrylate (PMMA). Both models allow its use in various models of cone beam computed tomography (CBCT), independent of the size of the field of view. The two phantoms are used to evaluate seven parameters of quality control: intensity or average value of pixels of five different plastics, signal to noise ratio, resolution, low contrast, spatial resolution, the accuracy of distances on the z axis, the geometric distortion and image uniformity. Spatial resolution is a critical parameter that differs significantly in the various imaging modalities in dental radiology, and in these two phantoms, it can be evaluated by two different methods: MTF obtained by Fourier transformation of the function LSF (line spread function) and/or by analysis visual image to a pattern of bars up to 16 1p/cm. Validation was performed in three models CBCT: CareStream CS 9000 3D, i-CAT Classical and Sirona Orthophos XG 3D. The two phantom models were viable and also complied with the specifications and data available in the literature. It is concluded that the two phantom models are viable for use in CBCT with different characteristics.
引用
收藏
页码:29 / 34
页数:6
相关论文
共 50 条
  • [1] Development and applicability of a quality control phantom for dental cone-beam CT
    Pauwels, Ruben
    Stamatakis, Harry
    Manousaridis, Giorgos
    Walker, Adrian
    Michielsen, Koen
    Bosmans, Hilde
    Bogaerts, Ria
    Jacobs, Reinhilde
    Horner, Keith
    Tsiklakis, Kostas
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (04): : 245 - 260
  • [2] Image quality of flat panel cone-beam CT
    Rose, G
    Wiegert, J
    Schäfer, D
    Fiedler, K
    Conrads, N
    Timmer, J
    Rasche, V
    Noordhoek, N
    Klotz, E
    Koppe, R
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 677 - 683
  • [3] Image quality and dose for a multisource cone-beam CT extremity scanner
    Gang, Grace J.
    Zbijewski, Wojciech
    Mahesh, Mahadevappa
    Thawait, Gaurav
    Packard, Nathan
    Yorkston, John
    Demehri, Shadpour
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2018, 45 (01) : 144 - 155
  • [4] The influence of bowtie filtration on cone-beam CT image quality
    Mail, N.
    Moseley, D. J.
    Siewerdsen, J. H.
    Jaffray, D. A.
    MEDICAL PHYSICS, 2009, 36 (01) : 22 - 32
  • [5] Development of a quality control program for a Cone Beam CT Imaging System
    Benitez, Ricardo Betancourt
    Ning, Ruola
    Conover, David
    Zhang, Yan
    Cai, Weixing
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [6] Crescent artifacts in cone-beam CT
    Giles, William
    Bowsher, James
    Li, Hao
    Yin, Fang-Fang
    MEDICAL PHYSICS, 2011, 38 (04) : 2116 - 2121
  • [7] Image quality optimization using a narrow vertical detector dental cone-beam CT
    Brasil, Danieli Moura
    Pauwels, Ruben
    Coucke, Wim
    Haiter-Neto, Francisco
    Jacobs, Reinhilde
    DENTOMAXILLOFACIAL RADIOLOGY, 2019, 48 (03)
  • [8] Flying focal spot (FFS) in cone-beam CT
    Kachelriess, Marc
    Knaup, Michael
    Penssel, Christian
    Kalender, Willi A.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) : 1238 - 1247
  • [9] A Validation Framework for Head-Motion Artifacts in Dental Cone-Beam CT
    Ens, S.
    Bruder, R.
    Ulrici, J.
    Hell, E.
    Buzug, T. M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 658 - 661
  • [10] Two methods for reducing moving metal artifacts in cone-beam CT
    Hahn, Andreas
    Knaup, Michael
    Brehm, Marcus
    Sauppe, Sebastian
    Kachelriess, Marc
    MEDICAL PHYSICS, 2018, 45 (08) : 3671 - 3680