Variation of depth of cure and intensity with distance using LED curing lights

被引:49
作者
Aravamudhan, Krishna
Rakowski, Duane
Fan, P. L.
机构
[1] Amer Dent Assoc, Div Sci, Res & Labs, Chicago, IL 60611 USA
[2] Amer Dent Assoc, Div Sci, Int Sci & Stand, Chicago, IL 60611 USA
关键词
dental material; intensity depth of cure; distance; light emitting diode; light curing; resin composite;
D O I
10.1016/j.dental.2005.11.031
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives. The purpose of the study was to determine the correlation between intensity of light-emitting diode (LED) and tungsten-halogen light sources, and depth of cure of a resin composite at different distances. Methods. Four LED curing lights (Flashlite 1001, Freelight 2, Smartlite IQ and Ultralume 5) and one tungsten halogen (Optilux 501, with 8 and 11 mm tips) were evaluated. Intensity was measured according a modified ISO Standard 10650 at distances of 0, 2, 4, 6, 8, 10 mm between the light tip and detector. Depth of cure (DOC) of TPH Spectrum shade A2 was measured according to the international standard ISO 4049 at the same distances. Results. For all lights, intensity decreased as distance increased. The authors documented a logarithmic correlation between intensity and distance for all lights except the Smartlite IQ, Ultralume 5 and the Optilux S01 with the 11 mm tip, which showed a linear relationship between intensity and distance. All lights demonstrated a logarithmic correlation between intensity and DOC, and a linear correlation between DOC and distance. Smartlite IQ and Optilux 501 (11 mm tip) also had the least reduction in intensity and DOC at 10 mm. Significance. Clinicians often an experience difficulty placing the light tip close to the resin surface when curing resin composites. While both intensity and DOC decrease with increasing distance, the relationship between these factors and distance may not be similar for all lights and may depend on the characteristics of individual lights. (C) 2005 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:988 / 994
页数:7
相关论文
共 21 条
[1]   Performance of two blue light-emitting-diode dental light curing units with distance and irradiation-time [J].
Bennett, AW ;
Watts, DC .
DENTAL MATERIALS, 2004, 20 (01) :72-79
[2]  
DUGAN W T, 1989, General Dentistry, V37, P40
[3]   IRRADIANCE OF VISIBLE LIGHT-CURING UNITS AND VOLTAGE VARIATION EFFECTS [J].
FAN, PL ;
WOZNIAK, WT ;
REYES, WD ;
STANFORD, JW .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 1987, 115 (03) :442-445
[4]   Curing-light intensity and depth of cure of resin-based composites tested according to international standards [J].
Fan, PL ;
Schumacher, RM ;
Azzolin, K ;
Geary, R ;
Eichmiller, FC .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2002, 133 (04) :429-434
[5]  
Felix CA, 2003, J ADHES DENT, V5, P283
[6]  
Friedman J, 1989, J Esthet Dent, V1, P189, DOI 10.1111/j.1708-8240.1989.tb00500.x
[7]  
HANSEN EK, 1993, SCAND J DENT RES, V101, P115
[8]   Visible-light curing units: Correlation between depth of cure and distance between exit window and resin surface [J].
Hansen, EK ;
Asmussen, E .
ACTA ODONTOLOGICA SCANDINAVICA, 1997, 55 (03) :162-166
[9]  
*INT ORG STAND, 2004, 106502004 ISO
[10]  
International Organization for Standardization, 2000, 40492000 ISO