Bringing back the old: time to reevaluate the high-frequency ventilation strategy

被引:23
作者
Mukerji, A. [1 ]
Belik, J. [1 ]
Sanchez-Luna, M. [2 ]
机构
[1] Univ Toronto, Hosp Sick Children, Dept Pediat, Div Neonatol, Toronto, ON M5G 1X8, Canada
[2] Univ Complutense Madrid, Hosp Gen Univ Gregorio Maranon, Inst Invest Sanitaria Gregorio Maranon, Div Neonatol, Madrid, Spain
关键词
pressure-volume index; ventilation efficacy; CO2; elimination; tidal volume; OSCILLATORY VENTILATION; GAS-EXCHANGE; PRESSURE; NEWBORN; INFANTS;
D O I
10.1038/jp.2014.39
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
OBJECTIVE: To examine the role of frequency in high-frequency ventilation (HFV) on carbon-dioxide (CO2) elimination and lung injury, independent of its effect on tidal volume. STUDY DESIGN: An anatomically representative lung model was attached to a mechanical ventilator capable of providing HFV with a constant volume. CO2 was infused directly into the lung, and a commercially available end-tidal CO2 detector was used to determine CO2 elimination. CO2 elimination and amplitude of pressure transmissions were evaluated using frequencies ranging from 5 to 15 Hz. The pressure-volume index (PVI) was described as the product of the volume and pressures delivered to the lung, a surrogate for lung injury. RESULT: The use of increasing frequencies directly correlated with improved CO2 clearance when keeping the tidal volume fixed, expressed as percent CO2 remaining in the lung at 255 (66.5 (+/- 1.1)%, 50.5 (+/- 0.1)% and 37.8 (+/- 0.3)% at 5, 10 and 15 Hz, respectively, P < 0.05). With a fixed tidal volume, there was a decrease in pressure amplitudes transmitted to the lung with a decline in the PVI (53.9 ( +/- 2.7) mmHg ml(-1), 41.1 (+/- 0.9) mmHg ml(-1) and 23.4 (+/- 3.6) mmHg ml(-1), at 5, 10 and 15 Hz, respectively, P < 0.05). CONCLUSION: Frequency has a direct relationship with CO2 elimination when tidal volume is fixed. Using low delivered tidal volumes and high frequencies may allow for improved ventilation efficacy, while minimizing lung injury.
引用
收藏
页码:464 / 467
页数:4
相关论文
共 21 条
[1]   GAS-EXCHANGE IN HEALTHY RABBITS DURING HIGH-FREQUENCY OSCILLATORY VENTILATION [J].
BOYNTON, BR ;
HAMMOND, MD ;
FREDBERG, JJ ;
BUCKLEY, BG ;
VILLANUEVA, D ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1989, 66 (03) :1343-1351
[2]   Volutrauma activates the clotting cascade in the newborn but not adult rat [J].
Chan, A ;
Jayasuriya, K ;
Berry, L ;
Kleiner, MR ;
Post, M ;
Belik, J .
AMERICAN JOURNAL OF PHYSIOLOGY-LUNG CELLULAR AND MOLECULAR PHYSIOLOGY, 2006, 290 (04) :L754-L760
[3]   Elective high-frequency oscillatory versus conventional ventilation in preterm infants: a systematic review and meta-analysis of individual patients' data [J].
Cools, Filip ;
Askie, Lisa M. ;
Offringa, Martin ;
Asselin, Jeanette M. ;
Calvert, Sandra A. ;
Courtney, Sherry E. ;
Dani, Carlo ;
Durand, David J. ;
Gerstmann, Dale R. ;
Henderson-Smart, David J. ;
Marlow, Neil ;
Peacock, Janet L. ;
Pillow, J. Jane ;
Soll, Roger F. ;
Thome, Ulrich H. ;
Truffert, Patrick ;
Schreiber, Michael D. ;
Van Reempts, Patrick ;
Vendettuoli, Valentina ;
Vento, Giovanni .
LANCET, 2010, 375 (9731) :2082-2091
[4]   High tidal volume ventilation causes different inflammatory responses in newborn versus adult lung [J].
Copland, IB ;
Martinez, F ;
Kavanagh, BP ;
Engelberts, D ;
McKerlie, C ;
Belik, J ;
Post, M .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2004, 169 (06) :739-748
[5]   Early changes in lung gene expression due to high tidal volume [J].
Copland, IB ;
Kavanagh, BP ;
Engelberts, D ;
McKerlie, C ;
Belik, J ;
Post, M .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2003, 168 (09) :1051-1059
[6]   High-frequency oscillatory ventilation versus conventional mechanical ventilation for very-low-birth-weight infants [J].
Courtney, SE ;
Durand, DJ ;
Asselin, JM ;
Hudak, ML ;
Aschner, JL ;
Shoemaker, CT .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 347 (09) :643-652
[7]   Volume delivery during high frequency oscillation [J].
Dimitriou, G ;
Greenough, A ;
Kavvadia, V ;
Laubscher, B ;
Milner, AD .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1998, 78 (02) :F148-F150
[8]  
FROESE AB, 1987, AM REV RESPIR DIS, V135, P1363, DOI 10.1007/BF03009896
[9]   OPTIMAL APPLICATION OF HIGH-FREQUENCY VENTILATION IN INFANTS - A THEORETICAL-STUDY [J].
GHAZANSHAHI, SD ;
KHOO, MCK .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (08) :788-796
[10]   Mechanical performance of clinically available, neonatal, high-frequency, oscillatory-type ventilators [J].
Hatcher, D ;
Watanabe, H ;
Ashbury, T ;
Vincent, S ;
Fisher, J ;
Froese, A .
CRITICAL CARE MEDICINE, 1998, 26 (06) :1081-1088