Diffusion-weighted echo planar MR imaging of the neck at 3 T using integrated shimming: comparison of MR sequence techniques for reducing artifacts caused by magnetic-field inhomogeneities

被引:21
|
作者
Gatidis, Sergios [1 ,4 ]
Graf, Hansjoerg [2 ]
Weiss, Jakob [1 ]
Stemmer, Alto [2 ]
Kiefer, Berthold [2 ]
Nikolaou, Konstantin [1 ]
Notohamiprodjo, Mike [1 ]
Martirosian, Petros [3 ]
机构
[1] Univ Tubingen, Dept Radiol Diagnost & Intervent Radiol, Tubingen, Germany
[2] Siemens Healthcare GmbH, Erlangen, Germany
[3] Univ Tubingen, Dept Radiol, Expt Radiol, Tubingen, Germany
[4] Univ Tubingen Hosp, Dept Radiol, Hoppe Seyler Str 3, D-72076 Tubingen, Germany
关键词
MRI; Diffusion weighted; Echo-planar Imaging; Artifacts; Neck; BRAIN; LESIONS; HEAD; EPI;
D O I
10.1007/s10334-016-0582-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Our objective was to compare available techniques reducing artifacts in echo planar imaging (EPI)-based diffusion-weighed magnetic resonance imaging MRI (DWI) of the neck at 3 Tesla caused by B0-field inhomogeneities. A cylindrical fat-water phantom was equipped with a Maxwell coil allowing for additional linear B0-field variations in z-direction. The effect of increasing strength of this superimposed gradient on image quality was observed using a standard single-shot EPI-based DWI sequence (sEPI), a zoomed single-shot EPI sequence (zEPI), a readout-segmented EPI sequence (rsEPI), and an sEPI sequence with integrated dynamic shimming (intEPI) on a 3-Tesla system. Additionally, ten volunteers were examined over the neck region using these techniques. Image quality was assessed by two radiologists. Scan durations were recorded. With increasing strength of the external gradient, marked distortions, signal loss, and failure of fat suppression were observed using sEPI, zEPI, and rsEPI. These artifacts were markedly reduced using intEPI. Significantly better in vivo image quality was also observed using intEPI compared with the other techniques. Scan time of intEPI was similar to sEPI and zEPI and shorter than rsEPI. The use of integrated 2D shim and frequency adjustment for EPI-based DWI results in a significant improvement in image quality of the head/neck region at 3 Tesla. Combining integrated shimming with rsEPI or zEPI can be expected to provide additional improvements.
引用
收藏
页码:57 / 63
页数:7
相关论文
共 50 条
  • [41] Correlation Between Apparent Diffusion Coefficient Values of the Renal Parenchyma and Estimated Glomerular Filtration Rates on 3-T Diffusion-Weighted Echo-Planar Magnetic Resonance Imaging
    Thiravit, Shanigarn
    Suwanchatree, Paphatsanan
    Skulratanasak, Peenida
    Thiravit, Phakphoom
    Suvannarerg, Voraparee
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2019, 43 (05) : 780 - 785
  • [42] Evaluation of image quality of diffusion weighted readout segmentation of long variable echo-trains MR pulse sequence for lumbosacral nerve imaging at 3T br
    Abdulaal, Osamah M.
    MacMahon, Peter J.
    Rainford, Louise
    Cradock, Andrea
    O'Driscoll, Dearbhail
    Galligan, Marie
    Alshoabi, Sultan A.
    Alsharif, Walaa
    McGee, Allison
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (01) : 196 - 209
  • [43] Comparison of accuracy of intravoxel incoherent motion and apparent diffusion coefficient techniques for predicting malignancy of head and neck tumors using half-Fourier single-shot turbo spin-echo diffusion-weighted imaging
    Sakamoto, Junichiro
    Imaizumi, Akiko
    Sasaki, Yoshinori
    Kamio, Takashi
    Wakoh, Mamoru
    Otonari-Yamamoto, Mika
    Sano, Tsukasa
    MAGNETIC RESONANCE IMAGING, 2014, 32 (07) : 860 - 866
  • [44] 3-T Breast Diffusion-Weighted MRI by Echo-Planar Imaging With Spectral Spatial Excitation or With Additional Spectral Inversion Recovery: An In Vivo Comparison of Image Quality
    Jacobsen, Megan C.
    Dogan, Basak E.
    Adrada, Beatriz E.
    Plaxco, Jeri Sue
    Wei, Wei
    Son, Jong Bum
    Hazle, John D.
    Ma, Jingfei
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (03) : 343 - 348
  • [45] Diffusion-weighted MRI of the lung at 3T evaluated using echo-planar-based and single-shot turbo spin-echo-based acquisition techniques for radiotherapy applications
    Tyagi, Neelam
    Cloutier, Michelle
    Zakian, Kristen
    Deasy, Joseph O.
    Hunt, Margie
    Rimner, Andreas
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01): : 284 - 292
  • [46] Evaluation of a new Spin-echo diffusion-weighted sequence on a 0.35 T open magnetic resonance imaging (MRI)-system: first experiences within 3 h after acute stroke
    W. A. Wohlgemuth
    G. Schulte-Altedorneburg
    T. Becker
    L. Zha
    D. Kramer
    K. Kirchhof
    Neuroradiology, 2005, 47 : 532 - 538
  • [47] MR imaging of the liver using an ultrafast 3D multi-shot EPI sequence - Comparison with four other T1-weighted sequences
    Papanikolaou, N
    Karantanas, A
    Lavdas, E
    ACTA RADIOLOGICA, 1999, 40 (03) : 322 - 325
  • [48] Evaluation of a new Spin-echo diffusion-weighted sequence on a 0.35 T open magnetic resonance imaging (MRI)-system: first experiences within 3 h after acute stroke
    Wohlgemuth, WA
    Schulte-Altedorneburg, G
    Becker, T
    Zha, L
    Kramer, D
    Kirchhof, K
    NEURORADIOLOGY, 2005, 47 (07) : 532 - 538
  • [49] Evaluation of diffusion-weighted magnetic resonance imaging of the rectal cancers: comparison between modified reduced field-of-view single-shot echo-planar imaging with tilted two-dimensional radiofrequency excitation pulses and conventional full field-of-view readout-segmented echo-planar imaging
    Inoue, Atsuo
    Tanabe, Masahiro
    Ihara, Kenichiro
    Hideura, Keiko
    Higashi, Mayumi
    Benkert, Thomas
    Imai, Hiroshi
    Yamane, Masatoshi
    Yamaguchi, Takahiro
    Ueda, Takaaki
    Ito, Katsuyoshi
    RADIOLOGIA MEDICA, 2023, 128 (10): : 1192 - 1198
  • [50] Cervical carotid plaque evaluation using 3D T1-weighted black-blood magnetic resonance imaging: Comparison of turbo field-echo and turbo spin-echo sequences
    Inoue, Katsuhiro
    Maeda, Masayuki
    Umino, Maki
    Takase, Shinichi
    Yamahata, Tsunehiro
    Sakuma, Hajime
    EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (05) : 1035 - 1039