On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry

被引:58
作者
Andreo, Pedro [1 ]
Palmans, Hugo [2 ,3 ]
Marteinsdottir, Maria [1 ]
Benmakhlouf, Hamza [1 ]
Carlsson-Tedgren, Asa [1 ]
机构
[1] Karolinska Univ Hosp, Dept Med Phys, SE-17176 Stockholm, Sweden
[2] Natl Phys Lab, Acoust & Ionising Radiat Div, Teddington TW11 0LW, Middx, England
[3] EBG MedAustron GmbH, Dept Med Phys, A-2700 Wiener Neustadt, Austria
关键词
small field dosimetry; Monte Carlo; output correction factors; detector geometry accuracy; CHAMBERS;
D O I
10.1088/0031-9155/61/1/L1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Monte Carlo (MC) calculated detector-specific output correction factors for small photon beam dosimetry are commonly used in clinical practice. The technique, with a geometry description based on manufacturer blueprints, offers certain advantages over experimentally determined values but is not free of weaknesses. Independent MC calculations of output correction factors for a PTW-60019 micro-diamond detector were made using the EGSnrc and PENELOPE systems. Compared with published experimental data the MC results showed substantial disagreement for the smallest field size simulated (5 mm x 5 mm). To explain the difference between the two datasets, a detector was imaged with x rays searching for possible anomalies in the detector construction or details not included in the blueprints. A discrepancy between the dimension stated in the blueprints for the active detector area and that estimated from the electrical contact seen in the x-ray image was observed. Calculations were repeated using the estimate of a smaller volume, leading to results in excellent agreement with the experimental data. MC users should become aware of the potential differences between the design blueprints of a detector and its manufacturer production, as they may differ substantially. The constraint is applicable to the simulation of any detector type. Comparison with experimental data should be used to reveal geometrical inconsistencies and details not included in technical drawings, in addition to the well-known QA procedure of detector x-ray imaging.
引用
收藏
页码:L1 / L10
页数:10
相关论文
共 20 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]   PENGEOM-A general-purpose geometry package for Monte Carlo simulation of radiation transport in material systems defined by quadric surfaces [J].
Almansa, Julio ;
Salvat-Pujol, Francesc ;
Diaz-Londono, Gloria ;
Carnicer, Artur ;
Lallena, Antonio M. ;
Salvat, Francesc .
COMPUTER PHYSICS COMMUNICATIONS, 2016, 199 :102-113
[3]   On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry [J].
Andreo, Pedro ;
Burns, David T. ;
Salvat, Francesc .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (08) :2117-2136
[4]   Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams [J].
Azangwe, Godfrey ;
Grochowska, Paulina ;
Georg, Dietmar ;
Izewska, Joanna ;
Hopfgartner, Johannes ;
Lechner, Wolfgang ;
Andersen, Claus E. ;
Beierholm, Anders R. ;
Helt-Hansen, Jakob ;
Mizuno, Hideyuki ;
Fukumura, Akifumi ;
Yajima, Kaori ;
Gouldstone, Clare ;
Sharpe, Peter ;
Meghzifene, Ahmed ;
Palmans, Hugo .
MEDICAL PHYSICS, 2014, 41 (07)
[5]   Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo study [J].
Benmakhlouf, Hamza ;
Sempau, Josep ;
Andreo, Pedro .
MEDICAL PHYSICS, 2014, 41 (04)
[6]   Validation of a Monte Carlo model of a NACP-02 plane-parallel ionization chamber model using electron backscatter experiments [J].
Chin, E. ;
Shipley, D. ;
Bailey, M. ;
Seuntjens, J. ;
Palmans, H. ;
DuSautoy, A. ;
Verhaegen, F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (08) :N119-N126
[7]   A methodological approach to reporting corrected small field relative outputs [J].
Cranmer-Sargison, Gavin ;
Charles, Paul H. ;
Trapp, Jamie V. ;
Thwaites, David I. .
RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) :350-355
[8]   Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields (vol 58, pg 2431, 2013) [J].
Czarnecki, D. ;
Zink, K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (03) :791-794
[9]   Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields [J].
Czarnecki, D. ;
Zink, K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08) :2431-2444
[10]   Calculation of kQclin,Qmsrfclin,fmsr for several small detectors and for two linear accelerators using Monte Carlo simulations [J].
Francescon, P. ;
Cora, S. ;
Satariano, N. .
MEDICAL PHYSICS, 2011, 38 (12) :6513-6527