Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes

被引:39
作者
Utine, Canan Asli [4 ]
Altin, Feyza [1 ]
Cakir, Hanefi [2 ]
Perente, Irfan [3 ]
机构
[1] SB Okmeydani Res & Training Hosp, Dept Ophthalmol, Istanbul, Turkey
[2] Turkiye Hosp Eye Clin, Dept Ophthalmol, Istanbul, Turkey
[3] Beyoglu Eye Res & Training Hosp, Dept Ophthalmol, Istanbul, Turkey
[4] Yeditepe Univ, Hosp Eye, Turkiye Hosp Eye Clin, Dept Ophthalmol, Istanbul, Turkey
关键词
anterior chamber depth; IOLMaster; Orbscan; Pentacam; ANGLE-CLOSURE GLAUCOMA; A-SCAN; BIOMETRY; REPEATABILITY; ULTRASONOGRAPHY; AGREEMENT; CORNEAS; CHINESE;
D O I
10.1111/j.1755-3768.2008.01278.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
This study determined to assess the degree of agreement between anterior chamber depth (ACD) measurements obtained using three different devices and to analyse the relationship between ACD and spherical equivalent (SE) refraction. In this cross-sectional study, 42 eyes of 42 patients with a mean SE of - 4.69 +/- 4.61 D (range 0.00 D to - 14.88 D) were analysed. Measurements of ACD between the corneal epithelium and the anterior surface of the crystalline lens, obtained using the Pentacam, Orbscan IIz and IOLMaster, were compared. The relationships between SE and ACD measurements obtained with different devices were also investigated. The results were analysed using Bland-Altman analyses, single-sample t-test and Pearson's correlation test. Orbscan ACD measurements were an average of 0.05 mm less than Pentacam measurements (p = 0.01). IOLMaster measurements were an average of 0.06 mm less than Orbscan measurements (p < 0.001). None of the ACD values measured by any of the devices were correlated with increasing SE (p > 0.05 for all). There was a weak positive correlation between SE and the difference in ACD measurements with Pentacam and Orbscan (p = 0.04); however, the differences between Pentacam and IOLMaster ACD measurements and Orbscan and IOLMaster ACD measurements seemed to be independent of SE (p = 0.17 and p = 0.54, respectively). The ACD in clinically normal eyes is measured differently by various non-ultrasonic devices. However, the observed mean error between these modalities is too small to create any noticeable difference in refractive outcome. No significant relationship was found between SE and ACD measurements obtained by Pentacam, Orbscan or IOLMaster.
引用
收藏
页码:386 / 391
页数:6
相关论文
共 37 条
[1]   Measuring anterior chamber depth with the Orbscan Topography System [J].
Auffarth, GU ;
Tetz, MR ;
Biazid, Y ;
Volcker, HE .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (09) :1351-1355
[2]   Measurement of the internal diameter and depth of the anterior chamber: IOLMaster versus anterior chamber optical coherence tomographer [J].
Baikoff, G ;
Jodai, HJ ;
Bourgeon, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1722-1728
[3]   Anterior chamber depth measurement in clinical practice [J].
Barrett, BT ;
McGraw, PV ;
Murray, LA ;
Murgatroyd, P .
OPTOMETRY AND VISION SCIENCE, 1996, 73 (07) :482-486
[4]  
Barrio V, 1998, NEFROLOGIA, V18, P32
[5]   COMPARING 2 METHODS OF CLINICAL MEASUREMENT - A PERSONAL HISTORY [J].
BLAND, JM ;
ALTMAN, DG .
INTERNATIONAL JOURNAL OF EPIDEMIOLOGY, 1995, 24 :S7-S14
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]   Accuracy of Orbscan optical pachymetry in corneas with haze [J].
Boscia, F ;
La Tegola, MG ;
Alessio, G ;
Sborgia, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :253-258
[8]  
Cho P, 2002, OPTOMETRY VISION SCI, V79, P462
[9]   The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting [J].
Cho, P ;
Lam, AKC ;
Mountford, J ;
Ng, L .
OPTOMETRY AND VISION SCIENCE, 2002, 79 (03) :175-183
[10]   Biometry and primary angle-closure glaucoma among Chinese, white, and black populations [J].
Congdon, NG ;
Qi, YL ;
Quigley, H ;
Hung, PT ;
Wang, TH ;
Ho, TC ;
Tielsch, JM .
OPHTHALMOLOGY, 1997, 104 (09) :1489-1495