More accurate quantification of pulmonary blood flow by magnetic resonance imaging than by lung perfusion scintigraphy in patients with Fontan circulation

被引:77
作者
Fratz, S
Hess, J
Schwaiger, M
Martinoff, S
Stern, HC
机构
[1] Klin Tech Univ Munich, Deutsch Herzzentrum Munchen, Dept Pediat Cardiol & Congenital Heart Dis, Munich, Germany
[2] Klin Tech Univ Munich, Deutsch Herzzentrum Munchen, Inst Radiol & Nucl Med, Munich, Germany
[3] Klin Tech Univ Munich, Klinikum Rechts Isar, Dept Nucl Med, Munich, Germany
关键词
Fontan procedure; magnetic resonance imaging; scintigraphy; lung;
D O I
10.1161/01.CIR.0000029103.26029.1E
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Quantitative evaluation of pulmonary perfusion using lung perfusion scintigraphy in patients with atriopulmonary anastomosis (APA) or total cavopulmonary connection (TCPC) or partial cavopulmonary connection (PCPC) is difficult because of preferential draining of the venae cavae to one lung. Scintigraphy is the gold standard. Phase-velocity MRI (PV-MRI) is a new technique for determining pulmonary perfusion. The aim of this study was therefore to determine whether PV-MRI is more accurate than scintigraphy for quantitative evaluation of pulmonary perfusion ratios in patients with APA, TCPC, or PCPC. Methods and Results-We studied 15 patients with APA, TCPC, or PCPC (16 7 years old, 4 female). Twelve patients (15+/-8 years old, 3 female) with a single pulmonary blood source supplied by a subpulmonary ventricle, ensuring complete mixing of the radioactive tracer before entering the pulmonary circulation, served as controls. Pulmonary scintigraphy and PV-MRI were performed in all patients. Bland-Altman analysis showed a clinically unacceptable difference of 7.1% right pulmonary blood flow (27.2% upper and - 13.0% lower limit of agreement) between the two methods in the study group. The two methods agreed excellently in the control group (difference, 1.6%; 4.0% upper and -7.2% lower limit of agreement), showing that the bad agreement in the study group was caused by the problems encountered using pulmonary scintigraphy in patients with APA, TCPC, or PCPC. Conclusions-Because of preferential caval flow into either lung, PV-MRI is more accurate for evaluating pulmonary perfusion ratios than lung perfusion scintigraphy in patients with Fontan-like circulation.
引用
收藏
页码:1510 / 1513
页数:4
相关论文
共 16 条
  • [1] In vivo evaluation of Fontan pathway flow dynamics by multidimensional phase-velocity magnetic resonance imaging
    Be'eri, E
    Maier, SE
    Landzberg, MJ
    Chung, T
    Geva, T
    [J]. CIRCULATION, 1998, 98 (25) : 2873 - 2882
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Lung perfusion scintigraphy in patients with congenital heart disease: Sensitivity and important pitfalls
    Boothroyd, AE
    McDonald, EA
    Carty, H
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 1996, 17 (01) : 33 - 39
  • [4] Use of computational fluid dynamics in the design of surgical procedures: Application to the study of competitive flows in cavopulmonary connections
    deLeval, MR
    Dubini, G
    Migliavacca, F
    Jalali, H
    Camporini, G
    Redington, A
    Pietrabissa, R
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 111 (03) : 502 - 510
  • [5] DELEVAL MR, 1988, J THORAC CARDIOV SUR, V96, P682
  • [6] Echocardiographic and radionuclide pulmonary blood flow patterns after transcatheter closure of patent ductus arteriosus
    Dessy, H
    Hermus, JPS
    vandenHeuvel, F
    Oei, HY
    Krenning, EP
    Hess, J
    [J]. CIRCULATION, 1996, 94 (02) : 126 - 129
  • [7] PALLIATION OF COMPLEX CARDIAC ANOMALIES WITH SUBAORTIC OBSTRUCTION - NEW OPERATIVE APPROACH
    DIDONATO, R
    DICARLO, DC
    GIANNICO, S
    MARCELLETTI, C
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (02) : 406 - 412
  • [8] Caval contribution to flow in the branch pulmonary arteries of Fontan patients with a novel application of magnetic resonance presaturation pulse
    Fogel, MA
    Weinberg, PM
    Rychik, J
    Hubbard, A
    Jacobs, M
    Spray, TL
    Haselgrove, J
    [J]. CIRCULATION, 1999, 99 (09) : 1215 - 1221
  • [9] SURGICAL REPAIR OF TRICUSPID ATRESIA
    FONTAN, F
    BAUDET, E
    [J]. THORAX, 1971, 26 (03) : 240 - +
  • [10] Kozerke S, 1999, MAGNET RESON MED, V42, P970, DOI 10.1002/(SICI)1522-2594(199911)42:5<970::AID-MRM18>3.0.CO