Failure of a cemented all-polyethylene patellar component of a Press-Fit Condylar total knee arthroplasty

被引:16
作者
Francke, EI [1 ]
Lachiewicz, PF [1 ]
机构
[1] Univ N Carolina, Dept Orthopaed, Chapel Hill, NC 27599 USA
关键词
Press-Fit Condylar modular total knee; all-polyethylene patella;
D O I
10.1016/S0883-5403(00)90396-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A 65-year-old woman had a posterior cruciate-retaining cemented Press-Fit Condylar (PFC; Johnson & Johnson Professional, Inc., Raynham, MA) knee replacement with resurfacing of the patella using a cemented all-polyethylene, 3-peg component. Five years postoperatively, the patient developed anterior knee pain. The patella tracked normally, but radiographs showed a shift of the patellar component. At revision, there was shear failure of all 3 fixation pegs and loosening of the component. Revision with a cemented Press-Fit Condylar Sigma patellar 3-pegs component (Johnson & Johnson Professional, Inc., Raynham, MA) was performed.
引用
收藏
页码:234 / 237
页数:4
相关论文
共 9 条
[1]  
DENNIS DA, 1995, ORTHOPAEDIC KNOWLEDG, P283
[2]  
ELBERT K, 1995, CLIN ORTHOP RELAT R, P71
[3]   FATIGUE-CRACK PROPAGATION BEHAVIOR OF ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE UNDER MIXED-MODE CONDITIONS [J].
ELBERT, KE ;
WRIGHT, TM ;
RIMNAC, CM ;
KLEIN, RW ;
INGRAFFEA, AR ;
GUNSALLUS, K ;
BARTEL, DL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (02) :181-187
[4]  
Kitsugi T, 1994, J Arthroplasty, V9, P151, DOI 10.1016/0883-5403(94)90064-7
[5]  
MCNAMARA JL, 1994, CLIN ORTHOP RELAT R, P104
[6]  
Oishi CS, 1996, CLIN ORTHOP RELAT R, P283
[7]   The press-fit condylar modular total knee system - Four-to-six-year results with a posterior-cruciate-substituting design [J].
Ranawat, CS ;
Luessenhop, CP ;
Rodriguez, JA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (03) :342-348
[8]   Total knee arthroplasty with the PFC system - Results at a minimum of ten years and survivorship analysis [J].
Schai, PA ;
Thornhill, TS ;
Scott, RD .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (05) :850-858
[9]   PATELLOFEMORAL CONTACT PRESSURES EXCEED THE COMPRESSIVE YIELD STRENGTH OF UHMWPE IN TOTAL KNEE ARTHROPLASTIES [J].
TAKEUCHI, T ;
LATHI, VK ;
KHAN, AM ;
HAYES, WC .
JOURNAL OF ARTHROPLASTY, 1995, 10 (03) :363-368