The Reliability of a Scoring System for Corrosion and Fretting, and Its Relationship to Material Loss of Tapered, Modular Junctions of Retrieved Hip Implants

被引:63
作者
Hothi, Harry S.
Matthies, Ashley K.
Berber, Reshid
Whittaker, Robert K.
Skinner, John A.
Hart, Alister J.
机构
[1] Univ Coll London, Inst Orthopaed & Musculoskeletal Sci, Stanmore, Middx, England
[2] Royal Natl Orthopaed Hosp, Stanmore HA7 4LP, Middx, England
关键词
taper; corrosion; fretting; metal-on-metal; hip; wear; LOCAL TISSUE REACTION; WEIGHTED KAPPA; NECK; AGREEMENT; BEARINGS; SIZE;
D O I
10.1016/j.arth.2013.12.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
It has been suggested that corrosion and fretting at the tapered, modular junctions of hip arthroplasties may contribute to implant failure. In this study the reliability of a commonly used peer-reviewed scoring system for visual assessment of corrosion and fretting at these junctions was evaluated. Volumetric material loss at the tapered head surface was measured and associations with the visual scores were investigated. We found that the inter-observer reproducibility and single-observer repeatability of the corrosion scores were substantial using Cohen's weighted Kappa statistic (k = 0.64-0.71). The reproducibility and repeatability of the fretting scores however were slight to fair (k = 0.18-0.31). Taper corrosion scores were significantly and moderately correlated with the volume of material loss measured (Spearman's r = 0.59; P < 0.001). We recommend the continued use of this scoring system but it should not be a substitute for measurement of material loss. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:1313 / 1317
页数:5
相关论文
共 23 条
[1]  
Abbott EJ., 1933, ASME Mech. Engr, V55, P569, DOI DOI 10.4236/WJM.2013.32007
[2]  
Australian Orthopaedic Association National Joint Replacement Registry, 2013, ANN REPORT
[3]  
Bills PJ, 2013, 19 INT C WEAR MAT 14
[4]   Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings [J].
Bishop, Nicholas ;
Witt, Florian ;
Pourzal, Robin ;
Fischer, Alfons ;
Ruetschi, Marcel ;
Michel, Markus ;
Morlock, Michael .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (07) :1116-1122
[5]   High friction moments in large hard-on-hard hip replacement bearings in conditions of poor lubrication [J].
Bishop, Nicholas E. ;
Hothan, Arne ;
Morlock, Michael M. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (05) :807-813
[7]   Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck [J].
Cooper, H. John ;
Urban, Robert M. ;
Wixson, Richard L. ;
Meneghini, R. Michael ;
Jacobs, Joshua J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10) :865-872
[8]  
Cross MB, 2013, BONE JOINT J B S15, V95-B, P127
[9]   The Influence of Head Size on Corrosion and Fretting Behaviour at the Head-Neck Interface of Artificial Hip Joints [J].
Dyrkacz, Richard M. R. ;
Brandt, Jan-M. ;
Ojo, Olanrewaju A. ;
Turgeon, Thomas R. ;
Wyss, Urs P. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (06) :1036-1040
[10]   EQUIVALENCE OF WEIGHTED KAPPA AND INTRACLASS CORRELATION COEFFICIENT AS MEASURES OF RELIABILITY [J].
FLEISS, JL ;
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1973, 33 (03) :613-619